[Paper Review] Phonon and charge signals from IR and X excitation in the SELENDIS Ge cryogenic detector

H. Lattaud, Q. Arnaud|arXiv (Cornell University)|Dec 22, 2021
Dark Matter and Cosmic PhenomenaPhysics and Astronomy16 references3 citations
TL;DR

This study demonstrates single electron-hole pair detection in a 3.3g cryogenic germanium detector using infrared laser excitation and the Neganov-Trofimov-Luke (NTL) effect. At 1550 nm, quantum efficiency for pair creation reaches nearly 100%, and the detector resolves the Poisson distribution of pair events at 54 V bias, confirming single-pair sensitivity despite energy resolution limitations (40 eV RMS).

ABSTRACT

The aim of the SELENDIS project within the EDELWEISS collaboration is to observe single $e^- h^+$ pairs in lightweight (3.3 g) cryogenic germanium bolometers with charge and phonon readout at biases up to $\sim 100$ V. These devices are ideal to characterize in detail the mechanism of charge creation and collection in cryogenic germanium detectors. Electron-hole pairs are produced in the bulk of the detector either by the injection of pulsed IR laser or by neutron activation of germanium inducing the K, L and M lines from $^{71}$Ge electron capture decays. Low-energy laser pulses are also used to probe the single $e^- h^+$ pair sensitivity of Ge bolometers. Preliminary results are used to compare these two modes of charge creation, an important step toward a detailed characterization of Ge bolometers for their use in sub-MeV Dark Matter (DM) searches.

Motivation & Objective

  • To characterize charge creation and collection mechanisms in lightweight cryogenic Ge bolometers for sub-MeV Dark Matter searches.
  • To compare IR-laser-induced and neutron-activated X-ray excitation for probing single e−h+ pair sensitivity.
  • To validate the use of the Neganov-Trofimov-Luke effect for amplifying charge signals in low-energy Ge detectors.
  • To establish baseline performance for future detectors in the EDELWEISS-SubGeV project.

Proposed method

  • Irradiated 3.3g high-purity Ge crystals with pulsed IR laser diodes at 1310, 1550, and 1650 nm to probe quantum efficiency of e−h+ pair creation.
  • Applied high biases (up to 100 V) to enhance charge drift and amplify signals via the NTL effect.
  • Used 2D heat and ionization energy spectra from K- and L-lines (10.37, 1.3 keV) for low-energy calibration.
  • Measured phonon and charge signals from laser pulses with controlled energy deposits (1.4–2.8 pairs) at 54 V bias.
  • Fitted observed energy spectra with a Poisson distribution convolved with a Gaussian of width σ ≈ 40 eV to model resolution.
  • Performed data analysis using ROOT framework and pulser-triggered coincidence detection to suppress background.

Experimental results

Research questions

  • RQ1What is the quantum efficiency of e−h+ pair creation in Ge at IR wavelengths near the bandgap (1310–1650 nm)?
  • RQ2Can the Neganov-Trofimov-Luke effect enable detection of single electron-hole pairs in cryogenic Ge detectors?
  • RQ3How well does the observed distribution of charge signals match the expected Poisson statistics for single-pair events?
  • RQ4What is the energy resolution and signal fidelity of the detector at low-energy laser excitation?
  • RQ5How does the laser-induced signal compare to X-ray calibration from 71Ge decay in terms of linearity and reproducibility?

Key findings

  • The 1550 nm laser achieved near-100% quantum efficiency for e−h+ pair creation, outperforming 1310 and 1650 nm wavelengths.
  • The detector resolved the Poisson distribution of pair events at 54 V bias, with fitted mean values of 2.82 ± 0.08 and 1.38 ± 0.04 pairs for two pulse widths.
  • The observed energy resolution was 40 eV RMS, insufficient to resolve individual Npair peaks but sufficient to confirm Poisson-like statistics.
  • The NTL effect amplified charge signals, enabling detection of sub-3 eV energy deposits corresponding to single e−h+ pairs.
  • The ratio of mean pair counts (2.82 vs. 1.38) matched the expected factor of 2 when pulse width was halved, confirming signal linearity.
  • No degradation in K-line charge collection was observed over time, indicating stable detector performance during IR calibration.