[Paper Review] Automating LC-MS/MS mass chromatogram quantification. Wavelet transform based peak detection and automated estimation of peak boundaries and signal-to-noise ratio using signal processing methods

Florian Rupprecht, Sören Enge|arXiv (Cornell University)|Jan 1, 2021
Metabolomics and Mass Spectrometry StudiesBiochemistry, Genetics and Molecular Biology22 references13 citations
TL;DR

This paper presents a wavelet transform-based algorithm for automated peak detection, boundary estimation, and signal-to-noise ratio (SNR) calculation in LC-MS/MS chromatograms. By applying continuous wavelet transformation and digital signal processing, the method achieves high correlation with expert manual quantification, reduces non-detectable results, and enables reliable, objective, and reproducible quantification of six steroid hormones in hair samples.

ABSTRACT

While there are many different methods for peak detection, no automatic methods for marking peak boundaries to calculate area under the curve (AUC) and signal-to-noise ratio (SNR) estimation exist. An algorithm for the automation of liquid chromatography tandem mass spectrometry (LC-MS/MS) mass chromatogram quantification was developed and validated. Continuous wavelet transformation and other digital signal processing methods were used in a multi-step procedure to calculate concentrations of six different analytes. To evaluate the performance of the algorithm, the results of the manual quantification of 446 hair samples with 6 different steroid hormones by two experts were compared to the algorithm results. The proposed approach of automating mass chromatogram quantification is reliable and valid. The algorithm returns less nondetectables than human raters. Based on signal to noise ratio, human non-detectables could be correctly classified with a diagnostic performance of AUC = 0.95. The algorithm presented here allows fast, automated, reliable, and valid computational peak detection and quantification in LC- MS/MS.

Motivation & Objective

  • To develop an open-source, automated method for LC-MS/MS peak quantification that eliminates manual intervention in peak boundary marking and SNR estimation.
  • To improve productivity and objectivity in LC-MS/MS data analysis by replacing time-consuming manual peak integration with algorithmic processing.
  • To validate the algorithm’s performance against expert human raters in real-world hair sample analysis for six steroid hormones.
  • To assess the algorithm’s reliability, consistency, and validity in measuring analyte concentrations compared to manual expert quantification.
  • To evaluate the diagnostic performance of SNR and peak area under the curve (AUC) in classifying non-detectable results.

Proposed method

  • The algorithm uses a continuous wavelet transform (CWT) to detect peak features in LC-MS/MS chromatograms by analyzing time-series intensity data.
  • A multi-step signal processing pipeline includes Gaussian smoothing, high-pass filtering, second derivative approximation, and CWT to decompose chromatographic signals into background, peak, and noise components.
  • Peak boundaries are estimated using a partial convex hull method that mimics expert rater behavior, improving accuracy in area under the curve (AUC) calculation.
  • Signal-to-noise ratio (SNR) is computed based on peak height and background noise, enabling automated detection of low-intensity or non-detectable signals.
  • The method accounts for retention time shifts and drift by modeling background as a linear trend and applying calibration to reference standards.
  • Latent variable modeling (state-trait theory) is used to decompose variance into reliability, specificity, and consistency components for method comparison.
Figure 1 : Generic model for chromatographic peaks. The actual LC–MS/MS sensor data ( $y(t)$ ) is combined background and drift ( $B(t)$ ), peak distribution ( $P(t)$ ) and noise ( $N(t)$ ). Note that this Figure does not include peak skew or nearby peaks. Peak range is the time interval between the
Figure 1 : Generic model for chromatographic peaks. The actual LC–MS/MS sensor data ( $y(t)$ ) is combined background and drift ( $B(t)$ ), peak distribution ( $P(t)$ ) and noise ( $N(t)$ ). Note that this Figure does not include peak skew or nearby peaks. Peak range is the time interval between the

Experimental results

Research questions

  • RQ1Does the automated algorithm produce analyte concentrations that correlate strongly with those manually quantified by expert raters?
  • RQ2Can the algorithm reduce the number of non-detectable results compared to human raters while maintaining accuracy?
  • RQ3Is the signal-to-noise ratio (SNR) and peak AUC effective in classifying non-detectable results with high diagnostic performance?
  • RQ4How does the algorithm’s reliability and consistency compare to those of expert human raters across multiple analytes?
  • RQ5To what extent does the algorithm measure the same underlying biological construct (hormone concentration) as human raters, as validated by latent variable modeling?

Key findings

  • The algorithm demonstrated strong correlation with expert manual quantification, with reliability values ranging from 0.776 to 0.986 across analytes.
  • The algorithm produced significantly fewer non-detectable results than human raters, indicating improved sensitivity in low-abundance signal detection.
  • Signal-to-noise ratio (SNR) and peak AUC were highly effective in classifying non-detectables, achieving an AUC of 0.95 in diagnostic performance.
  • Latent variable modeling confirmed that the algorithm measures the same underlying construct (hormone concentration) as human raters, with high consistency and reliability.
  • The model showed good fit for cortisol, cortisone, and DHEA (RMSEA ≤ 0.059, SRMR ≤ 0.019), but poor fit for testosterone and progesterone due to method-specific and replicate-related residual covariances.
  • The algorithm’s consistency (0.801–0.869) and reliability (0.959–0.986) were comparable or superior to those of expert raters, especially for cortisol and DHEA.
Figure 2 : Chromatogram time series transformations. The LC–MS/MS chromatographic time series $Y(t)$ is smoothed using a gaussian kernel. High-pass filtered data is obtained by differentiating smoothed and original time series. The second derivative of the smoothed data is approximated and the conti
Figure 2 : Chromatogram time series transformations. The LC–MS/MS chromatographic time series $Y(t)$ is smoothed using a gaussian kernel. High-pass filtered data is obtained by differentiating smoothed and original time series. The second derivative of the smoothed data is approximated and the conti