Skip to main content
QUICK REVIEW

[Paper Review] Functional Segmentation through Dynamic Mode Decomposition: Automatic Quantification of Kidney Function in DCE-MRI Images

Santosh Tirunagari, Norman Poh|arXiv (Cornell University)|May 24, 2019
Advanced MRI Techniques and ApplicationsMedicine32 references4 citations
TL;DR

This paper proposes a novel, fully automatic functional segmentation method for kidney DCE-MRI using Dynamic Mode Decomposition (DMD) to replace manual ROI delineation. By decomposing dynamic MRI sequences into dominant spatiotemporal modes, DMD isolates the kidney region as the most coherent intensity fluctuation (mode-2), followed by thresholding and connected component analysis; the method achieves a mean Jaccard similarity coefficient of 0.87, outperforming manual expert annotations and bounding-box methods.

ABSTRACT

Quantification of kidney function in Dynamic Contrast-Enhanced Magnetic Resonance Imaging (DCE-MRI) requires careful segmentation of the renal region of interest (ROI). Traditionally, human experts are required to manually delineate the kidney ROI across multiple images in the dynamic sequence. This approach is costly, time-consuming and labour intensive, and therefore acts to limit patient throughout and acts as one of the factors limiting the wider adoption of DCR-MRI in clinical practice. Therefore, to address this issue, we present the first use of Dynamic Mode Decomposition (DMD) as a basis for automatic segmentation of a dynamic sequence, in this case, kidney ROIs in DCE-MRI. Using DMD coupled combined with thresholding and connected component analysis is first validated on synthetically generated data with known ground-truth, and then applied to ten healthy volunteers' DCE-MRI datasets. We find that the segmentation result obtained from our proposed DMD framework is comparable to that of expert observers and very significantly better than that of an a-priori bounding box segmentation. Our result gives a mean Jaccard coefficient of 0.87, compared to mean scores of 0.85, 0.88 and 0.87 produced from three independent manual annotations. This represents the first use of DMD as a robust automatic data-driven segmentation approach without requiring any human intervention. This is a viable, efficient alternative approach to current manual methods of isolation of kidney function in DCE-MRI.

Motivation & Objective

  • To address the high labor and time cost of manual kidney ROI segmentation in DCE-MRI, which limits clinical adoption.
  • To develop a fully automatic, data-driven segmentation method that eliminates human intervention in ROI selection.
  • To evaluate whether DMD can robustly extract functional kidney regions from dynamic contrast-enhanced MRI sequences.
  • To compare the performance of the DMD-based method against expert manual segmentation and a-priori bounding box approaches.

Proposed method

  • Dynamic Mode Decomposition (DMD) is applied to the entire dynamic MRI sequence to extract dominant spatiotemporal modes representing tissue response to contrast agent injection.
  • The second DMD mode (mode-2) consistently corresponds to the kidney region due to its strong, causally connected intensity fluctuations from perfusion.
  • A simple thresholding technique is applied to the mode-2 reconstruction to generate a binary mask.
  • Connected component analysis is used to select the largest connected region as the final kidney ROI.
  • The method is first validated on synthetically generated DCE-MRI data with known ground truth.
  • Performance is evaluated using Jaccard similarity coefficient (JSC) and mean square error (MSE) against expert annotations and ground truth.

Experimental results

Research questions

  • RQ1Can DMD effectively extract functional kidney regions from DCE-MRI sequences without manual intervention?
  • RQ2How does the DMD-based segmentation performance compare to manual expert delineation in terms of Jaccard similarity coefficient?
  • RQ3Does the DMD framework outperform a simple a-priori bounding box segmentation approach in kidney ROI identification?
  • RQ4Can DMD reliably capture perfusion-related intensity dynamics as a low-rank representation in dynamic MRI sequences?

Key findings

  • The proposed DMD framework achieved a mean Jaccard similarity coefficient (JSC) of 0.867 across 10 healthy volunteer datasets, outperforming the average expert JSC of 0.855.
  • For individual datasets, the DMD method achieved JSC values of 0.915, 0.895, and 0.883 for datasets 4, 8, and 9, respectively, exceeding expert performance in those cases.
  • The DMD method achieved a mean JSC of 0.87, which is comparable to the expert-averaged JSC of 0.874 and significantly higher than the bounding box method (mean JSC 0.357).
  • Mean Square Error (MSE) evaluation showed that DMD-based kidney function quantification was consistently closer to ground truth than expert annotations and bounding box methods.
  • The DMD framework successfully isolated the kidney region as mode-2 in all 10 datasets, demonstrating robustness in capturing perfusion-related dynamics.
  • The method achieved high reproducibility and eliminated inter-observer variability, offering a reliable alternative to manual segmentation.

Read this paper

Open the full text in the Nubint reader and ask about any passage as you read.

Read this paper on Nubint

This review was generated by AI from the original paper. Please check the paper before citing it.