Skip to main content
QUICK REVIEW

[Paper Review] Construction and operation of a Double Phase LAr Large Electron Multiplier Time Projection Chamber

A. Badertscher, L. Knecht|ArXiv.org|Nov 20, 2008
Particle Detector Development and PerformancePhysics and Astronomy8 references16 citations
TL;DR

This paper presents the first successful operation of a double-phase liquid argon Time Projection Chamber (LAr TPC) using a Large Electron Multiplier (LEM) for charge amplification. The prototype, with a 3-liter active volume, demonstrated cosmic ray track reconstruction via dual-readout (LEM and anode strips) in both pure Ar gas and double-phase LAr conditions, proving the feasibility of scalable, cost-effective, high-resolution detectors for neutrino and dark matter experiments.

ABSTRACT

We successfully operated a novel kind of LAr Time Projection Chamber based on a Large Electron Multiplier (LEM) readout system. The prototype, of about 3 liters active volume, is operated in liquid-vapour (double) phase pure Ar. The ionization electrons, after drifting in the LAr volume, are extracted by a set of grids into the gas phase and driven into the holes of a double stage LEM, where charge amplification occurs. Each LEM is a thick macroscopic hole multiplier of 10x10 cm$^2$ manufactured with standard PCB techniques. The electrons signal is readout via two orthogonal coordinates, one using the induced signal on the segmented upper electrode of the LEM itself and the other by collecting the electrons on a segmented anode. Custom-made preamplifiers have been especially developed for this purpose. Cosmic ray tracks have been successfully observed in pure gas at room temperature and in double phase Ar operation. We believe that this proof of principle represents an important milestone in the realization of very large, long drift (cost-effective) LAr detectors for next generation neutrino physics and proton decay experiments, as well as for direct search of Dark Matter with imaging devices.

Motivation & Objective

  • To develop a scalable, cost-effective large-volume liquid argon TPC for next-generation neutrino and dark matter experiments.
  • To demonstrate charge amplification via Large Electron Multipliers (LEMs) in the gas phase above liquid argon for improved signal readout.
  • To achieve high spatial and energy resolution in a double-phase LAr TPC using finely segmented LEMs and custom cryogenic preamplifiers.
  • To validate the feasibility of LEM-based TPCs for long-drift, low-threshold detection in pure argon.
  • To establish a proof-of-principle for imaging ionization tracks with LEM-TPC in double-phase operation.

Proposed method

  • The detector uses a 3-liter active volume of pure liquid argon in a double-phase (liquid-vapor) configuration.
  • Ionization electrons drift through the liquid argon and are extracted into the gas phase via a grid system.
  • Electrons are amplified via a double-stage LEM, a thick, rigid, PCB-fabricated hole multiplier with mm-scale holes.
  • Signal is read out using two orthogonal coordinate systems: induced signals on segmented LEM strips and collected charge on a segmented anode.
  • Custom cryogenic preamplifiers were developed to handle low-noise, high-bandwidth signals in liquid argon.
  • Performance was characterized using 109Cd radioactive sources at room temperature and cosmic rays in both gas and double-phase operation.

Experimental results

Research questions

  • RQ1Can a LEM-based readout system achieve stable, high-gain electron multiplication in pure argon gas at cryogenic temperatures?
  • RQ2Can ionization electrons be efficiently extracted from liquid argon into the gas phase and amplified in a double-stage LEM?
  • RQ3Can a double-phase LAr LEM-TPC achieve high spatial and energy resolution for cosmic ray muon tracking?
  • RQ4Is the LEM robust and compatible with long-term cryogenic operation in pure argon?
  • RQ5Can the system achieve sufficient signal-to-noise ratio for low-threshold detection in dark matter and neutrino experiments?

Key findings

  • The LEM-TPC successfully detected cosmic ray tracks in pure argon gas at room temperature, demonstrating full signal chain functionality.
  • The system achieved a measurable gain of approximately 10^3 in double-phase operation at 87 K and 1 bar pressure, confirming effective electron multiplication.
  • The gain was tunable over a range of 14 ± 0.3 kV/cm electric field in the LEM holes, as confirmed by 109Cd source calibration.
  • Cosmic muon tracks were clearly reconstructed in double-phase operation, with orthogonal coordinate readout on LEM and anode strips showing excellent signal-to-noise ratio.
  • A PMT signal revealed three distinct light peaks: direct scintillation, proportional scintillation in the gas, and light from the avalanche process, confirming LEM operation.
  • Initial LEM wetting by liquid argon was observed but reversed after a few hours, indicating no permanent degradation and good compatibility with liquid argon.

Read this paper

Open the full text in the Nubint reader and ask about any passage as you read.

Read this paper on Nubint

This review was generated by AI from the original paper. Please check the paper before citing it.