[Paper Review] Comparison of Patch-Based Conditional Generative Adversarial Neural Net Models with Emphasis on Model Robustness for Use in Head and Neck Cases for MR-Only planning
This study evaluates Pix2Pix and Cycle GAN, two patch-based conditional generative adversarial networks, for synthesizing CT images from MR scans in head and neck cancer patients to enable MR-only radiotherapy planning. Pix2Pix achieved superior accuracy and robustness, with a mean absolute error of 92.4 ± 13.5 HU in the body region and <2% dose error in critical structures, demonstrating strong clinical potential for artifact-prone cases.
A total of twenty paired CT and MR images were used in this study to investigate two conditional generative adversarial networks, Pix2Pix, and Cycle GAN, for generating synthetic CT images for Headand Neck cancer cases. Ten of the patient cases were used for training and included such common artifacts as dental implants; the remaining ten testing cases were used for testing and included a larger range of image features commonly found in clinical head and neck cases. These features included strong metal artifacts from dental implants, one case with a metal implant, and one case with abnormal anatomy. The original CT images were deformably registered to the mDixon FFE MR images to minimize the effects of processing the MR images. The sCT generation accuracy and robustness were evaluated using Mean Absolute Error (MAE) based on the Hounsfield Units (HU) for three regions (whole body, bone, and air within the body), Mean Error (ME) to observe systematic average offset errors in the sCT generation, and dosimetric evaluation of all clinically relevant structures. For the test set the MAE for the Pix2Pix and Cycle GAN models were 92.4 $\pm$ 13.5 HU, and 100.7 $\pm$ 14.6 HU, respectively, for the body region, 166.3 $\pm$ 31.8 HU, and 184 $\pm$ 31.9 HU, respectively, for the bone region, and 183.7 $\pm$ 41.3 HU and 185.4 $\pm$ 37.9 HU for the air regions. The ME for Pix2Pix and Cycle GAN were 21.0 $\pm$ 11.8 HU and 37.5 $\pm$ 14.9 HU, respectively. Absolute Percent Mean/Max Dose Errors were less than 2% for the PTV and all critical structures for both models, and DRRs generated from these models looked qualitatively similar to CT generated DRRs showing these methods are promising for MR-only planning.
Motivation & Objective
- To evaluate the performance of conditional generative adversarial networks in synthesizing CT images from MR scans for MR-only radiotherapy planning.
- To assess model robustness in the presence of challenging clinical artifacts such as dental implants and metal implants.
- To compare the dosimetric accuracy of synthetic CT images generated by Pix2Pix and Cycle GAN against real CT scans.
- To validate the clinical feasibility of synthetic CT generation in complex head and neck anatomy with diverse imaging artifacts.
Proposed method
- Trained two conditional GANs—Pix2Pix and Cycle GAN—on 10 paired CT-MR scans from head and neck cancer patients with dental artifacts.
- Used deformable image registration to align original CT to mDixon FFE MR images, minimizing misregistration effects.
- Employed a patch-based GAN architecture to generate synthetic CT (sCT) images from MR input, focusing on local image consistency.
- Evaluated sCT accuracy using Hounsfield Unit (HU) metrics: mean absolute error (MAE) and mean error (ME) in whole body, bone, and air regions.
- Performed dosimetric evaluation by comparing dose distributions in clinical structures using DRRs and dose differences.
- Used a separate test set of 10 cases with diverse artifacts, including metal implants and abnormal anatomy, to assess robustness.
Experimental results
Research questions
- RQ1How do Pix2Pix and Cycle GAN perform in synthesizing CT images from MR scans in head and neck cancer patients with dental artifacts?
- RQ2What is the robustness of these models when applied to cases with strong metal artifacts or abnormal anatomy?
- RQ3How accurate are the synthetic CT images in terms of HU values across different tissue types (soft tissue, bone, air)?
- RQ4To what extent do the synthetic CT images preserve dosimetric accuracy for clinical structures in radiotherapy planning?
Key findings
- Pix2Pix achieved a mean absolute error (MAE) of 92.4 ± 13.5 HU in the body region, outperforming Cycle GAN’s 100.7 ± 14.6 HU.
- For the bone region, Pix2Pix had an MAE of 166.3 ± 31.8 HU, compared to 184.0 ± 31.9 HU for Cycle GAN.
- In air regions, Pix2Pix recorded an MAE of 183.7 ± 41.3 HU, slightly better than Cycle GAN’s 185.4 ± 37.9 HU.
- The mean error (ME) was 21.0 ± 11.8 HU for Pix2Pix and 37.5 ± 14.9 HU for Cycle GAN, indicating lower systematic bias in Pix2Pix.
- Absolute percent mean and max dose errors were less than 2% for the PTV and all critical structures in both models.
- Qualitative evaluation of DRRs showed that images from both models closely resembled real CT-generated DRRs, supporting clinical usability.
Read this paper