TL;DR

This study compares quartz tuning fork and length extensional resonator force sensors in atomic force microscopy, analyzing four noise sources—deflection detector, thermal, oscillator, and thermal drift noise—across stiffness, bandwidth, and amplitude. It confirms that optimal signal-to-noise ratio is achieved at oscillation amplitudes comparable to the force range, and identifies strategies to further improve sensor performance by minimizing stiffness-dependent noise sources.

ABSTRACT

The force sensor is key to the performance of atomic force microscopy (AFM). Nowadays, most AFMs use micro-machined force sensors made from silicon, but piezoelectric quartz sensors are applied at an increasing rate, mainly in vacuum. These self sensing force sensors allow a relatively easy upgrade of a scanning tunneling microscope to a combined scanning tunneling/atomic force microscope. Two fundamentally different types of quartz sensors have achieved atomic resolution: the 'needle sensor' that is based on a length extensional resonator and the 'qPlus sensor' that is based on a tuning fork. Here, we calculate and measure the noise characteristics of these sensors. We find four noise sources: deflection detector noise, thermal noise, oscillator noise and thermal drift noise. We calculate the effect of these noise sources as a factor of sensor stiffness, bandwidth and oscillation amplitude. We find that for self sensing quartz sensors, the deflection detector noise is independent of sensor stiffness, while the remaining three noise sources increase strongly with sensor stiffness. Deflection detector noise increases with bandwidth to the power of 1.5, while thermal noise and oscillator noise are proportional to the square root of the bandwidth. Thermal drift noise, however, is inversely proportional to bandwidth. The first three noise sources are inversely proportional to amplitude while thermal drift noise is independent of the amplitude. Thus, we show that the earlier finding that quoted optimal signal-to-noise ratio for oscillation amplitudes similar to the range of the forces is still correct when considering all four frequency noise contributions. Finally, we suggest how the signal-to-noise ratio of the sensors can be further improved and briefly discuss the challenges of mounting tips.

Motivation & Objective

  • To evaluate and compare the noise characteristics of two self-sensing quartz force sensors used in atomic force microscopy: the qPlus sensor (tuning fork-based) and the needle sensor (length extensional resonator-based).
  • To identify and quantify the contributions of four distinct noise sources—deflection detector noise, thermal noise, oscillator noise, and thermal drift noise—to sensor performance.
  • To determine the dependence of each noise source on sensor stiffness, oscillation bandwidth, and amplitude to establish optimal operating conditions.
  • To provide design guidelines for improving the signal-to-noise ratio in self-sensing quartz AFM sensors through parameter optimization.

Proposed method

  • Theoretical modeling of noise contributions using physical principles, including thermal noise from Brownian motion and oscillator phase noise from electronic feedback.
  • Derivation of analytical expressions for each noise source as a function of sensor stiffness, bandwidth, and oscillation amplitude.
  • Experimental measurement of noise characteristics in both tuning fork and length extensional resonator sensors under controlled conditions.
  • Systematic variation of sensor stiffness, bandwidth, and amplitude to isolate the influence of each parameter on signal-to-noise ratio.
  • Use of the signal-to-noise ratio as the primary metric to evaluate sensor performance across different configurations.
  • Comparison of theoretical predictions with experimental data to validate the noise model and identify dominant noise sources.

Experimental results

Research questions

  • RQ1How do deflection detector noise, thermal noise, oscillator noise, and thermal drift noise scale with sensor stiffness, bandwidth, and oscillation amplitude in quartz-based AFM sensors?
  • RQ2What is the optimal oscillation amplitude for maximizing signal-to-noise ratio when all four noise sources are considered?
  • RQ3How does sensor stiffness affect the relative contribution of each noise source to overall sensor performance?
  • RQ4Can the signal-to-noise ratio be further improved by optimizing sensor parameters such as bandwidth and amplitude?
  • RQ5What are the practical challenges in mounting tips on these self-sensing quartz sensors without degrading noise performance?

Key findings

  • Deflection detector noise is independent of sensor stiffness but increases with bandwidth to the power of 1.5.
  • Thermal noise and oscillator noise are proportional to the square root of the bandwidth and increase with stiffness.
  • Thermal drift noise is inversely proportional to bandwidth and is independent of oscillation amplitude.
  • Thermal drift noise is unaffected by oscillation amplitude, while the other three noise sources decrease with increasing amplitude.
  • The earlier conclusion that optimal signal-to-noise ratio occurs at amplitudes comparable to the force range remains valid when all four noise sources are considered.
  • Improving signal-to-noise ratio requires minimizing stiffness-dependent noise sources and optimizing bandwidth and amplitude to balance competing noise contributions.

Read this paper