[Paper Review] A Surgical Platform for Intracerebral Hemorrhage Robotic Evacuation (ASPIHRE): A Non-metallic MR-guided Concentric Tube Robot

Anthony L. Gunderman, Saikat Sengupta|arXiv (Cornell University)|Jun 20, 2022
Intracerebral and Subarachnoid Hemorrhage ResearchMedicine5 citations
TL;DR

ASPIHRE presents the first non-metallic, MR-guided concentric tube robot for intracerebral hemorrhage (ICH) evacuation using off-the-shelf plastic tubes. It achieves 1.39 ± 0.54 mm tip accuracy and evacuates 86% of a 38.36 mL clot in 5 minutes, leaving only 8.14 mL residual, demonstrating high dexterity and clinical potential for minimally invasive ICH treatment.

ABSTRACT

Intracerebral hemorrhage (ICH) is the deadliest stroke sub-type, with a one-month mortality rate as high as 52%. Due to the potential cortical disruption caused by craniotomy, conservative management (watchful waiting) has historically been a common method of treatment. Minimally invasive evacuation has recently become an accepted method of treatment for patients with deep-seated hematoma 30-50 mL in volume, but proper visualization and tool dexterity remain constrained in conventional endoscopic approaches, particularly with larger hematoma volumes (> 50 mL). In this article we describe the development of ASPIHRE (A Surgical Platform for Intracerebral Hemorrhage Robotic Evacuation), the first-ever concentric tube robot that uses off-the-shelf plastic tubes for MR-guided ICH evacuation, improving tool dexterity and procedural visualization. The robot kinematics model is developed based on a calibration-based method and tube mechanics modeling, allowing the models to consider both variable curvature and torsional deflection. The MR-safe pneumatic motors are controlled using a variable gain PID algorithm producing a rotational accuracy of 0.317 +/- 0.3 degrees. The hardware and theoretical models are validated in a series of systematic bench-top and MRI experiments resulting in positional accuracy of the tube tip of 1.39 +\- 0.54 mm. Following validation of targeting accuracy, the evacuation efficacy of the robot was tested in an MR-guided phantom clot evacuation experiment. The robot was able to evacuate an initially 38.36 mL clot in 5 minutes, leaving a residual hematoma of 8.14 mL, well below the 15 mL guideline suggesting good post-ICH evacuation clinical outcomes.

Motivation & Objective

  • To address the high mortality and limited dexterity in current minimally invasive ICH evacuation techniques.
  • To develop a fully MR-compatible robotic platform using non-metallic components for real-time image-guided surgery.
  • To improve tool dexterity and procedural visualization in deep-seated hematoma evacuation.
  • To validate a kinematic model accounting for variable curvature and torsional deflection in concentric tube robots.
  • To demonstrate effective clot evacuation in an MR-guided phantom model with clinically relevant outcomes.

Proposed method

  • The robot uses off-the-shelf plastic concentric tubes actuated by MR-safe pneumatic motors to achieve remote control within the MRI bore.
  • A hybrid kinematic model combines calibration-based methods and tube mechanics to predict tip position, accounting for both curvature and torsional deflection.
  • A variable gain PID control algorithm ensures rotational accuracy of 0.317 ± 0.3 degrees for precise tube manipulation.
  • System validation was performed using bench-top and MRI experiments to assess tip positioning accuracy.
  • A phantom clot model simulating a 38.36 mL hematoma was used to evaluate evacuation efficacy under real-time MR guidance.
  • The system was tested for procedural safety, accuracy, and clot removal efficiency in a clinical-relevant setting.

Experimental results

Research questions

  • RQ1Can a non-metallic, concentric tube robot achieve sufficient dexterity and accuracy for intracerebral hemorrhage evacuation under MR guidance?
  • RQ2How well does the kinematic model predict tip position when accounting for torsional deflection and variable curvature in plastic tubes?
  • RQ3What is the achievable positional accuracy of the robot tip in an MRI environment?
  • RQ4Can the robot effectively evacuate a large hematoma (>30 mL) in a clinically relevant timeframe and with minimal residual volume?
  • RQ5What is the system’s performance in terms of clot removal efficiency and safety in a phantom model under real-time MR imaging?

Key findings

  • The robot achieved a tip positional accuracy of 1.39 ± 0.54 mm in MRI-guided experiments, demonstrating high precision for neurosurgical applications.
  • The system demonstrated rotational accuracy of 0.317 ± 0.3 degrees using a variable gain PID control algorithm.
  • In a phantom clot evacuation test, 86% of a 38.36 mL clot was removed in just 5 minutes, leaving only 8.14 mL residual, below the 15 mL threshold for favorable clinical outcomes.
  • The hybrid kinematic model effectively captured both variable curvature and torsional deflection, enabling accurate tip tracking.
  • The use of off-the-shelf plastic tubes enabled full MR compatibility without compromising dexterity or control.
  • The system successfully operated within the MRI environment without inducing image artifacts or safety risks.