[Paper Review] Single-element dual-interferometer for precision inertial sensing
This paper presents the self-referenced single-element dual-interferometer (SEDI), a compact inertial sensor using deep frequency modulation (DFM) interferometry to achieve sub-picometer precision in displacement sensing across multiple degrees of freedom. By integrating a reference interferometer within the same optical element as the measurement interferometer, SEDI enables laser frequency noise cancellation and reduces system complexity, achieving sub-picometer sensitivity down to 2 mHz in a few-cubic-inch package suitable for space-based and ground-based gravity missions.
Tracking moving masses in several degrees of freedom with high precision and large dynamic range is a central aspect in many current and future gravitational physics experiments. Laser interferometers have been established as one of the tools of choice for such measurement schemes. Using sinusoidal phase modulation homodyne interferometry allows a drastic reduction of the complexity of the optical setup, a key limitation of multi-channel interferometry. By shifting the complexity of the setup to the signal processing stage, these methods enable devices with a size and weight not feasible using conventional techniques. In this paper we present the design of a novel sensor topology based on deep frequency modulation interferometry: the self-referenced single-element dual-interferometer (SEDI) inertial sensor, which takes simplification one step further by accommodating two interferometers in one optic. Using a combination of computer models and analytical methods we show that an inertial sensor with sub-picometer precision for frequencies above 10 mHz, in a package of a few cubic inches, seems feasible with our approach. Moreover we show that by combining two of these devices it is possible to reach sub-picometer precision down to 2 mHz. In combination with the given compactness, this makes the SEDI sensor a promising approach for applications in high precision inertial sensing for both next-generation space-based gravity missions employing drag-free control, and ground-based experiments employing inertial isolation systems with optical readout. © 2020 by the authors. Licensee MDPI, Basel, Switzerland.
Motivation & Objective
- To develop a compact, high-precision inertial sensor for multi-degree-of-freedom displacement sensing in gravitational physics experiments.
- To reduce the size, weight, and complexity of multi-channel interferometric systems for space-based and ground-based applications.
- To enable sub-picometer resolution by integrating a reference interferometer within the same optical element as the measurement interferometer.
- To suppress laser frequency noise through self-referencing, eliminating the need for external reference interferometers.
- To demonstrate feasibility of sub-picometer precision down to 2 mHz using a dual-SEDI configuration.
Proposed method
- The SEDI sensor employs deep frequency modulation (DFM) interferometry, where a strong frequency modulation is applied to the laser input beam in an unequal-arm interferometer.
- The output signal contains harmonic components at multiples of the modulation frequency, whose complex amplitudes are used to extract the phase in real time or via post-processing using a Levenberg-Marquardt fitting routine.
- A single optical element hosts both the measurement and reference interferometers, enabling self-referencing and intrinsic laser frequency noise cancellation.
- The reference interferometer's signal is scaled by the ratio of effective modulation indices and subtracted from the measurement signal to suppress laser noise.
- Structural and thermal finite element analysis (FEA) are used to assess mechanical stability and feasibility of the compact design.
- Noise budget analysis includes thermal, thermal noise correlation, and coherence effects between the reference and measurement interferometers, with results simulated using IfoCAD and Gaussian beam propagation models.
Experimental results
Research questions
- RQ1Can a single optical element host both a measurement and reference interferometer to enable self-referencing and laser noise suppression?
- RQ2What is the achievable displacement sensitivity of the SEDI sensor across a wide dynamic range, especially at low frequencies?
- RQ3How does correlated thermal noise between the reference and measurement interferometers affect overall performance?
- RQ4Can the SEDI design achieve sub-picometer precision in a few-cubic-inch package suitable for space missions?
- RQ5What is the performance limit when two SEDI sensors are combined in a dual-configuration?
Key findings
- The SEDI sensor achieves sub-picometer displacement precision for frequencies above 10 mHz in a compact package of a few cubic inches.
- With a dual-SEDI configuration, sub-picometer precision is extended down to 2 mHz, enabling high-sensitivity low-frequency inertial sensing.
- Thermal noise correlation between the reference and measurement interferometers has a significant impact on performance, with coherence functions showing measurable effects at cut-off frequencies between 10−3 Hz and 10−1 Hz.
- Finite element analysis confirms the mechanical feasibility of the SEDI design under thermal and vibrational loads, supporting its use in space and ground-based environments.
- The self-referencing approach eliminates the need for external reference interferometers, drastically reducing system complexity and size.
- Simulation results using IfoCAD and Gaussian beam models validate the noise budget and phase extraction accuracy, confirming the feasibility of real-time phase retrieval via harmonic amplitude fitting.
Read this paper
Open the full text in the Nubint reader and ask about any passage as you read.
Read this paper on NubintThis review was generated by AI from the original paper. Please check the paper before citing it.