[Paper Review] Calibration of Advanced Virgo and Reconstruction of the Gravitational Wave Signal h(t) during the Observing Run O2
This paper presents the calibration of Advanced Virgo during O2 and the reconstruction of the gravitational wave signal h(t), using laser wavelength as a primary reference and a calibration transfer method. It validates the reconstructed h(t) signal's amplitude and phase against photodiode calibrator (PCal) injections, achieving 5.1% amplitude uncertainty and 40 mrad phase uncertainty, with residual bias identified and included in systematic errors for future improvement.
In August 2017, Advanced Virgo joined Advanced LIGO for the end of the O2 run, leading to the first gravitational waves detections with the three-detector network. This paper describes the Advanced Virgo calibration and the gravitational wave strain h(t) reconstruction during O2. The methods are the same as the ones developed for the initial Virgo detector and have already been described in previous publications, this paper summarizes the differences and emphasis is put on estimating systematic uncertainties. Three versions of the h(t) signal have been computed for the Virgo O2 run, an online version and two post-run reprocessed versions with improved detector calibration and reconstruction algorithm. A photon calibrator has been used to establish the sign of h(t) and to make an independent partial cross-check of the systematic uncertainties. The uncertainties reached for the latest h(t) version are 5.1% in amplitude, 40 mrad in phase and 20 microseconds in timing.
Motivation & Objective
- To calibrate the mirror actuators and optical components (BS, NE, WE, PR, marionette) in Advanced Virgo during O2 using a laser wavelength reference.
- To reconstruct the gravitational wave signal h(t) by subtracting actuator control signals from the dark fringe signal.
- To quantify and characterize systematic uncertainties in the reconstructed h(t) signal for accurate gravitational wave parameter estimation.
- To validate the reconstructed h(t) against photodiode calibrator (PCal) injections to assess signal fidelity and uncertainty.
- To reduce future calibration uncertainties to the percent level to support high-SNR detections in O3 and beyond.
Proposed method
- Used the interferometer's laser wavelength as a primary calibration standard to anchor the transfer function of the system.
- Applied a calibration transfer procedure to propagate the laser-based calibration to mirror actuators and optical components.
- Injected known signals via the photodiode calibrator (PCal) to compare with reconstructed h(t) and validate signal reconstruction.
- Fitted the transfer function between PCal injections and reconstructed h(t) with a notch filter at 2 kHz to model frequency-dependent response.
- Estimated uncertainties in amplitude (5.1%) and phase (40 mrad) using error bands from timing and phase measurements.
- Accounted for timing delays (±7 μs) from PCal and dark fringe photodiodes, calibrated via GPS timing.
Experimental results
Research questions
- RQ1How accurately can the gravitational wave signal h(t) be reconstructed in Advanced Virgo during O2 using actuator control signals?
- RQ2To what extent does the photodiode calibrator (PCal) injection validate the reconstructed h(t) signal's amplitude and phase?
- RQ3What are the dominant sources of systematic uncertainty in the reconstructed h(t) channel, and how are they quantified?
- RQ4How do the frequency-dependent deviations in h(t) reconstruction compare between PCal and standard mirror actuator calibrations?
- RQ5What improvements are needed to reduce calibration uncertainties to the percent level for future O3 observations?
Key findings
- The reconstructed h(t) signal's amplitude is consistent with PCal injections within a ±4% error band, validating the relative amplitude calibration.
- The phase of the reconstructed h(t) signal lies within a ±40 mrad uncertainty band, with timing uncertainty contributing ±7 μs to the phase error budget.
- Frequency-dependent deviations in the h(t) reconstruction match those observed with standard mirror actuators, indicating a real bias in the signal reconstruction.
- The PCal validation confirms the correct sign and phase of h(t), with the bias in amplitude response included as a systematic uncertainty.
- The calibration uncertainty for the latest reprocessing version is 5.1% in amplitude and 40 mrad in phase, with 20 μs timing uncertainty.
- Future work will reduce uncertainties via improved actuator calibration, a new Newtonian calibrator, and refined optical modeling to achieve sub-percent-level accuracy.