[Paper Review] Performance assessment of the 2$\gamma$ positronium imaging with the total-body PET scanners
This study demonstrates the feasibility of ortho-positronium mean lifetime imaging using two-photon annihilation events in total-body PET scanners. By leveraging triple coincidence events (two 511 keV photons from ortho-positronium decay and one prompt gamma from 44Sc), the method achieves high sensitivity and sub-20 ps time resolution for lifetime mapping, enabling non-invasive in-vivo diagnostics of tissue microstructure via positronium dynamics.
In living organisms the positron-electron annihilation (occurring during the PET imaging) proceeds in about 30% via creation of a metastable ortho-positronium atom. In the tissue, due to the pick-off and conversion processes, over 98% of ortho-positronia annihilate into two 511~keV photons. In this article we assess the feasibility for reconstruction of the mean ortho-positronium lifetime image based on annihilations into two photons. The main objectives of this work include: (i) estimation of the sensitivity of the total-body PET scanners for the ortho-positronium mean lifetime imaging using $2\gamma$ annihilations, and (ii) estimation of the spatial and time resolution of the ortho-positronium image as a function of the coincidence resolving time (CRT) of the scanner. Simulations are conducted assuming that radiopharmaceutical is labelled with $^{44}Sc$ isotope emitting one positron and one prompt gamma. The image is reconstructed on the basis of triple coincidence events. The ortho-positronium lifetime spectrum is determined for each voxel of the image. Calculations were performed for cases of total-body detectors build of (i) LYSO scintillators as used in the EXPLORER PET, and (ii) plastic scintillators as anticipated for the cost-effective total-body J-PET scanner. To assess the spatial and time resolution the three cases were considered assuming that CRT is equal to 140ps, 50ps and 10ps. The estimated total-body PET sensitivity for the registration and selection of image forming triple coincidences is larger by a factor of 12.2 (for LYSO PET) and by factor of 4.7 (for plastic PET) with respect to the sensitivity for the standard $2\gamma$ imaging by LYSO PET scanners with AFOV=20cm.
Motivation & Objective
- To assess the sensitivity of total-body PET scanners for ortho-positronium mean lifetime imaging using 2γ annihilation events.
- To evaluate spatial and temporal resolution of the reconstructed ortho-positronium image as a function of coincidence resolving time (CRT).
- To determine whether existing time-of-flight (TOF) PET systems can be adapted for this novel imaging modality.
- To explore the potential of this method as a complementary diagnostic tool to standard SUV-based PET, sensitive to tissue nanostructure and bioactive molecules.
Proposed method
- Simulations were conducted using Monte Carlo methods to model positron emission from 44Sc, which emits one positron and one prompt gamma.
- Triple coincidence events were reconstructed, combining two 511 keV photons from ortho-positronium decay and one prompt gamma to tag the formation time.
- The ortho-positronium lifetime spectrum was calculated per voxel using time-of-flight information and event-by-event reconstruction.
- Two scanner configurations were simulated: LYSO scintillators (as in EXPLORER PET) and plastic scintillators (as in J-PET), to compare performance.
- Time resolution was evaluated at CRT values of 140 ps, 50 ps, and 10 ps to assess impact on image quality.
- A TOF-FBP reconstruction algorithm was applied to estimate spatial resolution and image fidelity.
Experimental results
Research questions
- RQ1Can total-body PET scanners detect and reconstruct ortho-positronium mean lifetime images using only two 511 keV photons from decay, without requiring three-photon events?
- RQ2What is the sensitivity gain of total-body PET over conventional 20 cm axial FOV PET for detecting triple coincidence events involving 2γ + γprompt?
- RQ3How does the coincidence resolving time (CRT) affect the spatial and temporal resolution of the reconstructed ortho-positronium lifetime image?
- RQ4Can this method achieve sub-20 ps time resolution for mean lifetime determination within a clinically feasible 20-minute scan duration?
- RQ5Is attenuation correction necessary for ortho-positronium lifetime images, given that lifetime is independent of photon attenuation?
Key findings
- The sensitivity for detecting image-forming triple coincidences (2γ + γprompt) is 12.2 times higher in LYSO-based total-body PET and 4.7 times higher in plastic-based total-body PET compared to conventional 20 cm axial FOV LYSO PET.
- Spatial resolution of the ortho-positronium image reaches TOF-FBP quality levels at CRT = 50 ps, indicating high diagnostic potential.
- For a 20-minute scan, a time resolution better than 20 ps is achievable for mean ortho-positronium lifetime image determination.
- The method is inherently robust to attenuation, as the mean lifetime spectrum is unaffected by photon attenuation, eliminating the need for attenuation correction.
- The approach is directly applicable to existing TOF-PET systems, enabling near-term clinical translation.
- The 2γ-based method offers higher detection sensitivity and lower photon attenuation than 3γ-based positronium imaging, making it more suitable for in-vivo applications.