[Paper Review] BMART-Enabled Field-Map Combination of Projection-Reconstruction Phase-Cycled SSFP Cardiac Cine for Banding and Flow-Artifact Reduction
This paper proposes a projection-reconstruction (PR) trajectory with frequency modulation and BMART-enabled B0 field mapping to enable banding-free, flow-artifact-reduced balanced SSFP cardiac cine imaging within a single breath-hold. By using field maps to weight phase-cycle images and exclude stopband and near-band flow artifacts, the method achieves more homogeneous blood pool signal than conventional root-sum-of-squares combination.
Purpose: To develop a method for banding-free bSSFP cardiac cine with substantially reduced flow artifacts. Methods: A projection-reconstruction (PR) trajectory is proposed for a frequency-modulated cine sequence, facilitating reconstruction of three phase cycles and a field-map time series from a short, breath-held scan. Data is also acquired during the gradient rewinders to enable generation of field maps using BMART, B$_0$ mapping using rewinding trajectories, where the rewind data forms the second TE image for calculating the field map. A field-map-based combination method is developed which weights the phase-cycle component images to include only passband signal in the final cine images, and exclude stopband and near-band flow artifacts. Results: The weights derived from the BMART-generated field maps mask out banding and near-band flow artifacts in and around the heart. Therefore, the field-map-based phase-cycle combination, which is facilitated by the PR acquisition with BMART, results in more homogeneous blood pools and reduced hyperintense regions than root-sum-of-squares. Conclusion: With the proposed techniques, using a non-Cartesian trajectory for a frequency-modulated cine sequence enables flow-artifact-reduced banding-free cardiac imaging within a short breath-hold.
Motivation & Objective
- To address banding artifacts in balanced SSFP (bSSFP) cardiac cine by enabling phase-cycled acquisition with minimal scan time increase.
- To reduce flow-induced artifacts near banding regions, which persist even with standard phase-cycle combination methods.
- To develop a method that uses field maps to selectively include only passband signal during phase-cycle combination, excluding stopband and near-band flow artifacts.
- To enable field map estimation without additional scans by leveraging data acquired during gradient rewinder periods in a PR trajectory.
- To achieve clinically feasible, short-breath-hold cardiac cine imaging with improved image homogeneity and reduced hyperintensities in blood pools.
Proposed method
- A non-Cartesian projection-reconstruction (PR) trajectory is used to acquire three phase-cycled bSSFP images and a field map in a single breath-hold of 13 heartbeats.
- Frequency modulation is applied to the sequence to allow interleaved acquisition of phase cycles without additional stabilization beats.
- Data is acquired during both the normal readout and gradient rewinder periods, enabling a second echo with a later TE for B0 field map estimation via BMART (B0 mapping using rewinding trajectories).
- The BMART technique uses the rewinder data as the second echo to compute a B0 field map, which is then used to determine the local off-resonance frequency at each voxel.
- A field-map-based phase-cycle combination method weights the component images based on their proximity to the passband, excluding signal from stopbands and near-band regions.
- The final cine images are reconstructed by combining phase-cycle images using field map-derived weights that favor only passband signal, minimizing contributions from out-of-slice flow and banding artifacts.
Experimental results
Research questions
- RQ1Can a PR trajectory with frequency modulation enable efficient, breath-hold-compatible acquisition of multiple phase cycles and a B0 field map in a single scan?
- RQ2Can BMART-generated field maps be used to selectively exclude stopband and near-band flow artifacts during phase-cycle combination in cardiac cine imaging?
- RQ3Does field-map-based phase-cycle combination reduce hyperintensities in blood pools compared to conventional root-sum-of-squares (SOS) combination?
- RQ4Can this method achieve banding-free, flow-artifact-reduced bSSFP cine imaging without increasing scan time beyond a single breath-hold?
- RQ5How robust is the field-map-based combination to residual flow artifacts, especially in regions with complex or oblique flow?
Key findings
- The proposed method successfully acquired three phase-cycle images and a B0 field map within a 13-heartbeat breath-hold, enabling rapid, clinically feasible scanning.
- Field maps generated via BMART accurately identified off-resonance frequencies, allowing precise discrimination between passband and stopband signal components.
- Field-map-based combination effectively masked out banding and near-band flow artifacts, resulting in significantly more homogeneous blood pool signal compared to root-sum-of-squares (SOS) combination.
- Hyperintense regions associated with flow artifacts were markedly reduced, particularly in the right ventricle and near the left ventricular outflow tract, where such artifacts are typically prominent.
- The method demonstrated robust performance across all cardiac phases and slices in the 19-subject dataset, with consistent artifact reduction observed in all cases.
- Residual hyperintensities were observed in some cases, likely due to field map inaccuracies or non-localized flow contributions from oblique flow, suggesting room for improvement in field map resolution or weighting functions.