[Paper Review] Indian Payloads (RT-2 Experiment) Onboard CORONAS-PHOTON Mission
This paper presents the in-flight performance and initial scientific results of the RT-2 experiment, an Indian-Russian collaborative low-energy gamma-ray instrument aboard the CORONAS-PHOTON mission. It comprises three detectors—two phoswich scintillators (RT-2/S, RT-2/G) and a CZT-based solid-state imager (RT-2/CZT)—capable of detecting solar flares and GRBs in the 15 keV–1 MeV range with energy resolutions of 18% and 8%, respectively, and successfully detected GRB 090618 with a 1-second time resolution light curve.
RT-2 Experiment (RT - Roentgen Telescope) is a low energy gamma-ray instrument which is designed and developed as a part of Indo-Russian collaborative project of CORONAS-PHOTON Mission to study the Solar flares in wide energy band of electromagnetic spectrum ranging from UV to high-energy gamma-rays (~2000 MeV). RT-2 instruments will cover the energy range of 15 keV to 150 keV extendable up to ~1 MeV. It consists of three detectors (two Phoswich detectors, namely, RT-2/S, RT-2/G and one solid-state imaging detector RT-2/CZT) and one processing electronic device (RT-2/E). Both Phoswich detectors will have time resolved spectrum, whereas the solid-state imaging detector will have high resolved image of the solar flares in hard X-rays. We have used Co-57 (122 keV) radio-active source for onboard calibration of all three detectors. In this paper, we briefly discuss the in-flight performance of RT-2 instruments and present initial flight data from the instruments. This mission was launched into polar LEO (Low Earth Orbit) (~550 km) on 30th January 2009 from Plesetsk Cosmodrome, Russia.
Motivation & Objective
- To study solar flares across a broad energy spectrum from UV to high-energy gamma-rays using a low-energy gamma-ray instrument.
- To develop and fly a multi-detector system capable of time-resolved spectroscopy and hard X-ray imaging of solar flares.
- To validate in-orbit performance of RT-2 detectors using onboard calibration sources and shadow mode data.
- To detect and characterize transient high-energy events such as gamma-ray bursts (GRBs).
- To achieve high spectral and temporal resolution for solar flare monitoring in low Earth polar orbit.
Proposed method
- The RT-2 experiment uses three primary detectors: two phoswich scintillators (RT-2/S and RT-2/G) with NaI(Tl)/CsI(Na) crystals viewed by PMTs, and a CZT-based solid-state imaging detector (RT-2/CZT) with 256-pixel arrays.
- Phoswich detectors employ pulse shape discrimination (PSD) at 26 and 28 channels to separate NaI and CsI signals, enabling energy spectroscopy from 15 keV to 1 MeV.
- The RT-2/CZT detector uses three CZT modules (2.5 mm × 2.5 mm pixels) and a CMOS detector (50 μm pixels) with coded aperture and Fresnel zone plate collimators for hard X-ray imaging.
- All detectors are calibrated in-flight using a Co-57 source (122 keV) embedded in the collimator slat, with energy and time response validated over 1661 orbits.
- Data is processed by the RT-2/E electronic unit, stored in 10 MB memory, and downlinked twice daily via ground station communication.
- Instrument health and performance are monitored via temperature profiles and background data collected during 'shadow mode' (Sun not visible).
Experimental results
Research questions
- RQ1How do the RT-2 phoswich and solid-state detectors perform in the low Earth polar orbit environment during solar flare observations?
- RQ2Can the RT-2 instrument detect and resolve the temporal and spectral characteristics of gamma-ray bursts such as GRB 090618?
- RQ3What is the energy resolution and spectral stability of the CZT-based imaging detector (RT-2/CZT) during on-orbit operation?
- RQ4How effective is pulse shape discrimination in separating NaI and CsI signals in the phoswich detectors across the 15–150 keV range?
- RQ5What causes pixel noise in the CZT modules, and how does it affect imaging performance and background threshold calibration?
Key findings
- The RT-2/S and RT-2/G phoswich detectors successfully achieved time-resolved spectroscopy with 1-second and 100-millisecond time resolution, respectively, and demonstrated stable pulse shape discrimination.
- The RT-2/CZT detector achieved an energy resolution of 5.6% at 122 keV for the Co-57 calibration peak, with other modules showing 5.97% and 7.32% resolution at 124 keV and 127 keV, respectively.
- The CZT2 module detected the Co-57 calibration peak at 122 keV with high stability, confirming in-flight calibration accuracy and detector functionality.
- The RT-2 instruments detected GRB 090618 with a 1-second time resolution light curve, showing a multi-peak profile consistent with observations from SWIFT, KONUS-RF, and Fermi.
- Background spectra below 40 keV were dominated by noise in all CZT modules, with some noisy pixels persisting even after threshold adjustment at 40 keV.
- Temperature variations during orbit were within safe operational limits: 18–25 °C for RT-2/S and RT-2/G, and 2–11 °C for RT-2/CZT, indicating stable thermal performance.