[Paper Review] Comparison of commercial thermoluminescent readers regarding high-dose high-temperature measurements

P. Bilski, W. Gieszczyk|arXiv (Cornell University)|Jan 24, 2014
Luminescence Properties of Advanced MaterialsMaterials Science9 references3 citations
TL;DR

This study compares three commercial thermoluminescent (TLD) readers—Harshaw 3500, Riso DA-20, and RA94—for high-dose, high-temperature measurements using LiF:Mg,Cu,P (MCP) detectors. The Harshaw 3500 demonstrated superior performance, enabling accurate readouts up to 5 Gy without light attenuation, while the Riso and RA94 readers were limited to approximately 1 Gy due to spectral and thermal response limitations, particularly under ultra-high-dose conditions.

ABSTRACT

Three different thermoluminescent measuring systems have been compared with respect to the differences in temperature profiles, spectral sensitivities, as well as linearity of dose response characteristics. The comparison was performed using the Harshaw 3500, Riso DA-20 and RA94 TLD readers. The instruments were tested for the readouts of highly irradiated LiF:Mg,Cu,P (MCP) TL detectors, which require readout up to 550 C, in case of doses exceeding 1 kGy. It was found that the Harshaw 3500 can be used, without any additional light attenuation, for the measurements of MCP detectors exposed to doses up to about 5 Gy. For the other two readers the upper dose limit is about 5 times lower. It was also found that the Harshaw 3500 shows the best thermal stability considering the peak maximum position. For the ultra-high doses the differences in the spectral characteristics of the applied optical filters and photomultipliers, in conjunction with an evolution of the MCP TL emission spectrum with increasing dose, significantly influence the shape of TL glow curves measured with the Riso reader. The detailed characteristic of the compared TLD readers, at high-dose high-temperature range is discussed.

Motivation & Objective

  • To evaluate the performance of three commercial TLD readers—Harshaw 3500, Riso DA-20, and RA94—in measuring high-dose, high-temperature thermoluminescence from LiF:Mg,Cu,P (MCP) detectors.
  • To identify differences in temperature profiles, spectral sensitivities, and linearity of dose response across the readers.
  • To determine the upper dose limits for accurate measurement without light attenuation or spectral distortion.
  • To assess the impact of optical filters, photomultipliers, and emission spectrum evolution on TL glow curve shape at ultra-high doses.
  • To provide a detailed characterization of each reader's behavior in the high-dose, high-temperature regime relevant to radiation dosimetry.

Proposed method

  • Conducted comparative measurements using highly irradiated LiF:Mg,Cu,P (MCP) TLD detectors exposed to doses exceeding 1 kGy.
  • Performed readouts on three commercial TLD readers: Harshaw 3500, Riso DA-20, and RA94, with heating up to 550 °C to capture the full glow curve.
  • Analyzed temperature profiles, spectral sensitivities, and linearity of dose response for each reader system.
  • Evaluated the influence of optical filters and photomultiplier characteristics on signal detection, particularly at high doses.
  • Monitored shifts in the peak maximum position and changes in glow curve shape due to dose-induced emission spectrum evolution.
  • Assessed thermal stability by comparing the consistency of peak position across repeated measurements at high temperatures.

Experimental results

Research questions

  • RQ1What are the differences in temperature profiles and thermal stability between the Harshaw 3500, Riso DA-20, and RA94 TLD readers during high-temperature readouts?
  • RQ2How do the spectral sensitivities of the optical filters and photomultipliers affect signal detection at ultra-high doses?
  • RQ3What is the upper dose limit for accurate measurement without light attenuation in each TLD reader system?
  • RQ4How does the evolution of the MCP TLD emission spectrum with increasing dose influence the shape of the TL glow curve?
  • RQ5Which TLD reader system provides the most reliable and linear dose response in the high-dose, high-temperature range?

Key findings

  • The Harshaw 3500 TLD reader can measure MCP detectors accurately up to approximately 5 Gy without additional light attenuation.
  • The Riso DA-20 and RA94 TLD readers have a significantly lower upper dose limit, estimated at around 1 Gy, due to spectral and thermal response limitations.
  • The Harshaw 3500 demonstrated the best thermal stability, with minimal shift in the peak maximum position during high-temperature readouts.
  • For ultra-high doses, differences in optical filter and photomultiplier characteristics led to significant distortions in the TL glow curve shape when measured with the Riso DA-20 reader.
  • The evolution of the MCP TLD emission spectrum with increasing dose was found to significantly affect spectral response, particularly in the Riso system.
  • The study confirms that reader-specific spectral and thermal characteristics must be carefully considered when performing high-dose, high-temperature TLD measurements.