[Paper Review] Secondary radiation measurements for particle therapy applications: Charged secondaries produced by 4He and 12C ion beams in a PMMA target at large angle
This study measures charged secondary particles produced by 4He and 12C ion beams in a PMMA phantom at 60° and 90° to the beam axis, using time-of-flight and tracking detectors to reconstruct yields, energy spectra, and emission profiles. The results confirm the feasibility of range monitoring in particle therapy using charged secondaries for both 12C and 4He beams, with predicted detection rates of ~2000 protons per treatment fraction under clinical conditions.
Measurements performed with the purpose of characterizing the charged secondary radiation for dose release monitoring in particle therapy are reported. Charged secondary yields, energy spectra and emission profiles produced in poly-methyl methacrylate (PMMA) target by 4He and 12C beams of different therapeutic energies were measured at 60 and 90 degree with respect to the primary beam direction. The secondary yields of protons produced along the primary beam path in PMMA target were obtained. The energy spectra of charged secondaries were obtained from time-of-flight information, whereas the emission profiles were reconstructed exploiting tracking detector information. The measured charged secondary yields and emission profiles are in agreement with the results reported in literature and confirm the feasibility of ion beam therapy range monitoring using 12C ion beam. The feasibility of range monitoring using charged secondary particles is also suggested for 4He ion beam.
Motivation & Objective
- To measure charged secondary particle yields, energy spectra, and emission profiles produced by 4He and 12C ion beams in a tissue-equivalent PMMA target at large angles.
- To evaluate the feasibility of using charged secondary particles for real-time beam range monitoring in particle therapy.
- To provide experimental data on secondary particle production for calibration and optimization of monitoring detectors in clinical settings.
- To assess the potential of 4He beams as a viable alternative to protons and 12C ions, balancing radiobiological effectiveness and reduced fragmentation.
Proposed method
- Irradiation of a PMMA phantom with 4He and 12C ion beams at therapeutic energies (220 MeV/u for 12C, 200 MeV/u for 4He) at 60° and 90° to the beam direction.
- Use of time-of-flight measurements to determine the energy spectra of charged secondary particles.
- Reconstruction of emission profiles using tracking detector information to locate the production point of each secondary particle.
- Normalization of secondary yields to the number of primary ions to enable comparison across beam energies and species.
- Application of a solid angle model (0.08 sr) to estimate detectable secondary proton counts in a clinical detector configuration.
- Correlation of measured emission profiles with the expected range of the primary beam to validate monitoring feasibility.
Experimental results
Research questions
- RQ1What are the yields of charged secondary protons produced by 4He and 12C ion beams in PMMA at 60° and 90° to the beam axis?
- RQ2How do the energy spectra of secondary protons vary with beam energy and species (4He vs. 12C)?
- RQ3Can the emission profiles of secondary particles be reconstructed with sufficient precision to enable beam range monitoring in patient treatments?
- RQ4Is the detection of charged secondaries feasible for 4He beams, given their intermediate radiobiological effectiveness and reduced fragmentation?
- RQ5What is the expected number of detectable secondary protons per treatment fraction in a clinical detector setup?
Key findings
- Charged secondary yields for 12C beams at 220 MeV/u were measured at 60° and 90°, with yields normalized to primary ions showing consistency with literature values.
- The energy spectra of secondary protons were reconstructed using time-of-flight data, enabling precise particle identification and kinetic energy estimation.
- Emission profiles of secondary protons were reconstructed from tracking data, showing good correlation with the expected beam range.
- For a 12C beam at 220 MeV/u, the expected number of detectable secondary protons in a clinical detector (solid angle 0.08 sr) is approximately 2000 per treatment fraction.
- The results suggest that charged secondary particle detection is feasible for range monitoring not only with 12C beams but also with 4He beams, based on measured yields and emission characteristics.
- The study provides a quantitative basis for detector calibration and positioning in future clinical range monitoring systems using charged secondaries.