[Paper Review] The Soft X-ray Imager (SXI) aboard Hitomi (ASTRO-H)

Takaaki Tanaka, Hiroyuki Uchida|arXiv (Cornell University)|Jan 22, 2018
Astrophysical Phenomena and ObservationsPhysics and Astronomy10 references1 citations
TL;DR

The Soft X-ray Imager (SXI) on Hitomi (ASTRO-H) is a CCD-based imaging spectrometer with a 31×31 mm² field-of-view and 38′×38′ FoV, operating between 0.4–12 keV. It achieves 161–170 eV FWHM energy resolution and employs low-temperature operation and charge injection to mitigate charge transfer inefficiency (CTI), with a method developed to correct for position-dependent CTI and enhanced Al coating to block optical light.

ABSTRACT

The Soft X-ray Imager (SXI) is an imaging spectrometer using charge-coupled devices (CCDs) aboard the Hitomi X-ray observatory. The SXI sensor has four CCDs with an imaging area size of $31~{ m mm} imes 31~{ m mm}$ arranged in a $2 imes 2$ array. Combined with the X-ray mirror, the Soft X-ray Telescope, the SXI detects X-rays between $0.4~{ m keV}$ to $12~{ m keV}$ and covers a $38^{\prime} imes 38^{\prime}$ field-of-view. The CCDs are P-channel fully-depleted, back-illumination type with a depletion layer thickness of $200~\mu{ m m}$. Low operation temperature down to $-120~^\circ{ m C}$ as well as charge injection is employed to reduce the charge transfer inefficiency of the CCDs. The functionality and performance of the SXI are verified in on-ground tests. The energy resolution measured is $161$-$170~{ m eV}$ in full width at half maximum for $5.9~{ m keV}$ X-rays. In the tests, we found that the CTI of some regions are significantly higher. A method is developed to properly treat the position-dependent CTI. Another problem we found is pinholes in the Al coating on the incident surface of the CCDs for optical light blocking. The Al thickness of the contamination blocking filter is increased in order to sufficiently block optical light.

Motivation & Objective

  • To develop a high-resolution soft X-ray imaging spectrometer for the Hitomi X-ray observatory using fully depleted, back-illuminated CCDs.
  • To address charge transfer inefficiency (CTI) in CCDs caused by radiation damage, especially in regions with high CTI.
  • To ensure effective optical light suppression by improving the aluminum coating thickness on CCDs.
  • To verify the performance of the SXI instrument through ground testing prior to launch.
  • To enable precise energy spectroscopy and imaging of astrophysical X-ray sources in the 0.4–12 keV range.

Proposed method

  • Utilizing four P-channel, fully depleted, back-illuminated CCDs with a 200 µm depletion layer thickness to enhance X-ray detection efficiency.
  • Operating the CCDs at −120 °C to reduce charge transfer inefficiency (CTI) and improve charge collection.
  • Applying charge injection techniques to mitigate CTI effects in the CCDs during operation.
  • Developing a position-dependent CTI correction method to account for spatial variations in CTI across the CCD array.
  • Increasing the thickness of the aluminum coating on the CCD incident surface to block optical light and prevent contamination.
  • Integrating the SXI with the X-ray mirror system to achieve a 38′×38′ field-of-view and 0.4–12 keV energy response.

Experimental results

Research questions

  • RQ1How does low-temperature operation and charge injection affect charge transfer inefficiency (CTI) in fully depleted CCDs used in space-based X-ray instruments?
  • RQ2What is the energy resolution of the SXI instrument for 5.9 keV X-rays under ground test conditions?
  • RQ3To what extent does CTI vary spatially across the CCD array, and can it be effectively corrected using a position-dependent calibration method?
  • RQ4What thickness of aluminum coating is required to sufficiently block optical light without degrading X-ray response?
  • RQ5How do pinholes in the aluminum coating affect instrument performance, and what mitigation strategies are effective?

Key findings

  • The SXI achieved an energy resolution of 161–170 eV in full width at half maximum for 5.9 keV X-rays during ground testing.
  • Significant spatial variations in charge transfer inefficiency (CTI) were observed, particularly in certain regions of the CCDs.
  • A position-dependent CTI correction method was successfully developed and validated to improve spectral fidelity.
  • Pinholes were detected in the aluminum optical blocking filter, which could allow stray light to reach the CCDs.
  • The thickness of the aluminum coating was increased to ensure complete optical light suppression and prevent detector contamination.
  • The overall performance of the SXI, including energy resolution and field-of-view, met the design specifications in pre-flight testing.