[论文解读] Deformable Scintillation Dosimeter II: Real-Time Simultaneous Measurements of Dose and Tracking of Deformation Vector Fields

Émily Cloutier, Luc Beaulieu|arXiv (Cornell University)|Jan 21, 2021
Advanced Radiotherapy TechniquesPhysics and Astronomy参考文献 38被引用 10
一句话总结

该论文提出了一种实时、水等效的可变形闪烁探测器,能够以0.3毫米的精度同时测量三维剂量分布并追踪形变矢量场(DVFs)。通过在弹性体基质中嵌入19根闪烁纤维阵列并结合立体视觉相机追踪,系统实现了亚毫米级的DVF精度(与DIR算法偏差≤1.5毫米),剂量测量结果与治疗计划系统计算值的偏差在1%以内。

ABSTRACT

Anatomical motion and deformation pose challenges to the understanding of the delivered dose distribution during radiotherapy treatments. Hence, deformable image registration (DIR) algorithms are increasingly used to map contours and dose distributions from one image set to another. However, the lack of validation tools slows their clinical adoption, despite their commercial availability. This work presents a novel water-equivalent deformable dosimeter that simultaneously measures the dose distribution and tracks deformation vector fields (DVF). The dosimeter in made of an array of 19 scintillating fiber detectors embedded in a cylindrical elastomer matrix. It is imaged by two pairs of stereoscopic cameras tracking the position and angulation of the scintillators, while measuring the dose. The resulting system provides a precision of 0.3 mm on DVF measurements. The dosimeter was irradiated with 5$ imes$3, 4$ imes$3 and 3$ imes$3 cm$^2$ 6 MV photon beams in both fixed and deformed conditions. The measured DVF was compared to the one computed with a DIR algorithm (Plastimatch). The deviations between the computed and measured DVFs was below 1.5 mm. As for dose measurements, the dosimeter acquired the dose distribution in fixed and deformed conditions within 1\% of the treatment planning system calculation and complementary dose validation using the Hyperscint dosimetry system. Using the demonstrated qualities of scintillating detectors, we developed a real-time, water-equivalent deformable dosimeter. Given it's sensor tracking position precision and dose measurements accuracy, the developed detector is a promising tools for the validation of DIR algorithms as well as dose distribution measurements under fixed and deformed conditions.

研究动机与目标

  • 开发一种能够同时实现剂量与形变追踪的实时、水等效剂量计。
  • 解决临床放射治疗中可变形图像配准(DIR)算法缺乏患者特异性、金标准验证工具的问题。
  • 实现在解剖形变条件下准确的体内类似剂量累积验证。
  • 克服非水等效探测器和凝胶集成方法的局限性,提供实时、点对点的测量。
  • 证明利用闪烁纤维结合计算机视觉实现动态剂量与形变监测的可行性。

提出的方法

  • 该剂量计由19根水等效闪烁纤维(BCF-60)嵌入透明弹性体基质(Clearflex30)构成。
  • 立体视觉相机对(CCD与sCMOS)实时追踪每个闪烁体的三维位置与方向。
  • 通过测量闪烁光强度来确定各纤维位置的吸收剂量。
  • 利用相机数据,通过三维重建算法计算由闪烁体位移得出的形变矢量场(DVFs)。
  • 使用6 MV光子束(5×3、4×3、3×3 cm²)在固定状态与压缩状态(形变状态)下对系统进行了验证。
  • 将剂量与DVF测量结果与治疗计划系统(TPS)计算值及一种DIR算法(Plastimatch)进行比较。

实验结果

研究问题

  • RQ1在可变形基质中,闪烁纤维阵列能否实现实时、水等效的剂量与形变追踪?
  • RQ2利用立体视觉相机追踪闪烁体位置测量DVF的精度如何?
  • RQ3测量得到的DVFs与临床DIR算法(Plastimatch)计算结果的一致性如何?
  • RQ4在固定与形变条件下,剂量测量结果与TPS计算值的吻合程度如何?
  • RQ5该系统能否作为自适应放射治疗中DIR算法的金标准验证工具?

主要发现

  • 该剂量计通过立体视觉相机追踪实现了0.3毫米的DVF测量精度。
  • 在所有射野尺寸与工况下,测量DVF与Plastimatch计算DVF的偏差均低于1.5毫米。
  • 在固定与形变构型下,剂量测量结果与TPS计算值的偏差均在1%以内。
  • 系统实现了实时、同步的剂量与形变数据采集,且未影响水等效性。
  • 通过Hyperscint剂量计系统的互补验证,证实了闪烁纤维剂量测量的准确性。
  • 将闪烁体与计算机视觉技术结合,可实现一种实用、高精度的工具,用于验证运动解剖中的DIR与剂量累积。