세이이치 ヤマモ토 교수의 연구실은 주로 방사선 영상 진단 기술, 특히 흡수위치 측정(PET) 및 자기공명영상(MRI)과의 융합 기술에 초점을 맞추고 있습니다. 고해상도 소형 동물용 PET 시스템 개발과 함께, MR-가용성 및 자기장에 대한 강인성을 확보한 광검출기(예: Si-PM, GSO/LGSO/LSO 등)의 응용 연구를 진행하고 있습니다. 또한 뇌 기능 영상 및 탄소선 치료의 범위 예측을 위한 빛 방출 영상 기술 등 임상적 응용을 겨냥한 혁신적 기술 개발도 함께 추진하고 있습니다.
표시된 성과는 수집된 데이터 기준으로 산출되며, 일부 차이가 있을 수 있습니다.
A Geiger-mode avalanche photodiode (Si-PM) is a promising photodetector for PET, especially for use in a magnetic resonance imaging (MRI) system, because it has high gain and is less sensitive to a static magnetic field. We developed a Si-PM-based depth-of-interaction (DOI) PET system for small animals. Hamamatsu 4 × 4 Si-PM arrays (S11065-025P) were used for its detector blocks. Two types of LGSO scintillator of 0.75 mol% Ce (decay time: ∼45 ns; 1.1 mm × 1.2 mm × 5 mm) and 0.025 mol% Ce (decay
The authors have developed a GSO depth of interaction detector. Three GSO scintillators with different concentration of Ce were chosen and stacked along depth of interaction direction and optically coupled to a photomultiplier tube (PMT) or a position sensitive photomultiplier tube (PSPMT). Pulse shape, light output and zero-cross time were measured to select a proper combination of GSO scintillators with different concentration of Ce. Selecting 3 GSO scintillators with different concentration o
Luminescence imaging during carbon-ion irradiation of water is not only possible but also a promising method for range estimation in carbon-ion therapy.
We investigated the magnetic susceptibility of several scintillators to select the best material for MR-compatible gamma detectors. NaI(Tl) in an Al container, NaI(Tl) in a Cu container and CsI(Tl) are candidates for MR-compatible gamma probes, imaging probes or SPECT, while BGO, LSO, LGSO, and GSO are the candidates for MR-compatible PET. In addition, we investigated image artifacts and distortion by scanning these scintillators with an MR-scanner. Results show that the magnetic susceptibility
Brain functional studies using PET have advantages over fMRI in some areas such as auditory research in part because PET systems produce no acoustic noise during acquisition. However commercially available PET systems are designed for whole body studies and are not optimized for brain functional studies. We developed a low cost, small, wearable brain PET system named PET-Hat dedicated for brain imaging. It employs double counter-balanced systems for mechanical supports of the detector ring while
Since a high resolution PET system is needed for small animal imaging, especially for mouse studies, we developed a new small animal PET system that decreased the size of the scintillators to less than 1 mm. Our developed PET system used 0.5 × 0.7 × 5 mm(3) LYSO pixels arranged in an 11 × 13 matrix to form a block with a 0.1 mm BaSO4 reflector between the pixels. Two LYSO blocks were optically coupled to two optical fiber based angled image guides. These LYSO blocks and image guides were coupled
Purpose: The authors previously reported successful luminescence imaging of water during proton irradiation and its application to range estimation. However, since the feasibility of this approach for carbon-ion irradiation remained unclear, the authors conducted luminescence imaging during carbon-ion irradiation and estimated the ranges. Methods: The authors placed a pure-water phantom on the patient couch of a carbon-ion therapy system and measured the luminescence images with a high-sensitivi
The silicon-photomultiplier (Si-PM) is a promising photodetector, especially for integrated PET/MRI systems, due to its small size, high gain, and low sensitivity to static magnetic fields. The major problem using a Si-PM-based PET system within the MRI system is the interference between the PET and MRI units. We measured the interference by combining a Si-PM-based PET system with a permanent-magnet MRI system. When the RF signal-induced pulse height exceeded the lower energy threshold level of
The silicon photomultiplier (Si-PM) is a promising photo-detector for PET for use in magnetic resonance imaging (MRI) systems because it has high gain and is insensitive to static magnetic fields. Recently we developed a Si-PM-based depth-of-interaction PET system for small animals and performed simultaneous measurements by combining the Si-PM-based PET and the 0.15 T permanent MRI to test the interferences between the Si-PM-based PET and an MRI. When the Si-PM was inside the MRI and installed a