Skip to main content
QUICK REVIEW

[Paper Review] Technical Report (v1.0)--Pseudo-random Cartesian Sampling for Dynamic MRI

Mihir Joshi, Aaron Pruitt|arXiv (Cornell University)|Jun 8, 2022
Advanced MRI Techniques and ApplicationsMedicine6 citations
TL;DR

This technical report presents five pseudo-random Cartesian sampling methods—VISTA, GRO, CAVA, OPRA, and PR4D—designed for dynamic MRI, including 2D cine, 3D volumetric cine, and 4D flow imaging. The methods leverage irrational number sequences (e.g., golden ratio, Fekete points) to generate incoherent, variable-density sampling patterns with fast on-the-fly computation and parameterized control over sampling density, enabling efficient compressed sensing in dynamic cardiac MRI.

ABSTRACT

For an effective application of compressed sensing (CS), which exploits the underlying compressibility of an image, one of the requirements is that the undersampling artifact be incoherent (noise-like) in the sparsifying transform domain. For cardiovascular MRI (CMR), several pseudo-random sampling methods have been proposed that yield a high level of incoherence. In this technical report, we present a collection of five pseudo-random Cartesian sampling methods that can be applied to 2D cine and flow, 3D volumetric cine, and 4D flow imaging. Four out of the five presented methods yield fast computation for on-the-fly generation of the sampling mask, without the need to create and store pre-computed look-up tables. In addition, the sampling distribution is parameterized, providing control over the sampling density. For each sampling method in the report, (i) we briefly describe the methodology, (ii) list default values of the pertinent parameters, and (iii) provide a publicly available MATLAB implementation.

Motivation & Objective

  • To develop efficient, incoherent sampling patterns for dynamic MRI that reduce aliasing artifacts and support compressed sensing.
  • To enable fast, on-the-fly generation of sampling masks without precomputed lookup tables, improving clinical feasibility.
  • To provide parameterized control over sampling density and temporal resolution for flexible application across dynamic MRI sequences.
  • To support a range of dynamic MRI applications including 2D cine, 3D volumetric cine, and 4D flow imaging with variable density and temporal resolution.
  • To offer publicly available MATLAB implementations for reproducibility and integration into research and clinical workflows.

Proposed method

  • VISTA uses Fekete point optimization to minimize Riesz energy on the $k_y$-$t$ grid, ensuring incoherence and variable density with controlled k-space jumps.
  • GRO employs golden ratio shifts between frames to generate pseudo-random sampling in the $k_y$-$t$ domain, enabling fast on-the-fly mask generation.
  • CAVA extends GRO by allowing retrospective adjustment of temporal resolution through golden ratio offset sampling, with a trade-off in uniformity.
  • OPRA samples $k_y$-$k_z$ space using L-shaped leaflets rotated by a golden angle, with non-linear radial scaling to create variable density and reduce k-space jumps.
  • PR4D uses polar coordinates with two irrational numbers ($g_s = ilde{ heta}$, $2-g$ for radius) to sample $k_y$-$k_z$-$t$ space without leaflet constraints, enabling even coverage at the cost of larger k-space jumps.
  • All methods convert polar to Cartesian coordinates and support parameterized control via variables such as $s$ (density extent), $g_s$ (irrational step), and $ heta$ (angular jump), with MATLAB implementations provided.

Experimental results

Research questions

  • RQ1Can pseudo-random Cartesian sampling patterns be generated efficiently on-the-fly without precomputed lookup tables while maintaining incoherence for compressed sensing in dynamic MRI?
  • RQ2How does parameterized variable density in sampling patterns affect image reconstruction quality and temporal resolution control in 2D and 3D dynamic MRI?
  • RQ3What is the trade-off between k-space jump size and sequence compatibility, particularly for steady-state free precession (SSFP) sequences?
  • RQ4How do different sampling strategies (e.g., leaflet-based OPRA vs. full-plane PR4D) compare in terms of coverage uniformity and reconstruction performance?
  • RQ5To what extent can sampling patterns be retrospectively adjusted for temporal resolution without compromising image quality?

Key findings

  • GRO and CAVA enable fast, on-the-fly sampling mask generation without lookup tables, making them suitable for real-time clinical applications.
  • OPRA reduces k-space jumps through L-shaped leaflets rotated by a golden angle, enhancing compatibility with SSFP sequences.
  • PR4D provides more uniform coverage across $k_y$-$k_z$ space than OPRA but incurs larger k-space jumps, limiting its use with SSFP.
  • All methods support retrospective adjustment of temporal resolution (e.g., via $n$ and $F$ parameters), with CAVA offering explicit flexibility at the cost of marginal uniformity loss.
  • The use of irrational numbers (e.g., golden ratio, $ ilde{ heta}$, $ ilde{g}_s$) ensures pseudo-randomness and incoherence, critical for compressed sensing success.
  • A unified MATLAB implementation is publicly available at https://github.com/MihirJoe/cmr-sampling, with main.m as the entry point for all methods.

Read this paper

Open the full text in the Nubint reader and ask about any passage as you read.

Read this paper on Nubint

This review was generated by AI from the original paper. Please check the paper before citing it.