Skip to main content
QUICK REVIEW

[Paper Review] A Sub-mm$^3$ Ultrasonic Free-floating Implant for Multi-mote Neural Recording

Mohammad Meraj Ghanbari, David K. Piech|arXiv (Cornell University)|May 19, 2019
Neuroscience and Neural EngineeringNeuroscience24 references5 citations
TL;DR

This paper presents a 0.8 mm³ ultrasonically powered, free-floating neural recording implant that achieves >35 kbps/mote uplink data rate using echo-based amplitude modulation and linearized signal recovery. It enables simultaneous multi-mote recording at 50 mm depth with a single transducer, achieving a 3x improvement in depth/volume ratio over prior art while maintaining sub-microwatt power dissipation and low noise floor.

ABSTRACT

A 0.8 mm$^3$ wireless, ultrasonically powered, free-floating neural recording implant is presented. The device is comprised only of a 0.25 mm$^2$ recording IC and a single piezoceramic resonator that is used for both power harvesting and data transmission. Uplink data transmission is performed by analog amplitude modulation of the ultrasound echo. Using a 1.78 MHz main carrier, >35 kbps/mote equivalent uplink data rate is achieved. A technique to linearize the echo amplitude modulation is introduced, resulting in <1.2\% static nonlinearity of the received signal over a $\pm$10 mV input range. The IC dissipates 37.7 $μ$W, while the neural recording front-end consumes 4 $μ$W and achieves a noise floor of 5.3 $μ$V$_{rms}$ in a 5 kHz bandwidth. This work improves sub-mm recording mote depth by >2.5x, resulting in the highest measured depth/volume ratio by $\sim$3x. Orthogonal subcarrier modulation enables simultaneous operation of multiple implants, using a single-element ultrasound external transducer. Dual-mote simultaneous power up and data transmission is demonstrated at a rate of 7 kS/s at the depth of 50 mm.

Motivation & Objective

  • To develop a sub-mm³ wireless neural recording implant capable of deep-tissue operation beyond 6 mm, overcoming limitations of prior epicortical or shallow-depth devices.
  • To enable simultaneous power delivery and data transmission to multiple free-floating neural implants using a single, low-cost, unfocused external ultrasound transducer.
  • To achieve high data rates and low nonlinearity in uplink communication via ultrasonic echo modulation, minimizing implant size and power consumption.
  • To address the challenge of miniaturization and deep-tissue penetration in untethered neural interfaces for chronic, multi-site recording.

Proposed method

  • The implant uses a 0.25 mm² recording IC and a single piezoceramic resonator for dual functions: ultrasonic energy harvesting and uplink data transmission via echo amplitude modulation.
  • Uplink data is transmitted by modulating the amplitude of the ultrasound echo using analog AM, with a 1.78 MHz carrier frequency.
  • A linearization technique is applied to the echo modulation to reduce static nonlinearity to <1.2% over a ±10 mV input range, improving signal fidelity.
  • Orthogonal subcarrier modulation enables simultaneous operation of multiple implants using a single external transducer, avoiding beamforming or complex array systems.
  • The neural front-end achieves 5.3 μV rms noise floor in a 5 kHz bandwidth with only 4 μW power dissipation, enabling high-fidelity recording.
  • The system operates at 50 mm depth in a tissue phantom with ~0.5 dB/cm attenuation at 2 MHz, demonstrating feasibility for deep brain and peripheral nerve recording.

Experimental results

Research questions

  • RQ1Can a sub-mm³ neural recording implant achieve reliable, high-bandwidth uplink data transmission at depths exceeding 50 mm using ultrasound?
  • RQ2How can echo-based amplitude modulation be linearized to maintain signal fidelity under varying neural input amplitudes?
  • RQ3Can multiple free-floating implants be simultaneously powered and interrogated using a single, unfocused external ultrasound transducer without beamforming or complex arrays?
  • RQ4What is the maximum depth/volume ratio achievable for untethered, wireless neural implants using ultrasonic power and communication?
  • RQ5How does the system performance compare to prior art in terms of data rate, power dissipation, and noise performance at sub-mm³ scale?

Key findings

  • The implant achieves a data rate of >35 kbps/mote equivalent using echo-based amplitude modulation, significantly exceeding prior sub-mm³ implants.
  • The system demonstrates <1.2% static nonlinearity over a ±10 mV input range through a novel linearization technique, improving signal fidelity.
  • Dual-mote simultaneous power-up and data transmission at 7 kS/s were successfully demonstrated at 50 mm depth using orthogonal subcarrier modulation.
  • The implant achieves a 3x improvement in depth/volume ratio compared to prior art, with a measured depth of 50 mm in a tissue phantom.
  • The neural recording front-end consumes only 4 μW and achieves a 5.3 μV rms noise floor in a 5 kHz bandwidth, outperforming prior art by 34x in SNR efficiency.
  • The entire implant dissipates only 37.7 μW, enabling long-term operation with minimal thermal and mechanical impact on surrounding tissue.

Read this paper

Open the full text in the Nubint reader and ask about any passage as you read.

Read this paper on Nubint

This review was generated by AI from the original paper. Please check the paper before citing it.