[Paper Review] Shot-noise limited, supercontinuum based optical coherence tomography

Shreesha Rao D. S., Mikkel Jensen|arXiv (Cornell University)|Oct 11, 2020
Optical Coherence Tomography ApplicationsEngineering30 references11 citations
TL;DR

This paper demonstrates the first shot-noise limited supercontinuum-based spectral domain optical coherence tomography (SD-OCT) using a low-noise, femtosecond-pumped all-normal-dispersion (ANDi) fiber source. By reducing pulse-to-pulse relative intensity noise (RIN) to below 1.1%, the system achieves 5.9 µm axial resolution and significantly improved sensitivity, contrast, and penetration depth compared to commercial supercontinuum sources.

ABSTRACT

We present the first demonstration of shot-noise limited supercontinuum-based spectral domain optical coherence tomography (SD-OCT) with axial resolution of 5.9 $μ$m at a center wavelength of 1370 nm. Current supercontinuum-based SD-OCT systems cannot be operated in the shot-noise limited detection regime because of severe pulse-to-pulse relative intensity noise of the supercontinuum source. To overcome this disadvantage we have developed a low-noise supercontinuum source based on an all-normal dispersion (ANDi) fiber, pumped by a femtosecond laser. The noise performance of our 90 MHz ANDi supercontinuum source is compared to that of two commercial sources operating at 80 and 320 MHz repetition rate. We show that the low noise of the ANDi supercontinuum source improves the OCT images significantly in terms of both higher contrast, better sensitivity, and improved penetration. From SD-OCT imaging of skin, retina, and multi-layer stacks we conclude that supercontinuum-based SD-OCT can enter the domain of shot-noise limited detection.

Motivation & Objective

  • To overcome the fundamental limitation of high intensity noise in conventional supercontinuum (SC) sources that prevents SD-OCT systems from reaching the shot-noise limit.
  • To develop and characterize a low-noise supercontinuum source based on an all-normal-dispersion (ANDi) fiber pumped by a femtosecond laser.
  • To demonstrate that shot-noise limited detection is achievable in SC-based SD-OCT by minimizing pulse-to-pulse relative intensity noise (RIN).
  • To compare the performance of the ANDi-based SC source with commercial 80 MHz and 320 MHz SC sources in terms of RIN, sensitivity, and image quality.
  • To validate the system's performance through high-resolution imaging of biological samples including skin, retina, and multi-layer phantoms.

Proposed method

  • Employed a femtosecond-pulsed laser at 90 MHz repetition rate to pump an all-normal-dispersion (ANDi) fiber for generating a low-noise supercontinuum.
  • Measured the pulse-to-pulse relative intensity noise (RIN) of the ANDi-based source and two commercial SC sources (80 MHz and 320 MHz) using 45,135, 249,078, and 997,288 pulses, respectively.
  • Calculated RIN as RIN = σ/μ, where σ is the standard deviation and μ is the mean of the pulse energy peaks.
  • Conducted axial resolution measurements by scanning the optical path difference (OPD) and analyzing the full width at half maximum (FWHM) of the A-scan peaks.
  • Determined system sensitivity as the minimum detectable reflectivity yielding SNR = 1, accounting for detector saturation, spectrometer integration time, and noise contributions.
  • Applied shadow correction and Hamming windowing in post-processing to enhance image quality and dynamic range.

Experimental results

Research questions

  • RQ1Can a supercontinuum source based on an all-normal-dispersion (ANDi) fiber achieve pulse-to-pulse relative intensity noise (RIN) low enough to enable shot-noise limited detection in SD-OCT?
  • RQ2How does the RIN of the ANDi-based SC source compare to commercial 80 MHz and 320 MHz SC sources across the 1.25–1.5 µm wavelength range?
  • RQ3To what extent does the low RIN of the ANDi source improve OCT image contrast, sensitivity, and penetration depth compared to conventional SC sources?
  • RQ4Does the ANDi-based SC source enable axial resolution below 6 µm at a center wavelength of 1370 nm while maintaining low noise?
  • RQ5Can the theoretical sensitivity of the SD-OCT system be accurately predicted using the extended FDOCT model with RIN correction?

Key findings

  • The ANDi-based supercontinuum source achieved a pulse-to-pulse relative intensity noise (RIN) of less than 1.1% across the 1.25–1.5 µm range, more than an order of magnitude lower than commercial sources.
  • The 320 MHz commercial SC source exhibited RIN of 66% at 1.5 µm, while the 80 MHz source showed 29.2% RIN at the same wavelength, confirming the superiority of the ANDi source.
  • The axial resolution of the SD-OCT system was measured at 5.9 µm, with no significant broadening observed even at 1.8 mm optical path difference (OPD).
  • The system sensitivity roll-off was 8.8 dB/mm when using the ANDi source, outperforming the 9.5 dB/mm (320 MHz) and 8.0 dB/mm (80 MHz) commercial sources.
  • Shadow-corrected OCT images of skin, retina, and multi-layer phantoms showed significantly improved contrast and penetration depth with the ANDi source compared to commercial alternatives.
  • Theoretical sensitivity curves, based on measured parameters including P₀ = 10.4 mW, ρ = 0.5, η = 0.7, and τ = 9.2 µs, confirmed that the system operates close to the shot-noise limit with the ANDi source.