Skip to main content
QUICK REVIEW

[Paper Review] Noninvasive three-dimensional optoacoustic localization microangiography of deep tissues

Xosé Luís Deán‐Ben, Justine Robin|arXiv (Cornell University)|Jul 1, 2020
Photoacoustic and Ultrasonic ImagingEngineering51 references12 citations
TL;DR

This study presents noninvasive three-dimensional optoacoustic microangiography of deep tissues using highly absorbing dichloromethane microdroplets (5 μm) that enable sub-20 μm resolution beyond the acoustic diffraction limit in intact mouse brains. By leveraging single-particle sensitivity and high 3D frame rates, the method achieves deep-tissue capillary-level imaging and enables in situ estimation of light fluence in heterogeneous tissues.

ABSTRACT

Structural microvascular alterations and dysfunction serve as key disease indicators of cancer, diabetes, ischemic stroke, neurodegenerative disorders and many other conditions. In vivo visualization of the microvasculature has traditionally been restricted to millimeter scale depths accessible with optical microscopy. Optoacoustic imaging has enabled breaking through the barrier imposed by light diffusion to map functional hemodynamic parameters in deep-seated vessels, but diffraction and dispersion of ultrasound waves in heterogeneous living tissues prevents reaching capillary resolution. Herein, we demonstrate three-dimensional microangiography of deep mouse brain beyond the acoustic diffraction limit (<20$μ$m resolution) through the intact scalp and skull via optoacoustic localization of sparsely-distributed highly absorbing microparticles. This was enabled by devising 5$μ$m sized extremely absorbing dichloromethane microdroplets exhibiting four orders of magnitude higher optical absorption than red blood cells at near-infrared wavelengths, thus facilitating in vivo compatibility and single particle sensitivity in the presence of highly absorbing blood background. Accurate mapping of the blood flow velocity within microvascular structures was also facilitated by the high 3D frame rate of the optoacoustic tomography system. We further show that the detected optoacoustic signal intensities from the localized particles may serve for estimating the light fluence distribution within optically heterogeneous tissues, a long-standing quantification challenge in biomedical optics. Given the intrinsic sensitivity of optoacoustics to various functional, metabolic and molecular events in living tissues, this new approach paves the way for non-invasive deep-tissue microscopic observations with unrivaled resolution, contrast and speed.

Motivation & Objective

  • To overcome the resolution and depth limitations of conventional optoacoustic imaging in deep tissues.
  • To enable noninvasive, three-dimensional microvascular imaging at capillary resolution (below 20 μm) through intact scalp and skull.
  • To develop a method for in vivo quantification of light fluence distribution in optically heterogeneous tissues.
  • To achieve high-speed 3D mapping of blood flow velocity in microvasculature using optoacoustic tomography.
  • To demonstrate single-particle sensitivity for microparticles in the presence of highly absorbing blood background.

Proposed method

  • Employed 5 μm diameter dichloromethane microdroplets with four orders of magnitude higher optical absorption than red blood cells at near-infrared wavelengths.
  • Utilized a high-frame-rate optoacoustic tomography system to enable real-time 3D imaging of microvascular structures.
  • Applied optoacoustic localization of sparse, highly absorbing microparticles to achieve sub-diffraction-limited resolution (<20 μm) in deep tissues.
  • Leveraged the high sensitivity of optoacoustics to detect individual microparticles against a strong blood background.
  • Used detected optoacoustic signal intensities to estimate light fluence distribution within optically heterogeneous tissues.
  • Combined spatial localization of microparticles with 3D reconstruction to generate high-contrast, high-resolution microangiograms.

Experimental results

Research questions

  • RQ1Can optoacoustic imaging achieve capillary-level resolution in deep tissues through intact skull and scalp?
  • RQ2Can highly absorbing microparticles enable single-particle sensitivity in the presence of a strongly absorbing blood background?
  • RQ3Can optoacoustic signal intensities be used to estimate light fluence in optically heterogeneous tissues?
  • RQ4Can high 3D frame rates enable accurate in vivo mapping of blood flow velocity in microvasculature?
  • RQ5Is it feasible to achieve noninvasive, three-dimensional microangiography with sub-20 μm resolution in deep brain tissue?

Key findings

  • The method achieved three-dimensional microangiography of deep mouse brain with sub-20 μm resolution, surpassing the acoustic diffraction limit.
  • The 5 μm dichloromethane microdroplets exhibited four orders of magnitude higher optical absorption than red blood cells at near-infrared wavelengths.
  • Single-particle sensitivity was achieved in the presence of a highly absorbing blood background, enabling detection of sparse microparticles.
  • High 3D frame rates allowed for accurate in vivo mapping of blood flow velocity within microvascular structures.
  • Detected optoacoustic signal intensities enabled estimation of light fluence distribution in optically heterogeneous tissues.
  • The technique enabled noninvasive, high-contrast, and high-speed 3D visualization of microvasculature in deep tissues through intact scalp and skull.

Read this paper

Open the full text in the Nubint reader and ask about any passage as you read.

Read this paper on Nubint

This review was generated by AI from the original paper. Please check the paper before citing it.