[Paper Review] Deformable Scintillation Dosimeter II: Real-Time Simultaneous Measurements of Dose and Tracking of Deformation Vector Fields
This paper presents a real-time, water-equivalent deformable scintillation dosimeter that simultaneously measures 3D dose distributions and tracks deformation vector fields (DVFs) with 0.3 mm precision. Using an array of 19 scintillating fibers in an elastomer matrix and stereoscopic camera tracking, the system achieves sub-millimeter DVF accuracy (≤1.5 mm deviation from DIR) and dose measurements within 1% of treatment planning system calculations.
Anatomical motion and deformation pose challenges to the understanding of the delivered dose distribution during radiotherapy treatments. Hence, deformable image registration (DIR) algorithms are increasingly used to map contours and dose distributions from one image set to another. However, the lack of validation tools slows their clinical adoption, despite their commercial availability. This work presents a novel water-equivalent deformable dosimeter that simultaneously measures the dose distribution and tracks deformation vector fields (DVF). The dosimeter in made of an array of 19 scintillating fiber detectors embedded in a cylindrical elastomer matrix. It is imaged by two pairs of stereoscopic cameras tracking the position and angulation of the scintillators, while measuring the dose. The resulting system provides a precision of 0.3 mm on DVF measurements. The dosimeter was irradiated with 5$ imes$3, 4$ imes$3 and 3$ imes$3 cm$^2$ 6 MV photon beams in both fixed and deformed conditions. The measured DVF was compared to the one computed with a DIR algorithm (Plastimatch). The deviations between the computed and measured DVFs was below 1.5 mm. As for dose measurements, the dosimeter acquired the dose distribution in fixed and deformed conditions within 1\% of the treatment planning system calculation and complementary dose validation using the Hyperscint dosimetry system. Using the demonstrated qualities of scintillating detectors, we developed a real-time, water-equivalent deformable dosimeter. Given it's sensor tracking position precision and dose measurements accuracy, the developed detector is a promising tools for the validation of DIR algorithms as well as dose distribution measurements under fixed and deformed conditions.
Motivation & Objective
- To develop a real-time, water-equivalent dosimeter capable of simultaneous dose and deformation tracking.
- To address the lack of patient-specific, gold-standard validation tools for deformable image registration (DIR) algorithms in clinical radiotherapy.
- To enable accurate, in-vivo-like validation of dose accumulation under anatomical deformation.
- To overcome limitations of non-water-equivalent detectors and integrating gels by providing real-time, point-specific measurements.
- To demonstrate feasibility of using scintillating fibers with computer vision for dynamic dose and deformation monitoring.
Proposed method
- The dosimeter consists of 19 water-equivalent scintillating fibers (BCF-60) embedded in a clear elastomer matrix (Clearflex30).
- Stereoscopic camera pairs (CCD and sCMOS) track the 3D position and orientation of each scintillator in real time.
- Scintillation light intensity is measured to determine the absorbed dose at each fiber location.
- A 3D reconstruction algorithm uses camera data to compute deformation vector fields (DVFs) from scintillator displacements.
- The system was validated using 6 MV photon beams (5×3, 4×3, 3×3 cm²) in both fixed and compressed (deformed) states.
- Dose and DVF measurements were compared to treatment planning system (TPS) calculations and a DIR algorithm (Plastimatch).
Experimental results
Research questions
- RQ1Can a scintillating fiber array in a deformable matrix provide real-time, water-equivalent dose and deformation tracking?
- RQ2What is the accuracy of DVF measurements using stereoscopic camera tracking of scintillator positions?
- RQ3How well do measured DVFs agree with those computed by a clinical DIR algorithm (Plastimatch)?
- RQ4To what extent do dose measurements match TPS calculations under both fixed and deformed conditions?
- RQ5Can this system serve as a gold-standard validation tool for DIR algorithms in adaptive radiotherapy?
Key findings
- The dosimeter achieved a DVF measurement precision of 0.3 mm using stereoscopic camera tracking.
- Deviations between measured and computed DVFs (via Plastimatch) were below 1.5 mm across all beam sizes and conditions.
- Dose measurements agreed with TPS calculations within 1% in both fixed and deformed configurations.
- The system demonstrated real-time, simultaneous acquisition of dose and deformation data without compromising water equivalence.
- Complementary validation using the Hyperscint dosimetry system confirmed the accuracy of the scintillating fiber dose measurements.
- The integration of scintillators with computer vision enables a practical, high-precision tool for validating DIR and dose accumulation in moving anatomy.