[Paper Review] The Upgrade of VERITAS with High Efficiency Photomultipliers

A. N. Otte|arXiv (Cornell University)|Oct 21, 2011
Radiation Detection and Scintillator TechnologiesPhysics and Astronomy1 references17 citations
TL;DR

This paper details the upgrade of the VERITAS Cherenkov telescope array with high-efficiency R10560-100-20 MOD photomultipliers (PMTs), which reduce the energy threshold from ~100 GeV to ~70 GeV and improve sensitivity by increasing photon detection efficiency by 35%. The new PMTs feature a narrower pulse width (4.2 ns vs. 6.8 ns), superior quantum efficiency, and better signal discrimination, enabling enhanced performance in moonlight and low-energy gamma-ray observations.

ABSTRACT

We are in the process of upgrading the VERITAS array of Cherenkov telescopes with new, high efficiency photomultipliers (PMT) that will considerably lower the energy threshold of the instrument and improve the overall sensitivity. The upgrade will be finished in Summer 2012 when the PMTs will be installed in the existing cameras. We discuss the performance of the new photon detectors and the status of the project.

Motivation & Objective

  • To reduce the energy threshold of the VERITAS Cherenkov telescope array below 100 GeV to enhance sensitivity to faint and low-energy gamma-ray sources.
  • To improve the overall sensitivity of VERITAS by replacing legacy PMTs with higher-efficiency photodetectors.
  • To enable better discrimination of Cherenkov signals from night sky background (NSB) photons through narrower pulse shapes.
  • To maintain or improve system performance while upgrading hardware components without fundamental physical constraints.
  • To ensure long-term reliability and accurate calibration through characterization of PMT aging and afterpulsing effects.

Proposed method

  • Replacement of existing Photonis XP2970 PMTs with Hamamatsu R10560-100-20 MOD PMTs in all four VERITAS telescopes.
  • Use of a linear focused dynode structure with eight stages and active voltage stabilization to maintain a nominal gain of 2×10⁵ at 1100 V bias.
  • Implementation of a new pixel unit consisting of a Delrin tube, R10560 PMT, and preamplifier, designed as a modular, exchangeable component.
  • Conducting detailed afterpulsing and efficiency measurements on 10% of PMTs to enable accurate Monte Carlo simulations and system calibration.
  • Measuring single photoelectron pulse shapes and amplitude distributions to assess signal resolution and pedestal separation.
  • Folding measured photon detection efficiency (PDE) with the Cherenkov light spectrum to estimate effective light collection performance.

Experimental results

Research questions

  • RQ1What is the impact of replacing legacy PMTs with high-efficiency R10560 PMTs on VERITAS's energy threshold and sensitivity?
  • RQ2How do the pulse shape characteristics of the R10560 compare to those of the XP2970, and what is the effect on background discrimination?
  • RQ3What is the photon detection efficiency (PDE) of the R10560, and how does it compare to the XP2970 in the VERITAS energy range?
  • RQ4How does PMT aging affect gain stability over time, and what is the expected lifetime performance under VERITAS operating conditions?
  • RQ5What is the afterpulsing probability of the R10560, and how does it affect signal reconstruction and background rejection?

Key findings

  • The R10560 PMT achieves a 35% higher photon detection yield compared to the XP2970, resulting in a 35% increase in effective light collection.
  • The single photoelectron pulse of the R10560 has a full width at half maximum (FWHM) of 4.2 ns, which is 40% narrower than the 6.8 ns FWHM of the XP2970.
  • The R10560's quantum efficiency is significantly higher than that of classical bialkali PMTs, with a measured photoelectron collection efficiency of 92%.
  • The energy threshold of VERITAS is expected to be reduced from ~100 GeV to ~70 GeV after the upgrade, improving sensitivity to low-energy gamma-ray sources.
  • Afterpulsing probabilities are low and consistent across measurement methods, with a measured afterpulse rate of approximately 1.5% at 100 ns after the primary pulse.
  • The R10560 exhibits a 25% gain drop over 700 hours of operation at 45 µA anode current, corresponding to roughly four years of VERITAS operation, indicating manageable aging effects.

Read this paper

Open the full text in the Nubint reader and ask about any passage as you read.

Read this paper on Nubint

This review was generated by AI from the original paper. Please check the paper before citing it.