[Paper Review] Does FLASH deplete Oxygen? Experimental Evaluation for Photons, Protons and Carbon Ions

Jeannette Jansen, Jan Knoll|arXiv (Cornell University)|Feb 25, 2021
Radiation Therapy and DosimetryMedicine29 references156 citations
TL;DR

This study experimentally investigates oxygen depletion in water during FLASH radiotherapy using photons, protons, and carbon ions at dose rates up to 340 Gy/s. Using non-invasive optical oxygen sensors in 3D-printed phantoms, it finds that oxygen consumption per 10 Gy ranges from 0.04% to 0.25% atm, decreases with increasing dose rate due to radical self-reactions, and never reaches total depletion—challenging the oxygen depletion theory as the sole explanation for the FLASH effect.

ABSTRACT

Purpose: To investigate experimentally, if FLASH irradiation depletes oxygen within water for different radiation types such as photons, protons and carbon ions. Methods: This study presents measurements of the oxygen consumption in sealed, 3D printed water phantoms during irradiation with X-rays, protons and carbon ions at varying dose rates up to 340 Gy/s. The oxygen measurement was performed using an optical sensor allowing for non-invasive measurements. Results: Oxygen consumption in water only depends on dose, dose rate and linear energy transfer (LET) of the irradiation. The total amount of oxygen depleted per 10 Gy was found to be 0.04 - 0.18 % atm for 225 kV photons, 0.04 - 0.25 % atm for 224 MeV protons and 0.09 - 0.17 % atm for carbon ions. consumption depends on dose rate by an inverse power law and saturates for higher dose rates because of self-interactions of radicals. Higher dose rates yield lower oxygen consumption. No total depletion of oxygen was found for clinical doses. Conclusions: FLASH irradiation does consume oxygen, but not enough to deplete all the oxygen present. For higher dose rates, less oxygen was consumed than at standard radiotherapy dose rates. No total depletion was found for any of the analyzed radiation types for 10 Gy dose delivery using FLASH.

Motivation & Objective

  • To experimentally assess whether FLASH irradiation depletes dissolved oxygen in water across different radiation types.
  • To evaluate the dependence of oxygen consumption on dose rate, total dose, and linear energy transfer (LET) for photons, protons, and carbon ions.
  • To determine if oxygen depletion could explain the FLASH effect's radioprotective benefit in healthy tissue.
  • To validate or challenge existing theoretical and simulation models of radiolytic oxygen consumption in pure water.

Proposed method

  • Non-invasive, real-time oxygen monitoring using TROXSP5 optical sensors embedded in 3D-printed, airtight, transparent water phantoms.
  • Irradiation with 225 kV photons, 224 MeV protons, and 12C ions at dose rates from 2 Gy/min to 340 Gy/s.
  • Use of FireStingO2 system for optical excitation and fluorescence readout with ~400 ms temporal resolution.
  • Measurement of oxygen consumption per unit dose (dO/dD) as a function of dose rate and radiation type.
  • Controlled experiments with varying phantom volumes and initial oxygen concentrations to assess linearity and independence of initial conditions.
  • Application of the inverse power law model (a + b·(dD/dt)^−0.5) to describe non-linear dose rate dependence of oxygen consumption.

Experimental results

Research questions

  • RQ1Does FLASH irradiation deplete oxygen in water during irradiation with photons, protons, and carbon ions?
  • RQ2How does oxygen consumption vary with dose rate, total dose, and radiation quality (LET) in pure water?
  • RQ3Is the observed oxygen consumption consistent with theoretical models and prior experimental data?
  • RQ4Can oxygen depletion alone explain the FLASH effect’s differential radioprotection in normal versus tumor tissue?
  • RQ5What is the role of radical self-reactions (e.g., e⁻aq + e⁻aq) in modulating oxygen consumption at ultra-high dose rates?

Key findings

  • Oxygen consumption per 10 Gy ranged from 0.04% to 0.18% atm for 225 kV photons, 0.04% to 0.25% atm for 224 MeV protons, and 0.09% to 0.17% atm for carbon ions, depending on dose rate.
  • Oxygen consumption decreased with increasing dose rate, following an inverse power law relationship, indicating radical self-reactions reduce steady-state radical concentrations.
  • The reduction in oxygen consumption at higher dose rates was attributed to self-reactions of solvated electrons (e⁻aq) and hydrogen atoms (H), which compete with oxygen for radicals.
  • No total oxygen depletion occurred for any radiation type at 10 Gy, even at the highest dose rates tested.
  • The experimental results are in good agreement with prior experimental and modeling studies, confirming non-linear dose rate dependence of oxygen consumption in pure water.
  • The findings challenge the hypothesis that oxygen depletion is the primary mechanism of the FLASH effect, as depletion is insufficient and actually reduced at FLASH dose rates.