[Paper Review] Characterization of a CdZnTe detector for a low-power CubeSat application

Giulio Lucchetta, M. Ackermann|arXiv (Cornell University)|Apr 1, 2022
Advanced Semiconductor Detectors and MaterialsEngineering23 references7 citations
TL;DR

This paper presents a low-power, high-performance pixelated CdZnTe detector system for the MeVCube CubeSat mission, using the VATA450.3 ASIC to achieve 0.25 mW/channel power consumption. It demonstrates 3% FWHM energy resolution at 662 keV and 2 mm 3D spatial resolution, enabling sensitive gamma-ray detection in the 200 keV–4 MeV range for compact space applications.

ABSTRACT

We report spectral and imaging performance of a pixelated CdZnTe detector custom designed for the \emph{MeVCube} project: a small Compton telescope on a CubeSat platform. \emph{MeVCube} is expected to cover the energy range between $200\;\mathrm{keV}$ and $4\;\mathrm{MeV}$, with performance comparable to the last generation of larger satellites. In order to achieve this goal, an energy resolution of few percent in full width at half maximum (FWHM) and a $3$-D spatial resolution of few millimeters for the individual detectors are needed. The severe power constraints present in small satellites require very low power read-out electronics for the detector. Our read-out is based on the VATA450.3 ASIC developed by \emph{Ideas}, with a power consumption of only $0.25\;\mathrm{mW/channel}$, which exhibits good performance in terms of dynamic range, noise and linearity. A $2.0\;\mathrm{cm} imes 2.0\;\mathrm{cm} imes 1.5\;\mathrm{cm}$ CdZnTe detector, with a custom $8 imes 8$ pixel anode structure read-out by a VATA450.3 ASIC, has been tested. A preliminary read-out system for the cathode, based on a discrete \emph{Amptek} A250F charge sensitive pre-amplifier and a DRS4 ASIC, has been implemented. An energy resolution around $3\%$ FWHM has been measured at a gamma energy of $662\;\mathrm{keV}$; at $200\;\mathrm{keV}$ the average energy resolution is $6.5\%$, decreasing to $\lesssim 2\%$ at energies above $1\;\mathrm{MeV}$. A $3$-D spatial resolution of $\approx 2\,\mathrm{mm}$ is achieved.

Motivation & Objective

  • Develop a low-power, high-performance CdZnTe detector system for a CubeSat-based Compton telescope.
  • Achieve energy resolution of a few percent FWHM across 200 keV–4 MeV and 3D spatial resolution of a few millimeters.
  • Integrate ultra-low power readout electronics (0.25 mW/channel) suitable for CubeSat power constraints.
  • Characterize spectral and imaging performance of a 2.0 cm × 2.0 cm × 1.5 cm CdZnTe detector with 8×8 pixel anode and discrete cathode readout.
  • Evaluate the impact of a steering grid on charge collection efficiency and system complexity for space applications.

Proposed method

  • Use a custom 8×8 pixel anode structure on a 2.0 cm × 2.0 cm × 1.5 cm CdZnTe detector with 2.25 mm × 2.25 mm pixels and 2.45 mm pitch.
  • Implement the VATA450.3 ASIC for anode readout, consuming only 0.25 mW/channel, with high dynamic range, low noise, and linearity.
  • Employ a discrete Amptek A250F preamplifier and DRS4 ASIC for cathode signal readout to enable depth-of-interaction measurement.
  • Apply depth-of-interaction correction using cathode-to-pixel signal ratio to improve energy resolution.
  • Conduct steering grid measurements by biasing the grid at −50 V and comparing detection efficiency and spectral response to floating grid conditions.
  • Measure energy resolution via 662 keV Cs-137 gamma-ray peaks and spatial resolution via 3D interaction position reconstruction.

Experimental results

Research questions

  • RQ1Can a CdZnTe detector with 8×8 pixel anode and discrete cathode readout achieve sub-3% FWHM energy resolution at 662 keV under low-power constraints?
  • RQ2What is the 3D spatial resolution of the detector system, and how does it perform across the 200 keV–4 MeV energy range?
  • RQ3How effective is a −50 V-biased steering grid in improving charge collection efficiency without introducing excessive system complexity?
  • RQ4To what extent does depth-of-interaction correction improve energy resolution, particularly at low and high energies?
  • RQ5Can the VATA450.3 ASIC deliver sufficient performance (noise, linearity, dynamic range) for space-based gamma-ray detection with minimal power draw?

Key findings

  • An energy resolution of 3% FWHM was achieved at 662 keV, with resolution improving to ≲2% at energies above 1 MeV.
  • At 200 keV, the average energy resolution was 6.5% FWHM, indicating moderate degradation at the low-energy end.
  • A 3D spatial resolution of approximately 2 mm was achieved in each dimension, meeting the target for Compton telescope reconstruction.
  • The use of a −50 V-biased steering grid resulted in a slight improvement in detection efficiency, but the benefit was deemed insufficient to justify added system complexity.
  • The VATA450.3 ASIC demonstrated excellent performance with 0.25 mW/channel power consumption, low noise, and high linearity across the full dynamic range.
  • The total system noise, measured from pedestal width, was 16.5 keV FWHM (equivalent to 355 e−), with consistent performance across all 63 active pixels.