[Paper Review] Afterpulsing in Silicon Photomultipliers: Impact on the Photodetectors Characterization
This paper models afterpulsing in silicon photomultipliers (SiPMs) using a statistical framework based on Poisson-distributed afterpulses with exponential decay, demonstrating that afterpulsing significantly degrades key performance metrics such as dark count rate, pulse height resolution, and timing characteristics. The key contribution is a quantitative framework showing that afterpulsing probability $P_{\text{aft}}$ increases with overvoltage, leading to a nonlinear rise in excess noise factor and systematic errors in dark count and photon calibration measurements.
Novel generation of silicon-based photodetectors are attractive alternatives to the traditional phototubes. They offer significant advantages but they present new challenges too. Presence of afterpulses may affect many characteristics of the photodetectors. Simple statistical model of afterpulsing is used to evaluate the contribution to the observed dark count rates, to examine the contribution to the pulse height resolution and to demonstrate the modification of the observed timing properties of the SiPMs.
Motivation & Objective
- To understand how afterpulsing in SiPMs distorts key detector characteristics such as dark count rate, pulse height resolution, and timing properties.
- To model the statistical behavior of afterpulsing using Poisson-distributed secondary pulses with exponential decay time constants.
- To quantify the impact of afterpulsing on the accuracy of dark count rate measurements and absolute photon calibration in SiPMs.
- To provide a correction framework for experimental data based on effective afterpulsing parameters ($P_{\text{aft}}$, $\tau_{\text{aft}}$) to retrieve true underlying detector parameters.
Proposed method
- A simplified statistical model treats afterpulsing as a Poisson process with mean $P_{\text{aft}}$, representing the average number of afterpulses per primary pulse.
- Afterpulses are modeled as exponentially decaying in time with a characteristic time constant $\tau_{\text{aft}}$, simulating trap release dynamics.
- Monte Carlo integration is used to simulate 50 million initial afterpulses, enabling convolution of Poisson and exponential distributions to generate pulse train statistics.
- The model accounts for iterative afterpulsing: each afterpulse can itself trigger further afterpulses, forming chains of events.
- The excess noise factor (ENF) is calculated as $\text{ENF} = \sigma^2 / \text{mean}$ to quantify degradation in pulse height resolution due to afterpulsing fluctuations.
- Experimental corrections are derived by comparing observed pulse distributions (e.g., $P(0)$) to true thermal rates, using gate length and $P_{\text{aft}}$ as variables.
Experimental results
Research questions
- RQ1How does afterpulsing affect the measured dark count rate in SiPMs, and what is its dependence on bias voltage?
- RQ2To what extent does afterpulsing degrade the pulse height resolution of SiPMs, as quantified by the excess noise factor (ENF)?
- RQ3How does afterpulsing distort the observed timing distribution of pulses, and what is the impact on time-of-flight measurements?
- RQ4How do standard calibration methods based on pulse height distribution width fail when afterpulsing is present?
- RQ5What corrections are required to recover the true thermal dark count rate and photon number from measurements affected by afterpulsing?
Key findings
- The excess noise factor (ENF) increases nonlinearly with afterpulsing probability $P_{\text{aft}}$, growing much faster than linearly, which degrades pulse height resolution significantly.
- The effective gain due to afterpulsing is $\frac{P_{\text{aft}}}{1 - P_{\text{aft}}}$, leading to a substantial increase in observed dark count rate with rising overvoltage.
- For gate lengths longer than $1\mu\text{s} \approx 20\tau_{\text{aft}}$, the true thermal dark count rate can be recovered with better than 20% accuracy, even at high $P_{\text{aft}}$.
- Photon number estimation using the standard method $\langle N \rangle = \text{Mean}^2 / \text{RMS}^2$ underestimates the true number by a factor of $(1 - P_{\text{aft}})$, leading to systematic calibration errors.
- The observed pulse rate is biased by afterpulsing, and the ratio of estimated to true thermal rate depends strongly on both gate length and $P_{\text{aft}}$, as shown in Figures 8 and 9.
- Despite simplifying assumptions, the effective $P_{\text{aft}}$ and $\tau_{\text{aft}}$ derived from the model provide a robust basis for correcting real SiPM data across different operating conditions.
Read this paper