[Paper Review] Percentage depth dose distributions in inhomogeneous phantoms with lung and bone equivalent media for small fields of CyberKnife
This study investigates percentage depth dose (PDD) distributions in inhomogeneous phantoms with lung (balsa wood) and bone (epoxy-calcium carbonate composite) equivalents using Gafchromic EBT film and GEANT4 Monte Carlo simulations for 6MV CyberKnife small fields. It demonstrates that EBT film significantly alters measured doses—especially in low-density regions—requiring its inclusion in simulations for accurate PDD estimation, with simulations including the film agreeing with measurements within 2.2% error across all field sizes.
The percentage depth dose distributions in inhomogeneous phantoms with lung and bone equivalent media are studied. For lung equivalent media a Balsa wood is used, and for a soft bone equivalent media a compound material with epoxy resin, hardener and calcium carbonate is used. Polystyrene slabs put together with these materials are used as an inhomogeneous phantom. Dose measurements are performed with Gafchromic EBT film by using photon beams from 6MV CyberKnife at the Seoul Uridul Hospital. The cone sizes of the photon beams are varied from 5, 10 to 30 mm. As a simulation tool GEANT4 Monte Carlo code v9.4.p02 is used. When the Balsa wood is inserted in the phantom, the dose measured with EBT film is found to be significantly different from the dose without the EBT film in and beyond the Balsa wood region, particularly for small field sizes. On the other hand, when the soft bone equivalent material is inserted in the phantom, discrepancy between the dose measured with EBT film and the dose without EBT film can be seen only in the region of bone equivalent material. GEANT4 simulations are done with and without the EBT film to compare the simulation results with measurements. We find that the simulations including EBT film agree with the measurements for all the cases within an error of $2.2\%$. Also, we find the "doses to phantom" without the EBT film differ from the "doses to film" up to $29\%$, which shows that for accurate dose estimations for inhomogeneous phantoms with EBT film the presence of the EBT film needs to be taken into account properly in particular for small fields.
Motivation & Objective
- To evaluate the impact of Gafchromic EBT film on percentage depth dose (PDD) measurements in inhomogeneous phantoms with lung and bone equivalent materials.
- To assess the accuracy of GEANT4 Monte Carlo simulations in modeling PDD distributions when EBT film is present or absent in small-field CyberKnife beams.
- To compare measured PDDs with those from the clinical treatment planning system Multiplan and validate them against GEANT4 simulations.
- To quantify the discrepancy between 'dose to phantom' and 'dose to film' in low- and high-density media, particularly for small fields.
- To verify the applicability of published correction factors for inhomogeneity in soft and cortical bone using experimental and simulated data.
Proposed method
- Measured PDD distributions using Gafchromic EBT film in phantoms with balsa wood (lung equivalent) and epoxy-calcium carbonate composite (soft bone equivalent) at 6MV CyberKnife beam output.
- Performed GEANT4 Monte Carlo simulations of the same setup, including and excluding the EBT film to isolate its dosimetric impact.
- Used a polystyrene phantom with embedded inhomogeneities to simulate clinical inhomogeneous tissue conditions.
- Compared experimental PDDs (with EBT film) to GEANT4-simulated PDDs (with and without film) and to Multiplan treatment planning system results.
- Calculated ratios of 'dose to film' (DTF) to 'dose to phantom' (DTP) to quantify film-induced dose perturbations.
- Applied correction factors from the literature (e.g., Siebers et al.) to validate their consistency with measured and simulated data.
Experimental results
Research questions
- RQ1How does the presence of Gafchromic EBT film affect measured percentage depth dose (PDD) in inhomogeneous phantoms with low-density (lung-like) and high-density (bone-like) materials?
- RQ2To what extent do GEANT4 Monte Carlo simulations that include the EBT film agree with experimental PDD measurements for small-field CyberKnife beams?
- RQ3How do the clinical treatment planning system Multiplan and the GEANT4 simulations compare with experimental data in inhomogeneous media?
- RQ4What is the magnitude of the discrepancy between 'dose to phantom' and 'dose to film' in different inhomogeneous regions, especially for 5 mm field sizes?
- RQ5Do published correction factors for inhomogeneity (e.g., for soft and cortical bone) align with the measured and simulated PDD values in this study?
Key findings
- GEANT4 simulations that include the EBT film agree with experimental measurements within 2.2% for all cone sizes (5, 10, 30 mm) and both inhomogeneous phantoms.
- The 'dose to phantom' (DTP) without EBT film differs from the 'dose to film' (DTF) by up to 29% for 5 mm cone size, particularly in the balsa wood (lung-equivalent) region.
- In the soft bone equivalent region, the DTF to DTP ratio is consistently 1.05 across all cone sizes, indicating a stable film-induced perturbation.
- Multiplan fails to reproduce measured PDDs, especially in inhomogeneous regions, with ratios to GEANT4 DTP of 0.94–0.99 for soft bone and 0.99–1.02 for balsa wood.
- The study confirms the correction factors from Siebers et al. (1.035 for soft bone, 1.117 for cortical bone) are consistent with the measured DTF/DTP ratios of 1.049 ± 0.015 and 1.129 ± 0.015, respectively, within 1.3%.
- For small fields (5 mm), the discrepancy between DTP and DTF is most pronounced, emphasizing the need to model EBT film explicitly in Monte Carlo simulations for accurate inhomogeneous dosimetry.