[Paper Review] Tests of PMT Signal Read-out in a Liquid Argon Dark Matter Detector with a New Fast Waveform Digitizer
This paper evaluates the CAEN V1751 waveform digitizer for readout of scintillation signals in liquid argon dark matter detectors using photomultiplier tubes (PMTs). Operating at 1–2 GS/s with low noise and extended dynamic range, it successfully captures the fast 6–7 ns scintillation component, demonstrating suitability for dark matter detection with high-fidelity signal reconstruction.
The CAEN V1751 is a new generation of Waveform Digitizer recently introduced by CAEN SpA. It features 8 Channels per board, 10 bit, 1 GS/s using Flash ADCs Waveform Digitizers (or 4 channels at 2 GS/s in Dual Edge Sampling mode) with threshold and Auto-Trigger capabilities. This provides a good basis for data acquisition in Dark Matter searches using PMTs to detect scintillation light in liquid argon, as it matches the requirements for measuring the fast scintillation component. The board was tested by operating it in real experimental conditions and by comparing it with a state of the art digital oscilloscope. We find that the sampling at 1 or 2 GS/s is appropriate for the reconstruction of the fast component of the scintillation light in argon (characteristic time of about 6-7 ns) and the extended dynamic range, after a small customization, allows for the detection of signals in the range of energy needed. The bandwidth is found to be adequate and the intrinsic noise is very low.
Motivation & Objective
- To assess the performance of the CAEN V1751 waveform digitizer in real experimental conditions for dark matter detection.
- To determine if the digitizer's sampling rate and dynamic range are sufficient for reconstructing fast scintillation signals in liquid argon.
- To validate the digitizer’s signal fidelity by comparing it with a state-of-the-art digital oscilloscope.
- To confirm the adequacy of the digitizer’s bandwidth and intrinsic noise for detecting low-energy PMT signals.
- To verify that minor hardware customization enables extended dynamic range for relevant energy detection.
Proposed method
- The CAEN V1751 digitizer was deployed in real experimental conditions to capture PMT signals from liquid argon scintillation light.
- The digitizer operates at 1 GS/s on 8 channels or 2 GS/s in dual-edge sampling mode on 4 channels using Flash ADCs.
- Signal reconstruction was validated by comparing digitizer outputs with those from a high-precision digital oscilloscope.
- A small hardware customization was applied to extend the dynamic range for improved sensitivity to low-energy signals.
- The bandwidth and intrinsic noise levels were measured and evaluated for suitability in dark matter detection.
- The system was tested for its ability to resolve the fast scintillation component with a characteristic decay time of ~6–7 ns.
Experimental results
Research questions
- RQ1Can the CAEN V1751 digitizer accurately sample the 6–7 ns scintillation component in liquid argon using 1–2 GS/s sampling rates?
- RQ2Does the digitizer’s extended dynamic range, after minor customization, allow for detection of signals within the energy range relevant to dark matter searches?
- RQ3Is the intrinsic noise level of the CAEN V1751 sufficiently low for reliable detection of weak scintillation signals?
- RQ4How does the digitizer’s performance compare to a state-of-the-art digital oscilloscope in real experimental conditions?
- RQ5Is the bandwidth of the CAEN V1751 adequate for capturing fast PMT signals in liquid argon detectors?
Key findings
- The CAEN V1751 digitizer successfully captures the fast scintillation component in liquid argon with a characteristic decay time of approximately 6–7 ns.
- Sampling at 1 or 2 GS/s is confirmed to be appropriate for high-fidelity reconstruction of the fast scintillation signal.
- The intrinsic noise of the digitizer is found to be very low, supporting reliable detection of weak signals.
- The digitizer’s bandwidth is adequate for the required signal bandwidth in liquid argon dark matter detectors.
- After minor hardware customization, the dynamic range is extended sufficiently to detect signals in the energy range relevant to dark matter searches.
- Comparison with a state-of-the-art digital oscilloscope confirms the digitizer’s performance is suitable for experimental use.