TL;DR

An AI system integrating FFDM, synthetic mammography (C-View), and DBT provides breast-level cancer predictions and bounding-box lesion localization; achieves internal AUROC of 0.945 and reduces recalls and radiologist workload, with strong external generalizability and prospective deployment results.

ABSTRACT

Although digital breast tomosynthesis (DBT) improves diagnostic performance over full-field digital mammography (FFDM), false-positive recalls remain a concern in breast cancer screening. We developed a multi-modal artificial intelligence system integrating FFDM, synthetic mammography, and DBT to provide breast-level predictions and bounding-box localizations of suspicious findings. Our AI system, trained on approximately 500,000 mammography exams, achieved 0.945 AUROC on an internal test set. It demonstrated capacity to reduce recalls by 31.7% and radiologist workload by 43.8% while maintaining 100% sensitivity, underscoring its potential to improve clinical workflows. External validation confirmed strong generalizability, reducing the gap to a perfect AUROC by 35.31%-69.14% relative to strong baselines. In prospective deployment across 18 sites, the system reduced recall rates for low-risk cases. An improved version, trained on over 750,000 exams with additional labels, further reduced the gap by 18.86%-56.62% across large external datasets. Overall, these results underscore the importance of utilizing all available imaging modalities, demonstrate the potential for clinical impact, and indicate feasibility of further reduction of the test error with increased training set when using large-capacity neural networks.

Motivation & Objective

  • Motivate reducing false-positive recalls and radiologist workload in US screening using AI.
  • Develop a multi-modal AI system that leverages FFDM, C-View, and DBT with bounding-box localization.
  • Train on pathology-confirmed cancer labels and bounding-box annotations to improve interpretability.
  • Assess generalizability across external datasets and in prospective clinical deployment across multiple sites.

Proposed method

  • Use a modified YOLOX architecture to generate bounding-box predictions and image-level cancer probabilities.
  • Process all DBT slices to preserve depth information and ensemble modality-specific predictions for breast-level output.
  • Train with dual signals: pathology-confirmed cancer labels and bounding-box annotations to guide learning.
  • Ensemble predictions across FFDM, C-View, and DBT using breast-level averaging and multi-modal bounding-box fusion.
  • Evaluate on internal test sets (AUROC 0.945; 95% CI 0.930–0.960) and seven external datasets; assess retrospective recall-reduction and prospective clinical impact.

Experimental results

Research questions

  • RQ1Can a three-modality mammography AI system (FFDM, C-View, DBT) improve breast cancer detection over single-modality approaches?
  • RQ2What is the impact of multi-modal ensembling on breast-level AUROC and bounding-box localization performance across diverse external datasets?
  • RQ3Does prospective AI-assisted interpretation reduce recall rates and radiologist workload in US screening workflows?
  • RQ4How does increasing training data size affect performance and generalizability across large external datasets?

Key findings

  • Internal AUROC of 0.945 (CI 0.930–0.960) on 38,368 breasts with V1, and 0.953 (CI 0.938–0.967) on V2 test set.
  • bounding-box AUFROC1 for malignant lesions: 0.848 (V1) and 0.945 (V2) on internal test sets.
  • External datasets show strong generalizability; V2 models surpass V1 and exceed baselines like GMIC and 3D-GMIC on several datasets.
  • Prospective deployment across 18 sites reduced recall rates for low-risk (green) exams and demonstrated AI-assisted adaptation in real-world workflow.
  • An improved model trained on 768,493 exams with additional labels further reduced performance gaps by 18.86%–56.62% on large external datasets.

Read this paper