Skip to main content
QUICK REVIEW

[Paper Review] Ultracold neutron detection with 6Li-doped glass scintillators, NANOSC: a fast ultracold neutron detector for the nEDM experiment at the Paul Scherrer Institute

G. Ban, G. Bison|arXiv (Cornell University)|Jun 23, 2016
Radiation Detection and Scintillator TechnologiesPhysics and Astronomy31 references13 citations
TL;DR

This paper presents a fast ultracold neutron (UCN) detector based on 6Li-doped glass scintillators, using optical contact-bonded stacks of 6Li-depleted and 6Li-enriched glass to achieve high neutron-gamma discrimination. The detector, named NANOSC, achieves detection efficiency comparable to 3He gas counters, handles up to 10⁶ counts/s, and has been successfully deployed in the nEDM experiment at PSI since 2010 with robust, background-insensitive counting via pulse charge measurement.

ABSTRACT

This paper summarizes the results from measurements aiming to characterize ultracold neutron detection with 6Li-doped glass scintillators. Single GS10 or GS20 scintillators, with a thickness of 100-200 micrometer, fulfill the ultracold neutron detection requirements with an acceptable neutron-gamma discrimination. This discrimination is clearly improved with a stack of two scintillators: a 6Li-depleted glass bonded to a 6Li-enriched glass. The optical contact bonding is used between the scintillators in order to obtain a perfect optical contact. The scintillator's detection efficiency is similar to that of a 3He Strelkov gas detector. Coupled to a digital data acquisition system, counting rates up to a few 10^5 counts/s can be handled. A detector based on such a scintillator stack arrangement was built and has been used in the neutron electric dipole moment experiment at the Paul Scherrer Institute since 2010. Its response for the regular runs of the neutron electric dipole moment experiment is presented.

Motivation & Objective

  • To develop a high-performance, fast ultracold neutron (UCN) detector as an alternative to 3He gas counters, which face supply constraints and slower response times.
  • To improve neutron-gamma discrimination in UCN detection using 6Li-doped glass scintillators with optimized scintillator stack configurations.
  • To achieve high detection efficiency with minimal Fermi potential and low background from edge events and gamma interactions.
  • To design a detector capable of handling high counting rates (up to 10⁶ counts/s) suitable for long-duration nEDM experiments.
  • To ensure robust, background-insensitive UCN counting via pulse charge measurement, minimizing impact on nEDM data analysis.

Proposed method

  • Using 6Li-doped glass scintillators (GS10, GS20, GS30) with scintillation decay times of ~70 ns for fast signal response.
  • Fabricating scintillators with thicknesses of 100–200 μm to reduce gamma interaction probability and energy deposition.
  • Implementing optical contact bonding between a 6Li-depleted glass (GS30) and a 6Li-enriched glass (GS20) to ensure perfect optical coupling and suppress edge events.
  • Employing photomultiplier tubes (PMTs) for light readout and a digital data acquisition system (FASTER) capable of handling high event rates.
  • Calibrating the system using UCN and gamma sources to measure pulse height response and optimize discrimination thresholds.
  • Monitoring pulse charge stability over time to ensure consistent counting performance under varying conditions.

Experimental results

Research questions

  • RQ1Can 6Li-doped glass scintillators with optimized thickness and 6Li concentration achieve sufficient neutron-gamma discrimination for UCN detection in high-background environments?
  • RQ2Does stacking a 6Li-depleted glass with a 6Li-enriched glass improve neutron-gamma discrimination by suppressing edge events and reducing background?
  • RQ3What is the maximum counting rate capability of the 6Li-doped glass scintillator detector, and can it handle the demands of long-term nEDM experiments?
  • RQ4How stable is the pulse charge measurement over time, and what impact does it have on UCN counting accuracy and nEDM data analysis?
  • RQ5Does the detection efficiency of the 6Li-doped glass detector match or exceed that of standard 3He gas detectors in real experimental conditions?

Key findings

  • Single 6Li-doped glass scintillators (GS10 or GS20) with thicknesses of 100–200 μm achieve detection efficiency comparable to standard 3He gas detectors.
  • The use of a 6Li-depleted (GS30) and 6Li-enriched (GS20) glass stack with optical contact bonding significantly improves neutron-gamma discrimination by suppressing edge events.
  • The detector system, named NANOSC, successfully handles counting rates up to 10⁶ counts/s, with the FASTER data acquisition system enabling precise time-stamping of events.
  • Pulse charge measurements remain stable over time, with variations of up to 2% across channels, resulting in negligible counting rate fluctuations (less than 0.025% per day).
  • The UCN counting procedure based on pulse charge is robust and does not influence the nEDM data analysis, even under long-term operation.
  • Since 2014, two identical NANOSC detectors have operated in the nEDM experiment with a simultaneous spin analyzer, confirming long-term reliability and performance.

Read this paper

Open the full text in the Nubint reader and ask about any passage as you read.

Read this paper on Nubint

This review was generated by AI from the original paper. Please check the paper before citing it.