[Paper Review] A prototype low-background multiwire proportional chamber for measuring alphas and low-energy betas
This paper presents a prototype low-background multiwire proportional chamber (MWPC) designed for high-sensitivity detection of alpha and low-energy beta particles from surface contaminants in ultra-rare-event experiments. Fabricated with radiopure materials and optimized for stability and energy resolution, the prototype achieves 15.8% FWHM energy resolution at 5.89 keV and a gas gain of ~10⁴, demonstrating feasibility for future low-background time projection chambers in materials screening.
A prototype multiwire proportional chamber (MWPC) was developed to demonstrate the feasibility of constructing a radiopure time projection chamber with MWPC track readout to assay materials for alpha- and beta-emitting surface contaminants for future rare-event-search experiments as well as other scientific fields. The design features and assembly techniques described here are motivated by the position and energy resolution required to reconstruct alpha and beta tracks while efficiently rejecting backgrounds. Results from a test setup using an $^{55}$Fe x-ray source indicate excellent operational stability and a near-ideal energy resolution of 15.8% FWHM at 5.89 keV and a gas gain of $\sim$10$^{4}$.
Motivation & Objective
- To develop a radiopure, low-background multiwire proportional chamber (MWPC) for detecting alpha and low-energy beta particles from surface contaminants in ultra-rare-event experiments.
- To achieve high energy and position resolution for identifying isotopes via their characteristic decay lines and track reconstruction.
- To ensure operational stability and low energy threshold through precision construction, radiopure materials, and gas gain control.
- To demonstrate feasibility of integrating the MWPC into a time projection chamber (TPC) design for fiducialized background rejection.
- To enable sensitive screening of materials for trace radioisotopes such as ²¹⁰Po, ²¹⁰Pb, and other beta-emitters with low-energy emissions.
Proposed method
- The prototype MWPC uses a three-grid design with two cathode grids and a central anode grid, fabricated from radiopure materials including aluminum, copper, and G10 (with plans for radiopure replacements).
- The chamber operates with P-10 gas (Ne/CH₄) to characterize performance, with the full screener planned to use neon-methane for lower intrinsic radioactivity.
- Electrical gain and energy resolution are measured using ⁵⁵Fe x-ray sources, with energy deposition calibrated at 5.89 keV.
- Gain stability is monitored over time, with corrections applied for pressure-induced gain variations using precise pressure monitoring.
- The Diethorn model is used to estimate intrinsic energy resolution, accounting for Fano and avalanche statistics, with b ≈ 0.44 derived from Alkhazov’s model.
- Performance is evaluated via FWHM energy resolution, gain stability, minimum electric field (E_min), and electron-ion potential (ΔV), with results compared to theoretical expectations.
Experimental results
Research questions
- RQ1Can a low-background MWPC achieve sufficient energy resolution (≤15.8% FWHM) at 5.89 keV to resolve low-energy beta and alpha lines from surface contaminants?
- RQ2Does the MWPC maintain stable gas gain over extended periods, with <2% variation over 1-day intervals, under controlled conditions?
- RQ3To what extent can gain variations due to pressure fluctuations be corrected using real-time pressure monitoring and calibration?
- RQ4Can the intrinsic energy resolution of the MWPC approach the statistical limit predicted by the Fano and avalanche models (12–13%)?
- RQ5Is the use of radiopure materials and low-outgassing construction techniques sufficient to minimize intrinsic backgrounds in a low-background TPC environment?
Key findings
- The prototype MWPC achieved an energy resolution of 15.8% FWHM at 5.89 keV, approaching the statistical limit predicted by the Fano and avalanche models.
- The gas gain was measured at approximately 10⁴, with stability of ∼2% over periods exceeding one day, indicating excellent operational consistency.
- Gain variations due to pressure changes were found to follow the expected dependence, with a slope of -6.7 ± 0.3 per unit pressure change, enabling effective correction via pressure monitoring.
- The measured minimum electric field (E_min) was 38.3 ± 1.3 kV/cm, slightly below the expected 48 ± 3 kV/cm, suggesting possible influence from gas impurities or geometric factors.
- The electron-ion potential (ΔV) was measured at 26.7 ± 0.6 V, consistent with theoretical expectations and indicating efficient electron collection.
- The intrinsic resolution estimate (FWHM₀) of 12–13% aligns with the measured 15.8%, supporting the validity of the Alkhazov-based model for low-gain, low-pressure, fine-wire configurations.