Professor Jaeman Son's research lab specializes in advancing image-guided and adaptive radiotherapy through innovative technologies and patient-specific solutions. The lab focuses on the clinical integration of magnetic resonance-guided radiotherapy (MRgRT), particularly investigating the dosimetric effects of low-field (0.35 T) magnetic resonance systems on stereotactic ablative radiotherapy (SABR) for lung and liver cancers. A key research direction involves developing and validating patient-specific bolus techniques using 3D printing and biocompatible materials to improve dose conformity in superficial targets. The lab also pioneers respiratory motion management systems to enhance the precision of breath-hold and gating techniques in external beam radiotherapy.
Research Overview
Research Output Trend
Figures are computed from collected data and may differ slightly.
Selected Papers
15Background: We investigated the impact of 0.35 T magnetic field on dose calculation for nonsmall cell lung cancer (NSCLC) stereotactic ablative radiotherapy (SABR) in the ViewRay system (ViewRay Inc.), which features a simultaneous use of magnetic resonance imaging (MRI) to guide radiotherapy for an improved targeting of tumors. Materials and Methods: Here, we present a comprehensive analysis of the effects induced by the 0.35 T magnetic field on various characteristics of SABR plans including t
Purpose: The objective of this study is to evaluate the geometrical accuracy of a patient-specific bolus based on a three-dimensional (3D) printed mold and casting method. Materials and Methods: Three breast cancer patients undergoing treatment for a superficial region were scanned using computed tomography (CT) and a designed bolus structure through a treatment planning system (TPS). For the fabrication of patient-specific bolus, we cast harmless certified silicone into 3D printed molds. The pr
To investigate the effect of low magnetic field on dose distribution in SABR plans for liver cancer, we calculated and evaluated the dose distribution to each organ with and without magnetic fields. Ten patients received a 50 Gy dose in five fractions using the ViewRay? treatment planning system. For planning target volume (PTV), the results were analyzed in the point minimum (Dmin), maximum (Dmax), mean dose (Dmean) and volume receiving at least 90% (V90%), 95% (V95%), and 100% (V100%) of the p
Purpose: This study aimed to develop a breath control training system for breath-hold technique and respiratory-gated radiation therapy wherein the patients can learn breath-hold techniques in their convenient environment. Methods: The breath control training system comprises a sensor device and software. The sensor device uses a loadcell sensor and an adjustable strap around the chest to acquire respiratory signals. The device connects via Bluetooth to a computer where the software is installed
Purpose: This study aimed to comprehensively investigate the diverse characteristics of a novel commercial bolus, CLEANBOLUS-WHITE (CBW), to ascertain its suitability for clinical application. Methods: The evaluation of CBW encompassed both physical and biological assessments. Physical parameters such as mass density and shore hardness were measured alongside analyses of element composition. Biological evaluations included assessments for skin irritation and cytotoxicity. Dosimetric properties w
CPAP demonstrated both safety and efficacy for breast cancer RT, achieving significant reductions in cardiac and pulmonary radiation exposure. These findings support further investigation of CPAP as a novel respiratory motion management strategy. Future studies are warranted to identify optimal CPAP pressure levels to facilitate broader clinical implementation.
Purpose: Brachytherapy is essential for treating gynecological cancers as it offers precise radiation delivery to tumors while minimizing radiation exposure to surrounding healthy tissues. The Geneva applicator, introduced in 2020 as a replacement for older models like the Utrecht applicator, enhances MRI-based brachytherapy with improved imaging capabilities and more accurate applicator placement. In 2021, updates to non-reimbursement policies in Korea for MRI-based 3D brachytherapy planning fu
Purpose: This study evaluated the quality of plans based on magnetic resonance-guided radiotherapy (MRgRT) tri-Co-60, linac, and conventional linac-based volumetric modulated arc therapy (linac-VMAT) for prostate cancer. Methods: Twenty patients suffering from prostate cancer with intermediate risk who were treated by MAT were selected. Additional treatment plans (primary and boost plans) were generated based on MRgRT-tri-Co-60 and MRgRT-linac. The planning target volume (PTV) of MRgRT-based pla
This study compared the dose calculated using the electron Monte Carlo (eMC) dose calculation algorithm employing the old version (eMC V13.7) of the Varian Eclipse treatment-planning system (TPS) and its newer version (eMC V16.1). The eMC V16.1 was configured using the same beam data as the eMC V13.7. Beam data measured using the VitalBeam linear accelerator were implemented. A box-shaped water phantom (30×30×30 cm3) was generated in the TPS. Consequently, the TPS with eMC V13.7 and eMC V16.1 ca
The proper position of a multi-leaf collimator (MLC) is essential for the quality of intensity-modulated radiation therapy (IMRT) and volumetric modulated arc radiotherapy (VMAT) dose delivery. Task Group (TG) 142 provides a quality assurance (QA) procedure for MLC position. Our study investigated the QA validation of the mechanical leaf gap measurement and the maintenance procedure. Two VitalBeamTM systems were evaluated to validate the acceptance of an MLC position. The dosimetric leaf gaps (D
The manufacturer of a linear accelerator (LINAC) has reported that the target melting phenomenon could be caused by a non-recommended output setting and the excessive use of monitor unit (MU) with intensity-modulated radiation therapy (IMRT). Due to these reasons, we observed an unexpected beam interruption during the treatment of a patient in our institution. The target status was inspected and a replacement of the target was determined. After the target replacement, the beam profile was adjust
The objective of this study is to monitor the radiation doses delivered to a cardiac implantable electronic device (CIED) by comparing the absorbed doses calculated by a commercial treatment planning system (TPS) to those measured by an in vivo dosimeter. Accurate monitoring of the radiation absorbed by a CIED during radiotherapy is necessary to prevent damage to the device. We conducted this study on three patients, who had the CIED inserted and were to be treated with radiotherapy. Treatment p
This study aimed to evaluate and verify a process for correcting the extended source-to-imager distance (SID) in portal dosimetry (PD). In this study, eight treatment plans (four volumetric modulated arc therapy and four intensity-modulated radiation therapy plans) at different treatment sites and beam energies were selected for measurement. A Varian PD system with portal dose image prediction (PDIP) was used for the measurement and verification. To verify the integrity of the plan, independent
Purpose: We investigated the properties of CLEANBOLUS based on silicone with suitable characteristics for clinical use. Methods: We evaluated the characteristics of CLEANBOLUS and compared the results with the commercial product (Super-Flex bolus). Also, we conducted physical evaluations, including shore hardness, element composition, and elongation break. Transparency was investigated through the measured absorbance within the visible region (400–700 nm). Also, dosimetric characteristics were i
This study examined the clinical use of two newly installed computed tomography (CT) simulators in the Department of Radiation Oncology. The accreditation procedure was performed by the Korean Institute for Accreditation of Medical Imaging. An Xi R/F dosimeter was used to measure the CT dose index for each plug of the CT dose index phantom. Image qualities such as the Hounsfield unit (HU) value of water, noise level, homogeneity, existence of artifacts, spatial resolution, contrast, and slice th