[Paper Review] Timing with the EPIC pn Camera of XMM-Newton
This paper evaluates the high-time-resolution capabilities of the EPIC pn camera on XMM-Newton, demonstrating millisecond-level timing accuracy for X-ray pulsars. It identifies critical limitations from detector readout modes, electronic choppers, and counting mode, which introduce aliasing and degrade time resolution, especially in Burst Mode and bright source observations.
The EPIC pn CCD camera on board of XMM-Newton is designed to perform high throughput imaging and spectroscopy as well as high resolution timing observations in the energy range of 0.1-15 keV. A temporal resolution of milliseconds or microseconds, depending on the instrument mode and detector, is outstanding for CCD based X-ray cameras. In order to calibrate the different observing modes of the EPIC pn CCD, XMM-Newton observations of the pulsars PSR B1509-58, PSR B0540-69 and the Crab were performed during the calibration and performance verification phase. To determine the accuracy of the on board clock against Coordinated Universal Time (UTC), PSR B1509-58 was observed simultaneously with XMM-Newton and RXTE in addition. The paper summarizes the current status of the clock calibration.
Motivation & Objective
- To assess the temporal resolution and timing accuracy of the EPIC pn camera during XMM-Newton's performance verification phase.
- To calibrate the on-board clock against UTC using simultaneous observations with RXTE, focusing on PSR B1509-58.
- To identify instrumental and observational constraints that degrade time resolution in high-time-resolution modes.
- To quantify the impact of detector configuration (e.g., electronic chopper, counting mode) on power spectral density (PSD) and QPO detectability.
- To provide guidance for observers on optimizing timing observations with the EPIC pn camera under practical limitations.
Proposed method
- Utilized fast readout modes—Timing (29.56 μs resolution) and Burst (7.2 μs resolution)—to achieve high temporal resolution in X-ray timing.
- Performed simultaneous observations of millisecond pulsars (PSR B1509-58, PSR B0540-69, Crab) with XMM-Newton and RXTE for cross-calibration.
- Applied time stamp corrections using OBT-to-UTC polynomial fits and barycentric correction via the SAS tool 'barycen' to transform to TDB.
- Simulated power spectra with Poissonian noise and AR[2] processes to model QPOs and assess aliasing effects from readout sampling.
- Analyzed PSDs from real observations (e.g., Crab in Rev. 0056) to detect distortions from electronic chopper (set to N=25) and counting mode.
- Used the XMM Science Analysis System (SAS) to correct event times based on source position on the CCD and readout mode.
Experimental results
Research questions
- RQ1What is the achievable time resolution of the EPIC pn camera in its fast readout modes, and how does it depend on source position?
- RQ2How accurate is the on-board clock of XMM-Newton relative to UTC, and what is the current uncertainty in timing calibration?
- RQ3To what extent do detector readout modes (e.g., Burst Mode) suppress detectability of quasi-periodic oscillations (QPOs) due to sampling effects?
- RQ4How do electronic chopper settings and counting mode degrade time resolution and distort power spectra in high-count-rate observations?
- RQ5What are the practical limitations for high-precision timing observations with the EPIC pn camera, and how can they be mitigated?
Key findings
- The EPIC pn camera achieves a time resolution of 7.2 μs in Burst Mode and 29.56 μs in Timing Mode, with optimal resolution dependent on source position on the CCD.
- The internal timing of the EPIC pn and MOS cameras functions as expected, but current timing accuracy is limited to ΔP/P ≈ 10⁻⁶, falling short of the target ΔP/P ≈ 10⁻⁸.
- In Burst Mode, the sampling rate completely masks QPOs above the critical frequency, rendering them undetectable due to severe aliasing.
- The electronic chopper (set to N=25) causes strong noise dominance above 10 Hz in the power spectrum, severely degrading high-frequency timing analysis.
- Counting mode introduces quasi-periodic dead time windows that distort the PSD at low frequencies, with broad peaks observed below 0.2 Hz.
- Rebinning can mitigate resolution loss but inherently reduces effective time resolution, making it a suboptimal solution for high-precision timing.