Skip to main content
QUICK REVIEW

[Paper Review] Multi-Grid Boron-10 detector for large area applications in neutron scattering science

K.H. Andersen, T. Bigault|arXiv (Cornell University)|Sep 4, 2012
Nuclear Physics and ApplicationsPhysics and Astronomy1 references18 citations
TL;DR

This paper presents a multi-grid boron-10 detector using 10B4C-coated aluminum blades and gas-filled cells as a viable replacement for 3He-based neutron detectors in large-area neutron scattering applications. The detector achieves 53% neutron detection efficiency with sub-1% uniformity across 0.16 m² active area, enabling high-resolution 3D imaging with 5760 voxels and improved gamma rejection through optimized gas pressure and anode-wire readout.

ABSTRACT

The present supply of 3He can no longer meet the detector demands of the upcoming ESS facility and continued detector upgrades at current neutron sources. Therefore viable alternative technologies are required to support the development of cutting-edge instrumentation for neutron scattering science. In this context, 10B-based detectors are being developed by collaboration between the ESS, ILL, and Linköping University. This paper reports on progress of this technology and the prospects applying it in modern neutron scattering experiments. The detector is made-up of multiple rectangular gas counter tubes coated with B4C, enriched in 10B. An anode wire reads out each tube, thereby giving position of conversion in one of the lateral co-ordinates as well as in depth of the detector. Position resolution in the remaining co-ordinate is obtained by segmenting the cathode tube itself. Boron carbide films have been produced at Linköping University and a detector built at ILL. The characterization study is presented in this paper, including measurement of efficiency, effects of the fill gas species and pressure, coating thickness variation on efficiency and sensitivity to gamma-rays.

Motivation & Objective

  • Address the global shortage of 3He by developing a scalable, high-efficiency alternative for large-area neutron detectors.
  • Enable large-area neutron detection for next-generation neutron scattering facilities like the European Spallation Source (ESS).
  • Achieve high spatial resolution and position sensitivity comparable to 3He detectors using solid 10B4C converters and segmented gas cells.
  • Minimize efficiency non-uniformity across the detector surface through systematic compensation of coating thickness gradients.
  • Optimize gamma-ray rejection performance to ensure reliable neutron signal detection in background-rich environments.

Proposed method

  • The detector uses 28 thin 10B4C-coated aluminum blades (1.0 μm thickness, 97% 10B enrichment) deposited via DC magnetron sputtering on 0.5 mm Al substrates.
  • Blades are arranged in 96 orthogonal grids with 0.4 mm spacing, forming 15 gas cells per row, each with a central anode wire for position readout.
  • Neutron detection relies on 10B(n,α)7Li reaction in 10B4C films; conversion products ionize gas in cells, generating detectable signals.
  • Position resolution is achieved via anode wire (1D) and segmented cathode grids (1D), enabling 3D voxel reconstruction (4×15×96 = 5760 voxels).
  • Coating thickness gradients (up to 11%) were compensated by alternating blade orientation in each frame to ensure uniform efficiency.
  • Detectors were tested with CF4 and Ar/CO2 gas mixtures at 1 bar; gamma response was measured using multiple gamma sources at varying bias voltages.

Experimental results

Research questions

  • RQ1Can a multi-grid 10B4C detector achieve neutron detection efficiency comparable to 3He detectors while maintaining large-area coverage?
  • RQ2How can coating thickness non-uniformity across large-area 10B4C films be compensated to ensure detector uniformity?
  • RQ3What is the optimal gas type and pressure for maximizing neutron-to-gamma discrimination in a solid converter-based detector?
  • RQ4To what extent does the detector’s gamma sensitivity affect neutron signal fidelity, and can it be mitigated through operational parameters?
  • RQ5Can the detector achieve sub-1% uniformity in efficiency across a 0.16 m² active area using blade orientation compensation?

Key findings

  • The detector achieved a neutron detection efficiency of 53% for 2.5 Å neutrons when accounting for Al substrate and vessel scattering losses.
  • The intrinsic efficiency without substrate losses was 58% for 28 layers of 1.0 μm thick 10B4C films with 97% 10B enrichment.
  • Coating thickness variation across the blade surface was measured at 11%, with a maximum 8% gradient in top and bottom blade positions.
  • By alternating blade orientation in each frame, efficiency non-uniformity was reduced to below 1% across the active area.
  • Gamma detection efficiency was below 10−6 at the lowest bias voltage, indicating excellent gamma rejection potential.
  • The detector demonstrated a 2% efficiency deviation in non-compensated frames where blade orientation was not alternated, confirming the effectiveness of the compensation strategy.

Read this paper

Open the full text in the Nubint reader and ask about any passage as you read.

Read this paper on Nubint

This review was generated by AI from the original paper. Please check the paper before citing it.