[Paper Review] Using artificial intelligence to detect chest X-rays with no significant findings in a primary health care setting in Oulu, Finland
This study evaluates a commercial AI system for detecting normal (no significant findings) chest X-rays in a Finnish primary care setting. Using retrospective analysis of 9,579 cases, the AI achieved 99.8% sensitivity and 36.4% specificity in identifying normal studies, with only nine false negatives—none involving critical findings—demonstrating high safety and potential for reducing radiologist workload in primary care imaging.
Objectives: To assess the use of artificial intelligence-based software in ruling out chest X-ray cases, with no significant findings in a primary health care setting. Methods: In this retrospective study, a commercially available artificial intelligence (AI) software was used to analyse 10 000 chest X-rays of Finnish primary health care patients. In studies with a mismatch between an AI normal report and the original radiologist report, a consensus read by two board-certified radiologists was conducted to make the final diagnosis. Results: After the exclusion of cases not meeting the study criteria, 9579 cases were analysed by AI. Of these cases, 4451 were considered normal in the original radiologist report and 4644 after the consensus reading. The number of cases correctly found nonsignificant by AI was 1692 (17.7% of all studies and 36.4% of studies with no significant findings). After the consensus read, there were nine confirmed false-negative studies. These studies included four cases of slightly enlarged heart size, four cases of slightly increased pulmonary opacification and one case with a small unilateral pleural effusion. This gives the AI a sensitivity of 99.8% (95% CI= 99.65-99.92) and specificity of 36.4 % (95% CI= 35.05-37.84) for recognising significant pathology on a chest X-ray. Conclusions: AI was able to correctly rule out 36.4% of chest X-rays with no significant findings of primary health care patients, with a minimal number of false negatives that would lead to effectively no compromise on patient safety. No critical findings were missed by the software.
Motivation & Objective
- To assess the performance of an AI-based software in identifying chest X-rays with no significant findings in a primary health care setting.
- To evaluate the safety and accuracy of AI in ruling out normal cases, minimizing the risk of missing critical pathology.
- To compare AI results against original radiologist readings and a consensus read by two board-certified radiologists to determine diagnostic reliability.
- To quantify the sensitivity and specificity of the AI system in detecting normal radiographic findings in a real-world primary care population.
- To identify any false-negative cases and assess whether critical findings were missed by the AI system.
Proposed method
- Retrospective analysis of 10,000 chest X-rays from Finnish primary care patients in Oulu, Finland.
- Use of a commercially available AI software to classify each X-ray as normal or abnormal based on predefined algorithms.
- Cases with discrepancies between AI output and original radiologist reports underwent a consensus reading by two board-certified radiologists to determine the final diagnosis.
- Statistical analysis of AI performance using sensitivity, specificity, and 95% confidence intervals.
- Exclusion of cases not meeting study criteria (e.g., incomplete data, non-diagnostic studies) to ensure data quality.
- Final evaluation focused on the AI’s ability to correctly identify normal studies and detect false negatives.
Experimental results
Research questions
- RQ1What is the sensitivity and specificity of the AI system in detecting chest X-rays with no significant findings in a primary care setting?
- RQ2How many false-negative cases does the AI system produce, and what types of pathology are missed?
- RQ3Are any critical or life-threatening findings missed by the AI system in the absence of significant radiological findings?
- RQ4How does the AI’s performance compare between original radiologist reports and a consensus read by two expert radiologists?
- RQ5To what extent can the AI system safely rule out normal cases in a primary care environment without compromising patient safety?
Key findings
- The AI system correctly identified 1,692 out of 4,644 cases with no significant findings after consensus reading, resulting in a sensitivity of 99.8% (95% CI: 99.65–99.92).
- Specificity for identifying normal cases was 36.4% (95% CI: 35.05–37.84), indicating that the AI flagged a relatively low proportion of truly normal cases as normal.
- Only nine false-negative cases were identified after consensus reading, including four with slightly enlarged heart size, four with slightly increased pulmonary opacification, and one with a small unilateral pleural effusion.
- No critical or life-threatening findings were missed by the AI, indicating strong safety performance in clinical screening.
- The AI system demonstrated high reliability in ruling out normal cases, with minimal risk of missing significant pathology.
- The results suggest that AI can be safely used in primary care to pre-screen chest X-rays, reducing the workload for radiologists while maintaining high diagnostic safety.