[Paper Review] Simultaneous Reduction of Two Common Autocalibration Errors in GRAPPA EPI Time Series Data

Daniel Sheltraw, Ben Inglis|arXiv (Cornell University)|Aug 5, 2012
Advanced MRI Techniques and ApplicationsMedicine13 references6 citations
TL;DR

This paper proposes a method to simultaneously reduce two major autocalibration errors in GRAPPA EPI time series: artifacts from inter-shot signal perturbations during R-shot ACS acquisition and suboptimal ACS data selection. By using phase correction scans as perturbation-sensitive navigators, the method computes a root-mean-square deviation metric to select the cleanest complete ACS dataset, significantly improving image quality and temporal SNR with minimal added scan time.

ABSTRACT

The GRAPPA (GeneRalized Autocalibrating Partially Parallel Acquisitions) method of parallel MRI makes use of an autocalibration scan (ACS) to determine a set of synthesis coefficients to be used in the image reconstruction. For EPI time series the ACS data is usually acquired once prior to the time series. In this case the interleaved R-shot EPI trajectory, where R is the GRAPPA reduction factor, offers advantages which we justify from a theoretical and experimental perspective. Unfortunately, interleaved R-shot ACS can be corrupted due to perturbations to the signal (such as direct and indirect motion effects) occurring between the shots, and these perturbations may lead to artifacts in GRAPPA-reconstructed images. Consequently we also present a method of acquiring interleaved ACS data in a manner which can reduce the effects of inter-shot signal perturbations. This method makes use of the phase correction data, conveniently a part of many standard EPI sequences, to assess the signal perturbations between the segments of R-shot EPI ACS scans. The phase correction scans serve as navigator echoes, or more accurately a perturbation-sensitive signal, to which a root-mean-square deviation perturbation metric is applied for the determination of the best available complete ACS data set among multiple complete sets of ACS data acquired prior to the EPI time series. This best set (assumed to be that with the smallest valued perturbation metric) is used in the GRAPPA autocalibration algorithm, thereby permitting considerable improvement in both image quality and temporal signal-to-noise ratio of the subsequent EPI time series at the expense of a small increase in overall acquisition time.

Motivation & Objective

  • To address artifacts in GRAPPA-reconstructed EPI time series caused by signal perturbations between shots in R-shot autocalibration scans.
  • To improve image quality and temporal signal-to-noise ratio (tSNR) in parallel imaging EPI by selecting the most stable ACS data set.
  • To leverage existing phase correction scans as navigators to detect inter-shot signal perturbations during ACS acquisition.
  • To enable robust GRAPPA reconstruction by selecting the ACS dataset with the lowest perturbation metric, minimizing error propagation.
  • To maintain high acceleration efficiency while reducing artifacts from field inhomogeneity and motion-induced signal drift.

Proposed method

  • Uses R-shot interleaved EPI trajectories to acquire autocalibration scan (ACS) data, enabling better handling of field inhomogeneity compared to 1-shot ACS.
  • Utilizes phase correction scans—already present in standard EPI sequences—as navigator echoes to monitor signal stability between ACS shot segments.
  • Applies a root-mean-square (RMS) deviation metric to quantify signal perturbations across multiple complete ACS datasets acquired prior to the time series.
  • Selects the ACS dataset with the smallest RMS perturbation value as the optimal input for GRAPPA autocalibration, minimizing error from corrupted data.
  • Implements the selected optimal ACS data in the GRAPPA reconstruction process to synthesize missing k-space lines using coil sensitivity information.
  • Employs a block-wise coefficient estimation strategy where synthesis coefficients are computed locally, assuming slow spatial variation of coil sensitivities and system operators.

Experimental results

Research questions

  • RQ1How do inter-shot signal perturbations during R-shot ACS acquisition affect GRAPPA reconstruction quality in EPI time series?
  • RQ2Can phase correction scans be effectively repurposed as navigators to detect and quantify signal instability in ACS data?
  • RQ3What impact does selecting the least-perturbed ACS dataset have on image quality and temporal SNR in GRAPPA-reconstructed EPI time series?
  • RQ4How does the RMS perturbation metric compare to other selection criteria in identifying the most reliable ACS dataset?
  • RQ5To what extent does the proposed method reduce artifacts compared to standard 1-shot or unselected R-shot ACS approaches?

Key findings

  • The proposed method significantly improves image quality in GRAPPA-reconstructed EPI time series by reducing artifacts caused by inter-shot signal perturbations.
  • Temporal signal-to-noise ratio (tSNR) is notably enhanced by selecting the ACS dataset with the lowest RMS perturbation metric.
  • The use of phase correction scans as navigators enables robust, real-time assessment of ACS data stability without additional hardware or sequence modifications.
  • The method maintains high acceleration efficiency and is compatible with standard EPI sequences, requiring only minor software-level changes for implementation.
  • The selection of the optimal ACS dataset leads to more accurate synthesis of missing k-space lines, reducing reconstruction errors and improving fidelity.
  • The approach is particularly effective at higher field strengths and higher reduction factors, where field inhomogeneity and signal drift are more pronounced.