[Paper Review] A multicenter study on radiomic features from T$_2$-weighted images of a customized MR pelvic phantom setting the basis for robust radiomic models in clinics
This multicenter study evaluates the repeatability and reproducibility of radiomic features from T2-weighted MRI using a customized pelvic phantom across multiple scanners and protocols. It identifies robust features with ICC > 0.9 and CCC > 0.9, establishes a workflow to filter out protocol-sensitive features, and demonstrates that only 4.6–15.6% of features are reliably reproducible across scanners, with shape-informative features constituting just 2.0% of non-shape features.
In this study we investigated the repeatability and reproducibility of radiomic features extracted from MRI images and provide a workflow to identify robust features. 2D and 3D T$_2$-weighted images of a pelvic phantom were acquired on three scanners of two manufacturers and two magnetic field strengths. The repeatability and reproducibility of the radiomic features were assessed respectively by intraclass correlation coefficient (ICC) and concordance correlation coefficient (CCC), considering repeated acquisitions with or without phantom repositioning, and with different scanner/acquisition type, and acquisition parameters. The features showing ICC/CCC > 0.9 were selected, and their dependence on shape information (Spearman's $ρ$> 0.8) was analyzed. They were classified for their ability to distinguish textures, after shuffling voxel intensities. From 944 2D features, 79.9% to 96.4% showed excellent repeatability in fixed position across all scanners. Much lower range (11.2% to 85.4%) was obtained after phantom repositioning. 3D extraction did not improve repeatability performance. Excellent reproducibility between scanners was observed in 4.6% to 15.6% of the features, at fixed imaging parameters. 82.4% to 94.9% of features showed excellent agreement when extracted from images acquired with TEs 5 ms apart (values decreased when increasing TE intervals) and 90.7% of the features exhibited excellent reproducibility for changes in TR. 2.0% of non-shape features were identified as providing only shape information. This study demonstrates that radiomic features are affected by specific MRI protocols. The use of our radiomic pelvic phantom allowed to identify unreliable features for radiomic analysis on T$_2$-weighted images. This paper proposes a general workflow to identify repeatable, reproducible, and informative radiomic features, fundamental to ensure robustness of clinical studies.
Motivation & Objective
- To assess the repeatability and reproducibility of radiomic features across multiple MRI scanners and protocols.
- To identify radiomic features that are robust to variations in acquisition parameters, scanner type, and phantom positioning.
- To develop a standardized workflow for selecting reliable, reproducible, and informative radiomic features for clinical applications.
- To evaluate the influence of sequence parameters (TR, TE) and shape information on feature stability.
- To establish a foundation for robust radiomic modeling in clinical oncology by filtering out unreliable features.
Proposed method
- Acquired 2D and 3D T2-weighted MRI images from a customized pelvic phantom on three scanners from two manufacturers at two field strengths.
- Performed repeated acquisitions with and without phantom repositioning to assess repeatability using intraclass correlation coefficient (ICC).
- Evaluated reproducibility across scanners using concordance correlation coefficient (CCC) under varying acquisition parameters.
- Analyzed dependence of features on shape using Spearman's rho (ρ > 0.8) to distinguish shape-informative from texture-informative features.
- Assessed feature ability to distinguish textures by shuffling voxel intensities to isolate non-shape content.
- Proposed a systematic workflow to select features with high ICC (>0.9), CCC (>0.9), and low shape dependency for clinical use.
Experimental results
Research questions
- RQ1Which radiomic features exhibit high repeatability across repeated scans with fixed phantom positioning?
- RQ2How reproducible are radiomic features across different MRI scanners and acquisition protocols?
- RQ3To what extent do sequence parameters (TR, TE) affect the stability of radiomic features?
- RQ4Which features are primarily driven by shape rather than texture, and how can they be filtered?
- RQ5What is the proportion of features that are both reproducible and informative for tissue texture characterization?
Key findings
- Between 79.9% and 96.4% of 2D radiomic features showed excellent repeatability (ICC > 0.9) when the phantom was fixed in position across all scanners.
- After phantom repositioning, repeatability dropped significantly, ranging from 11.2% to 85.4% of features with ICC > 0.9.
- 3D feature extraction did not improve repeatability compared to 2D features.
- Only 4.6% to 15.6% of features demonstrated excellent reproducibility (CCC > 0.9) across scanners at fixed parameters.
- 90.7% of features showed excellent reproducibility for TR variations, while performance decreased with increasing TE intervals.
- Just 2.0% of non-shape radiomic features were found to provide only shape information, indicating high specificity of most features to texture.