[论文解读] Development of a one-dimensional position sensitive detector for Compton X-ray polarimeters

Abhay Kumar, Santosh V. Vadawale|arXiv (Cornell University)|Mar 26, 2026
Particle Detector Development and PerformancePhysics and Astronomy被引用 0
一句话总结

本文提出使用 SiPM 的双端读出 NaI(Tl) 薄晶体吸收器,实现硬 X 射线 Compton 偏振计 (CXPOL) 的 1-D 相互作用位置与能量测量。

ABSTRACT

The scientific potential of X-ray polarimetry has long been recognized, but the challenges in measuring polarization have left it largely unexplored, particularly in the hard X-ray regime. While tremendous advancement has been made in soft X-ray polarimetery, the lack of sensitive hard X-ray polarimeters and polarisation measurements continues to limit our understanding of high-energy astrophysical processes. With the development of hard X-ray mirrors, it is now possible to develop a sensitive focal plane hard X-ray polarimeter. One such effort is CXPOL, a prototype developed at PRL, India, which consists of a plastic scintillator as active scatterer readout by PMT surrounded by CsI(Tl) scintillators in cylindrical array with SiPM readout from one side. First results of the prototype have been demonstrated in 20 to 80 keV energy range. The sensitivity of the instrument can be significantly enhanced using faster and better light yield scintillator like NaI as absorbers. Further, the use of a position-sensitive scatterer and absorbers, can also provide spectroscopic information by measuring the interaction position along the length and from the known energy depositions in the detectors. Position sensitive detectors are also helpful in mitigating the systematic effects introduced by the off-axis events in the polarisation measurements. Here, we demonstrate the detection sensitivity in the 100x20x5 mm^3 NaI(Tl) scintillator absorber readout on both ends by SiPM arrays operating in co-incidence. In this work, we characterize the first prototype of this detector system and investigate the variation in energy and position resolution, and light output with irradiation position along the length of the detector. The two end readout in co-incidence also reduces the overall SiPM background per absorber by an order of magnitude, further enhancing the polarimetric sensitivity of the instrument.

研究动机与目标

  • 推动硬 X 射线偏振测量及在 20–80 keV 范围内实现光谱-极化能力的需求。
  • 开发 1-D 位置敏感吸收探测器以提升偏振测量并提供谱信息。
  • 用 NaI(Tl) 代替 CsI(Tl 吸收体并实现双端 SiPM 读出以获得全长灵敏度。
  • 利用 Am-241 59.54 keV 光子沿晶体长度演示探测器性能。
  • 在 coincidenc e 模式下评估能量分辨率、位置定位、效率及背景抑制。

提出的方法

  • 将 CsI(Tl) 吸收体替换为 NaI(Tl) 以提升光产额和衰变时间。
  • 在 100×20×5 mm3 NaI(Tl) 薄晶体两端放置两组 SiPM 阵列,通过端对端信号比 ADC2/(ADC1+ADC2) 获得 1-D 位置信息。
  • 将晶体两端与石英窗和光凝胶耦合以改善光收集。
  • 设计两块板卡读出:电荷敏感前置放大+整形,以及基于 FPGA 的 coincid ence/峰值检测链,时间窗口为 1 µs。
  • 使用中心照射和 Am-241 同步读取来标定增益对称性,校正两端 ADC 增益差异。
  • 通过将两端信号相加 (ADC1+ADC2) 来分析能量,并用两个高斯拟合 59.54 keV 与碘逸出峰谱。

实验结果

研究问题

  • RQ1两端 1-D 位置敏感 NaI(Tl) 吸收器是否能在晶体长度上提供均匀灵敏度以用于硬 X 射线偏振测量?
  • RQ2在 20–80 keV 带内,双端 SiPM 读出下的能量分辨率、位置精度与 coincid ence 效率可达到何种水平?
  • RQ3两端 coincid ence 读出对 SiPM 背景与整体极化灵敏度有何影响?
  • RQ4为达到 ~20 keV 阈值和更好的背景抑制,需要在哪些方面(晶体、包裹、光学、电子学)进行优化?

主要发现

  • 59.54 keV 的能量质心在接近 SiPM 端部处显示较高值,晶体长度中心处较低,表明长度方向的光收集存在变化。
  • 59.54 keV 的能量分辨率在端部约为 30%,在中心约为 40%,沿长度的平均约为 34%。
  • 1-D 平均空间定位分辨率约为 1.55 cm,端部附近约为 1.04–1.09 cm。
  • 两端 coincid ence 可将 SiPM 背景降低约 10 倍,提升极化灵敏度。
  • 在晶体长度方向的两端 coincid ence 模式下,探测器在 60 keV 处相对探测效率 >60%。