[论文解读] Sensor-Movement-Robust Angle Estimation for 3-DoF Lower Limb Joints Without Calibration
本文提出了一种无需标定、对传感器运动具有鲁棒性的3-DoF下肢关节角估计方法,利用惯性测量单元(IMUs)实现。该方法采用基于反馈的迭代算法估计关节轴,并通过融合重力与磁场矢量的逐点参考帧校准,实现了在2分钟平地行走中误差低于2°的均方根误差,且对传感器运动和错位具有鲁棒性。
Inertial measurement unit (IMU)-based 3-DoF angle estimation methods for lower limb joints have been studied for decades, however the calibration motions and/or careful sensor placement are still necessary due to challenges of real-time application. This study proposes a novel sensormovement-robust 3-DoF method for lower-limb joint angle estimation without calibration. A realtime optimization process, which is based on a feedback iteration progress to identify three joint axes of a 3-DoF joint, has been presented with a reference frame calibration algorithm, and a safe-guarded strategy is proposed to detect and compensate for the errors caused by sensor movements. The experimental results obtained from a 3-DoF gimbal and ten healthy subjects demonstrate a promising performance on 3-DoF angle estimation. Specially, the experiments on ten subjects are performed with three gait modes and a 2-min level walking. The root mean square error is below 2 deg for level walking and 5 deg for other two gait modes. The result of 2-min level walking shows our algorithms stability under long run. The robustness against sensor movement are demonstrated through data from multiple sets of IMUs. In addition, results from the 3-DoF gimbal indicate that the accuracy of 3-DoF angle estimation could be improved by 84.9% with our reference frame calibration algorithm. In conclusion, our study proposes and validates a sensor-movement-robust 3-DoF angle estimation for lowerlimb joints based on IMU. To the best of our knowledge, our approach is the first experimental implementation of IMUbased 3-DoF angle estimation for lower-limb joints without calibration.
研究动机与目标
- 消除在使用IMU进行3-DoF下肢关节角估计时对功能标定姿势的需求。
- 解决IMU与地球坐标系之间参考帧偏差的问题,该偏差会导致关节角估计产生较大误差。
- 实现实时检测并补偿运动过程中的传感器移动,以维持估计精度。
- 提高IMU基关节角估计对传感器错位和动态干扰的鲁棒性。
- 开发一种自对齐、无标定的方法,适用于临床和可穿戴设备中的长期、实时应用。
提出的方法
- 基于反馈的迭代优化过程实时估计3-DoF下肢关节的三个关节轴,无需依赖功能标定姿势。
- 采用逐点参考帧校准算法,利用融合的重力与磁场矢量,校正传感器与参考帧之间的时变偏差。
- 通过监控一个变化度量 $ V $ 的安全防护检测策略,识别传感器移动,当该度量超过预设阈值时判定为发生移动。
- 在检测到移动时,利用检测窗口内的估计值,在临时区间 $ I_n^{det} $ 内维持连续性并减少误差传播。
- 该算法将实时姿态跟踪与生物力学约束相结合,在身体固定坐标系中计算3-DoF关节角。
- 采用滑动窗口方法,自适应调整区间长度 $ I_n ∈ [W_n, 800] $,在实时性能与估计精度之间取得平衡。
实验结果
研究问题
- RQ1能否仅使用IMU数据实现无需功能标定姿势的3-DoF下肢关节角估计?
- RQ2如何在实时中有效补偿IMU与地球坐标系之间参考帧的偏差?
- RQ3能否在运动过程中实时检测并在线校正传感器移动,以维持估计精度?
- RQ4该算法在长时间连续运动(如2分钟平地行走)下的性能如何?
- RQ5与现有基于IMU的方法相比,该方法在传感器错位和移动方面的鲁棒性如何?
主要发现
- 在2分钟平地行走过程中,该算法在髋关节屈伸方向的均方根误差(RMSE)为1.69° ± 0.90°,在内收/外展方向为0.52° ± 0.33°,在内旋/外旋方向为0.79° ± 0.43°。
- 在非平地步态模式下,RMSE保持在5°以下,最大误差为5.73°(膝关节屈伸方向)。
- 在传感器移动事件中,该算法将RMSE从检测前的43.60°降低至检测区间内的13.97°,显示出有效的误差缓解能力。
- 参考帧校准算法在3-DoF云台上的实验中,使3-DoF角度估计精度相比基线方法提升了84.9%。
- 该算法在长时间运行中保持稳定性能,跨受试者在2分钟测试中,膝关节屈伸方向的标准差仅从0.77°略微增加至0.83°。
- 计算时间始终在实时约束范围内,不同滑动窗口长度下每轮迭代的最大耗时为6 ms,证实了其在实时部署中的可行性。