Retinal Imaging and AnalysisMedicine14 references4 citations
TL;DR

This paper presents a computer-aided diagnostic (CAD) system for automated segmentation of sub-retinal surfaces and layers in optical coherence tomography (OCT) images from patients with diabetic macular edema (DME). The method combines a novel Fourier-domain Wiener deconvolution denoising technique with a model-based segmentation algorithm, achieving high correlation (r ≥ 0.7, R² ≥ 0.5) in sub-retinal layer thickness estimation, with mean errors of 2–18 μm in abnormal images and improved segmentation accuracy over wavelet-based denoising.

ABSTRACT

This paper presents a novel automated system that segments six sub-retinal layers from optical coherence tomography (OCT) image stacks of healthy patients and patients with diabetic macular edema (DME). First, each image in the OCT stack is denoised using a Wiener deconvolution algorithm that estimates the additive speckle noise variance using a novel Fourier-domain based structural error. This denoising method enhances the image SNR by an average of 12dB. Next, the denoised images are subjected to an iterative multi-resolution high-pass filtering algorithm that detects seven sub-retinal surfaces in six iterative steps. The thicknesses of each sub-retinal layer for all scans from a particular OCT stack are then compared to the manually marked groundtruth. The proposed system uses adaptive thresholds for denoising and segmenting each image and hence it is robust to disruptions in the retinal micro-structure due to DME. The proposed denoising and segmentation system has an average error of 1.2-5.8 $μm$ and 3.5-26$μm$ for segmenting sub-retinal surfaces in normal and abnormal images with DME, respectively. For estimating the sub-retinal layer thicknesses, the proposed system has an average error of 0.2-2.5 $μm$ and 1.8-18 $μm$ in normal and abnormal images, respectively. Additionally, the average inner sub-retinal layer thickness in abnormal images is estimated as 275$μm (r=0.92)$ with an average error of 9.3 $μm$, while the average thickness of the outer layers in abnormal images is estimated as 57.4$μm (r=0.74)$ with an average error of 3.5 $μm$. The proposed system can be useful for tracking the disease progression for DME over a period of time.

Motivation & Objective

  • To develop an automated CAD system for accurate segmentation of sub-retinal surfaces and layers in OCT images from patients with DME.
  • To improve the reliability of sub-retinal layer thickness estimation for clinical assessment of DME severity.
  • To compare the performance of a novel Fourier-domain denoising method with existing wavelet-based denoising (CDWT) in enhancing segmentation accuracy.
  • To enable faster, more efficient detection and monitoring of DME progression through automated thickness mapping.
  • To support clinical research by providing robust, reproducible metrics beyond macular thickness for evaluating treatment response.

Proposed method

  • A Fourier-domain structural error-based Wiener deconvolution algorithm is used to estimate noise parameters and denoise OCT images, enhancing signal-to-noise ratio (SNR) by over 12 dB.
  • The denoised images are segmented using a model-based approach that identifies 7 sub-retinal surfaces and 6 sub-retinal layers through optimized surface fitting.
  • The method preserves sub-retinal edge structures while suppressing background noise, achieving a PSNR of 23.5 dB.
  • Segmentation performance is evaluated using correlation coefficient (r) and coefficient of determination (R²) against manual segmentation.
  • The system is implemented on a 2.53 GHz Intel Core i3 with 3 GB RAM using MATLAB, processing each OCT image in under 35 seconds.
  • Thickness maps are generated by combining layer thickness distributions across the entire OCT image stack for longitudinal assessment.

Experimental results

Research questions

  • RQ1Does the proposed Fourier-domain denoising method improve sub-retinal surface segmentation accuracy compared to the CDWT wavelet-based method in OCT images with DME?
  • RQ2What is the correlation between automated and manual sub-retinal layer thickness measurements using the proposed segmentation algorithm?
  • RQ3How does segmentation accuracy vary between normal and DME-affected OCT images using the proposed denoising and segmentation pipeline?
  • RQ4To what extent does denoising quality affect the estimation of mean and distributional sub-retinal layer thicknesses in pathological images?
  • RQ5Can the proposed CAD system reliably estimate sub-retinal layer thicknesses with r ≥ 0.7 and R² ≥ 0.5 in both normal and DME-affected OCT images?

Key findings

  • The proposed Fourier-domain denoising method improves SNR by over 12 dB and achieves a PSNR of 23.5 dB, outperforming CDWT in segmentation accuracy despite lower PSNR.
  • The proposed denoising method preserves sub-retinal edges better than CDWT, which smooths edges and reduces segmentation reliability.
  • For abnormal OCT images, the proposed method achieves mean sub-retinal layer thickness estimation errors of 2–18 μm with r ≥ 0.7 and R² ≥ 0.5.
  • The CDWT-based approach results in significantly lower correlation (r as low as 0.23) and R² (as low as 0.1) for thickness distribution estimation, especially in abnormal images.
  • The proposed segmentation algorithm demonstrates robustness in normal images (0–2 μm error) and acceptable performance in abnormal images (2–18 μm error), with p-values < 0.001 for denoising and < 0.01 for CDWT-based segmentation.
  • The system processes each OCT image in under 35 seconds on a standard laptop, enabling practical clinical deployment for DME monitoring and research.