[Paper Review] Ultrasound imaging with three dimensional full-wave nonlinear acoustic simulations. Part 2: sources of image degradation in intercostal imaging
This study uses 3D full-wave nonlinear ultrasound simulations to investigate image degradation in intercostal imaging, revealing that while ribs act as beam apodizers improving sidelobe suppression, they also generate multiple reverberation clutter that degrades image quality. The optimal image quality occurs when ribs are in their anatomical position, balancing apodization benefits against reverberation penalties, with contrast-to-noise ratio (CNR) decreasing by 8% due to increased reverberation when ribs are artificially compressed.
Fullwave simulations are applied to an intercostal imaging scenario to determine the sources of fundamental and harmonic image degradation with respect to aberration and reverberation. These simulations are based on Part I of this two part paper, which established the Fullwave simulation methods to generate realistic ultrasound images based directly on the first principles of wave propagation in the human body. The ultasound images are generated based on the first principles of propagation and reflection and they describe interplay between distributed aberration and reverberation clutter. Three imaging scenarios that would not be realizable in vivo are investigated in silico. First, the ribs were completely removed and replaced with fat. Then, the ribs were maintained in their anatomically correct configuration to yield a reference image. Finally the ribs were placed closer together in elevation. The propagation based B-mode images show that of these three scenarios the second, anatomically correct configuration, has the best contrast-to-noise ratio. This is due to two competing effects. First the ribs effectively apodize the fundamental and harmonic beams by 3-5 dB. This effect alone would predict an improvement in image quality. However, the B-mode image quality, measured by the contrast-to-noise ratio degrades by 8%. To explain these changes, it is shown that a second effect, multiple reverberation, must be taken into account. A point spread function analysis shows that when the ribs are placed closer together they generate significantly more reverberation clutter (by 2.4 to 2.9 dB), degrading the image quality even though the beamplot has lower sidelobes. In this intercostal imaging scenario the effects of the ribs on beam shape and reverberation are therefore in competition in terms of image quality and there is an optimal acoustic window that balances them out.
Motivation & Objective
- To identify the primary sources of image degradation in intercostal ultrasound imaging, particularly aberration and reverberation.
- To isolate the effects of rib structure on beam shaping and reverberation using in silico simulations not realizable in vivo.
- To evaluate how rib positioning—specifically, closer spacing—affects image quality metrics like contrast-to-noise ratio (CNR).
- To demonstrate that beam profile alone is insufficient to predict image quality, and that multiple reverberation is a dominant, underappreciated factor.
- To validate the Fullwave simulation platform as a tool for quantitative image quality analysis in complex, heterogeneous anatomical environments.
Proposed method
- 3D full-wave simulations based on first principles of wave propagation in heterogeneous media were used to model ultrasound imaging in the intercostal window.
- The simulations used anatomical data from the Visible Human female cadaver dataset to generate realistic 3D acoustical maps of the thorax.
- Three imaging scenarios were simulated: ribs removed and replaced with fat, ribs in anatomical position, and ribs artificially compressed in elevation.
- Point spread functions (PSFs) were computed to isolate and quantify reverberation clutter from other image degradation sources.
- Contrast-to-noise ratio (CNR) was measured in B-mode images to evaluate image quality across the three scenarios.
- Beamplots and PSF analysis were used to disentangle the effects of aberration, apodization, and multiple reverberation on image quality.
Experimental results
Research questions
- RQ1What is the relative contribution of rib-induced aberration versus multiple reverberation to image degradation in intercostal ultrasound?
- RQ2How does rib positioning—specifically, closer spacing—affect beam shape and image quality metrics like CNR?
- RQ3Can beam apodization by ribs improve image quality despite introducing additional reverberation clutter?
- RQ4Why does image quality degrade when ribs are moved closer together, even though beam profiles show improved sidelobe suppression?
- RQ5To what extent can beam profile and aberration alone predict image quality, or is reverberation a dominant, unaccounted factor?
Key findings
- The anatomically correct rib configuration yielded the highest contrast-to-noise ratio (CNR), despite ribs causing 3–5 dB of apodization, indicating a net improvement in image quality.
- When ribs were removed and replaced with fat, CNR decreased by 8% compared to the anatomical case, despite improved beam shape, due to loss of apodization.
- Artificially compressing ribs by a few millimeters improved beam sidelobe suppression but increased reverberation clutter by 2.4–2.9 dB, degrading overall CNR.
- Reverberation clutter was quantified via PSFs: it increased from -35.4 dB (anatomical ribs) to -32.7 dB (compressed ribs) at the fundamental frequency.
- At the second harmonic, reverberation clutter increased from -51.2 dB (anatomical) to -48.9 dB (compressed), confirming that multiple reverberation dominates image degradation when ribs are closer.
- The study demonstrates that beam profile and aberration alone are insufficient predictors of image quality; multiple reverberation is a critical, often underappreciated, factor in intercostal imaging.