[Paper Review] Multi-Axis Force Sensing in Robotic Minimally Invasive Surgery With No Instrument Modification

A. H. Hadi-Hosseinabadi, Septimiu E. Salcudean|arXiv (Cornell University)|Mar 20, 2021
Soft Robotics and ApplicationsEngineering42 references5 citations
TL;DR

This paper presents a novel 6-axis optical force sensor mounted on the proximal shaft of a da Vinci EndoWrist instrument without modifying the surgical tool, using a dual-tube cannula with a compliant leaf spring to isolate tip forces. The sensor achieves near 10% accuracy in lateral force measurement and ~6% in axial torque, validated via model-based and data-driven calibration using a shallow neural network.

ABSTRACT

This paper presents a novel multi-axis force-sensing approach in robotic minimally invasive surgery with no modification to the surgical instrument. Thus, it is adaptable to different surgical instruments. A novel 6-axis optical force sensor, with local signal conditioning and digital electronics, was mounted onto the proximal shaft of a da Vinci EndoWrist instrument. A new cannula design comprising an inner tube and an outer tube was proposed. The inner tube is attached to the cannula interface to the robot base through a compliant leaf spring with adjustable stiffness. It allows bending of the instrument shaft due to the tip forces. The outer tube mechanically filters out the body forces from affecting the instrument bending behavior. A mathematical model of the sensing principle was developed and used for model-based calibration. A data-driven calibration based on a shallow neural network architecture comprising a single 5-nodes hidden layer and a 5x1 output layer is discussed. Extensive testing was conducted to validate that the sensor can successfully measure the lateral forces and moments and the axial torque applied to the instruments distal end within the desired resolution, accuracy, and range requirements.

Motivation & Objective

  • To enable multi-axis force sensing in robotic minimally invasive surgery without modifying existing surgical instruments.
  • To address the lack of haptic feedback in current robotic surgical systems, which limits tissue perception and force control.
  • To develop a sensor system that is adaptable across different instrument types and compatible with clinical workflows.
  • To minimize interference from body wall forces and grasping forces through mechanical filtering and sensor design.
  • To achieve high-accuracy force measurement using a combination of model-based and data-driven calibration techniques.

Proposed method

  • A 6-axis optical force sensor with local signal conditioning and digital electronics is mounted on the proximal shaft of a da Vinci EndoWrist instrument.
  • A novel dual-tube cannula design uses an inner tube attached via a 3-arm leaf spring to the robot base, allowing instrument bending under tip forces while isolating body wall forces.
  • The effective stiffness of the inner tube is adjustable by rotating the inner tube, preventing closure of the 1.5 mm gap under maximum lateral loads.
  • A mathematical model of the sensor’s mechanical behavior is developed for model-based calibration, capturing the changing boundary conditions during instrument insertion.
  • A data-driven calibration method uses a shallow neural network with one 5-node hidden layer and a 5×1 output layer to map sensor outputs to force/torque components.
  • Calibration is performed using an external reference sensor and controlled loading setup to train the network on real-world force application scenarios.

Experimental results

Research questions

  • RQ1Can a multi-axis force sensor be integrated into a robotic surgical system without modifying the surgical instrument?
  • RQ2How can body wall forces be effectively filtered out to isolate tip forces from instrument-tissue interactions?
  • RQ3To what extent can a data-driven neural network calibration improve force estimation accuracy compared to a model-based approach?
  • RQ4How do grasping forces and wrist actuation affect the accuracy of force measurements in the absence of instrument modification?
  • RQ5Can the sensor achieve clinical-level resolution and accuracy (e.g., within human Just Noticeable Difference) across all six degrees of freedom?

Key findings

  • The sensor achieved a calibration accuracy of approximately 10% for lateral forces across the 0.5–200 N range, which aligns with the human Just Noticeable Difference (JND).
  • Axial torque measurement achieved the highest accuracy at around 6%, due to the sensor’s optimal sensitivity and insensitivity to support boundary changes.
  • The data-driven calibration method outperformed the model-based approach in scenarios where the analytical model was invalid, particularly in capturing nonlinear effects like grasping force influence.
  • The modified cannula design successfully filtered out body wall forces, as confirmed by the overcoat test, demonstrating effective mechanical isolation.
  • Wrist maneuver tests revealed that grasping forces had a more dominant effect on measurement error than wrist actuation in pitch and yaw axes.
  • The sensor design is compatible with da Vinci X and XI systems but not with older S/SI models due to interference with the sterile adapter interface.