Skip to main content
QUICK REVIEW

[Paper Review] Admire vs. Safire: Objective comparison of CT reconstruction algorithms and their noise properties

Ingvild Dalehaug, Kirsten Nygaard Bolstad|arXiv (Cornell University)|Aug 31, 2017
Radiation Dose and ImagingMedicine13 references6 citations
TL;DR

This study objectively compares the noise reduction performance of Siemens' ADMIRE and SAFIRE iterative reconstruction algorithms in CT imaging using noise power spectrum (NPS) analysis and inter-image standard deviation maps. Results show ADMIRE reduces noise near edges more effectively than SAFIRE, but no significant improvement in noise texture preservation was observed, with ADMIRE producing more blotchy images at low dose due to a shift toward lower spatial frequencies.

ABSTRACT

Purpose: Siemens has developed several iterative reconstruction (IR) algorithms on their CT scanners. SAFIRE is available on most of their CT scanners. The latest algorithm, ADMIRE, is available on their newest high-end CT scanners. The aim of our study was to compare the noise reduction properties of the two IR algorithms using objective methods. Methods and Materials: The homogeneous module of the Catphan phantom was scanned on a Siemens AS+ and a Siemens Flash CT scanner using an axial abdomen protocol with fixed tube current at two dose levels. The images were reconstructed with an abdomen filter (B30) using filtered back projection (FBP) and a low, medium, and high level of SAFIRE or ADMIRE. Noise Power Spectrum (NPS) curves were calculated using these images. Then, an anthropomorphic abdomen phantom (Kyoto Kagaku PH-5) was scanned using the same setup and exposure parameters. Fifty axial images at the same slice location were used to calculate inter-image standard deviation maps. Results: At full dose, the median values of the NPS curves were similar for both scanners at all IR levels. At low dose the median values of the NPS curves were generally shifted towards lower spatial frequencies, usually resulting in a more blotchy image texture. This shift was more prominent for ADMIRE compared to SAFIRE for all IR levels. Based on the inter-image standard deviation maps of the anthropomorphic phantom, ADMIRE removed noise near edges more efficiently than SAFIRE. Conclusion: No significant improvement in maintaining noise structure were found for the ADMIRE algorithm. Based on the inter-image standard deviation maps, ADMIRE removed noise near edges more efficiently than SAFIRE.

Motivation & Objective

  • To objectively evaluate and compare the noise reduction properties of two iterative reconstruction algorithms, ADMIRE and SAFIRE, in CT imaging.
  • To assess how each algorithm affects image noise texture and spatial frequency distribution using quantitative metrics.
  • To investigate whether ADMIRE offers superior noise structure preservation compared to SAFIRE, particularly at low-dose protocols.
  • To evaluate the performance of both algorithms using both homogeneous and anthropomorphic phantoms under identical scanning conditions.

Proposed method

  • Scanning of a Catphan phantom's homogeneous module on two Siemens CT scanners (AS+ and Flash) using axial abdomen protocols at two dose levels.
  • Reconstruction of images using filtered back projection (FBP) and three levels (low, medium, high) of SAFIRE and ADMIRE with an abdomen-specific kernel (B30).
  • Computation of Noise Power Spectrum (NPS) curves from reconstructed images to analyze spatial frequency distribution of noise.
  • Scanning of an anthropomorphic abdomen phantom (Kyoto Kagaku PH-5) to assess inter-scan noise variability.
  • Calculation of inter-image standard deviation maps from 50 axial slices at the same anatomical location to quantify noise variation across repeated scans.
  • Statistical comparison of NPS median values and standard deviation maps between ADMIRE and SAFIRE at different dose levels.

Experimental results

Research questions

  • RQ1How do ADMIRE and SAFIRE compare in terms of noise power spectrum (NPS) median values at full and low dose?
  • RQ2Does ADMIRE preserve noise texture better than SAFIRE, particularly in terms of spatial frequency distribution?
  • RQ3To what extent does ADMIRE reduce noise near anatomical edges compared to SAFIRE?
  • RQ4How does the shift in NPS curves toward lower spatial frequencies differ between ADMIRE and SAFIRE at low dose?
  • RQ5Is there a measurable improvement in inter-scan noise consistency with ADMIRE compared to SAFIRE?

Key findings

  • At full dose, the median NPS values for ADMIRE and SAFIRE were similar across all IR levels on both scanners.
  • At low dose, both algorithms shifted NPS curves toward lower spatial frequencies, but this shift was more pronounced with ADMIRE than with SAFIRE.
  • The shift in NPS toward lower spatial frequencies with ADMIRE resulted in a more blotchy image texture compared to SAFIRE.
  • Inter-image standard deviation maps showed that ADMIRE reduced noise more effectively near edges than SAFIRE.
  • Despite superior edge noise reduction, ADMIRE did not demonstrate significant improvement in preserving noise structure compared to SAFIRE.
  • The study found no substantial benefit in noise texture preservation with ADMIRE, suggesting SAFIRE maintains a more natural noise appearance.

Read this paper

Open the full text in the Nubint reader and ask about any passage as you read.

Read this paper on Nubint

This review was generated by AI from the original paper. Please check the paper before citing it.