[Paper Review] Sorting of particles suspended in whole blood
This study investigates deterministic lateral displacement (DLD) for sorting microparticles in whole blood, demonstrating that DLD maintains effective separation despite hematocrit-induced trajectory shifts and dispersion near critical particle sizes. Simulations and experiments confirm the robustness of DLD in whole blood with minimal sample prep, enabling high-throughput diagnostics.
An important step in diagnostics is the isolation of specific cells and microorganisms of interest from blood. Since such bioparticles are often present at very low concentrations, throughput needs to be as high as possible. In addition, to ensure simplicity, a minimum of sample preparation is important. Therefore, sorting schemes that function for whole blood are highly desirable. Deterministic lateral displacement (DLD) has proven to be very precise and versatile in terms of a wide range of sorting parameters. To better understand how DLD performs for blood as the hematocrit increases, we have performed measurements and simulations for spherical particles in the micrometer range moving through DLD arrays for different flow velocities and hematocrits ranging from pure buffer to whole blood. We find that the separation function of the DLD array is sustained, even though blood cells introduce a shift in the trajectories and a significant dispersion for particles that are close to the critical size in the device. Simulations qualitatively replicate our experimental observations and help us identify fundamental mechanisms for the effect of hematocrit on the performance of the DLD device.
Motivation & Objective
- To evaluate the performance of deterministic lateral displacement (DLD) in whole blood for high-throughput isolation of bioparticles.
- To understand how increasing hematocrit affects particle trajectory and dispersion in DLD arrays.
- To assess the feasibility of using DLD for direct sorting of particles in whole blood without extensive preprocessing.
- To validate experimental results with simulations that capture hematocrit-dependent effects on particle separation.
Proposed method
- Experiments were conducted using spherical microparticles in DLD arrays under varying flow velocities and hematocrit levels, from pure buffer to whole blood.
- Particle trajectories were tracked in real time to quantify shifts and dispersion relative to the expected separation behavior.
- Numerical simulations were performed to model particle motion in DLD arrays under different hematocrit conditions.
- The simulations incorporated hydrodynamic interactions and cell-induced flow disturbances to replicate experimental observations.
- The separation function was evaluated by comparing particle focusing behavior across different hematocrit levels and particle sizes.
- Critical particle size was identified as the threshold where dispersion increases significantly due to hematocrit effects.
Experimental results
Research questions
- RQ1How does increasing hematocrit in whole blood affect the trajectory and focusing of microparticles in DLD arrays?
- RQ2To what extent is the DLD separation function preserved when processing whole blood versus diluted buffers?
- RQ3What mechanisms underlie the increased dispersion of particles near the critical size in high-hematocrit environments?
- RQ4Can simulations accurately reproduce the experimental behavior of particle sorting in whole blood?
Key findings
- The DLD separation function remains effective even in whole blood, demonstrating sustained particle focusing across a range of hematocrit levels.
- Hematocrit increases cause measurable shifts in particle trajectories, particularly for particles near the critical size for separation.
- Particles near the critical size exhibit significant dispersion in whole blood due to hydrodynamic interactions with red blood cells.
- Simulations qualitatively reproduce experimental observations, identifying hematocrit-induced flow heterogeneity as a key factor in trajectory deviation.
- The performance of DLD in whole blood is robust for particles significantly larger or smaller than the critical size, but degradation occurs near the critical threshold.
- Minimal sample preparation is feasible, supporting the use of DLD as a high-throughput diagnostic tool directly on whole blood.
Read this paper
Open the full text in the Nubint reader and ask about any passage as you read.
Read this paper on NubintThis review was generated by AI from the original paper. Please check the paper before citing it.