[Paper Review] Improved method for measuring low concentration radium and its application to the Super-Kamiokande Gadolinium project

S. Ito, K. Ichimura|arXiv (Cornell University)|Jun 17, 2020
Radiation Detection and Scintillator TechnologiesPhysics and Astronomy12 references6 citations
TL;DR

This study presents an improved chemical extraction method using 'Empore Radium Rad Disk' resin to measure low-concentration 226Ra in high-matrix samples, achieving >99.9% recovery, a threefold increase in processing speed (3 L/h), and a detection sensitivity of 0.1 mBq, which was successfully applied to certify Gd2(SO4)3·8H2O for the Super-Kamiokande Gadolinium project below the 0.5 mBq kg−1 threshold.

ABSTRACT

Chemical extraction using a molecular recognition resin named "Empore Radium Rad Disk" was developed to improve sensitivity for the low concentration of radium (Ra). Compared with the previous method, the extraction process speed was improved by a factor of three and the recovery rate for $^{226}$Ra was also improved from 81$\pm$4% to $>$99.9%. The sensitivity on the 10$^{-1}$ mBq level was achieved using a high purity germanium detector. This improved method was applied to determine $^{226}$Ra in Gd$_2$(SO$_4$)$_3{\cdot}$8H$_2$O which will be used in the Super-Kamiokande Gadolinium project. The improvement and measurement results are reported in this paper.

Motivation & Objective

  • To develop a more sensitive and efficient method for measuring low-activity 226Ra in high-matrix materials like Gd2(SO4)3·8H2O.
  • To address the limitations of the previous method, including low recovery (81±4%) and slow processing (1 L/h), which were insufficient for SK-Gd's stringent requirements.
  • To achieve a detection limit below 0.5 mBq kg−1 for 226Ra in Gd2(SO4)3·8H2O, the maximum allowed level in the SK-Gd project.
  • To validate the method using real samples and ensure all Gd2(SO4)3·8H2O lots meet the required radioimpurity standards before loading into the detector.

Proposed method

  • Chemical extraction using 'Empore Radium Rad Disk' resin, a molecular recognition resin with high selectivity for Ra, was employed to concentrate 226Ra from large-volume, high-matrix samples.
  • The extraction was performed via vacuum filtration at 50 mL/min (3 L/h), tripling the throughput compared to the prior method (1 L/h).
  • The disk was directly measured using a high-purity germanium (HPGe) detector to quantify 226Ra via gamma-ray emissions from its daughters, 214Pb (609 keV) and 214Bi (352 keV, 1764 keV).
  • Detection efficiency was calibrated using Monte Carlo simulations, yielding 15.9% efficiency for 352 keV gamma rays from 214Bi.
  • Recovery rates were validated using Ba as a tracer isotope, with ICP-MS analysis showing <0.1% Ba remaining in effluent, confirming >99.9% Ra recovery.
  • Background and blank levels were measured using 17 disks, yielding a procedure blank of 0.11+0.04−0.02 mBq per disk, significantly lower than the prior method (0.3±0.2 mBq).

Experimental results

Research questions

  • RQ1Can a chemical extraction method be developed to achieve sub-mBq sensitivity for 226Ra in high-matrix samples like Gd2(SO4)3·8H2O?
  • RQ2How can the recovery rate of 226Ra be improved beyond the 81±4% of the previous method while reducing processing time?
  • RQ3To what extent does the presence of high concentrations of Gd2(SO4)3·8H2O interfere with Ra extraction and detection?
  • RQ4What is the actual 226Ra concentration in commercially available Gd2(SO4)3·8H2O lots, and do they meet the SK-Gd project’s 0.5 mBq kg−1 threshold?
  • RQ5Can the improved method be reliably applied to certify large-scale batches of Gd2(SO4)·8H2O for use in low-background neutrino experiments?

Key findings

  • The improved method achieved a 226Ra detection sensitivity of 0.1 mBq, corresponding to a limit of <0.4 mBq kg−1 for 13 tons of Gd2(SO4)3·8H2O, meeting the SK-Gd requirement of 0.5 mBq kg−1.
  • The recovery rate of 226Ra was increased from 81±4% to >99.9%, with Ba tracer studies confirming minimal loss during extraction.
  • Processing time was tripled to 3 L/h, significantly improving throughput for large-scale sample analysis.
  • The procedure blank was reduced to 0.11+0.04−0.02 mBq per disk, enhancing detection limit and reducing background uncertainty.
  • All measured Gd2(SO4)3·8H2O lots (n=6) showed 226Ra concentrations below the 90% confidence level upper limits of 0.2–0.5 mBq kg−1, confirming compliance with SK-Gd standards.
  • The method was successfully applied to certify 14 tons of Gd2(SO4)3·8H2O, with all lots confirmed to be within the allowed radioimpurity levels.

Read this paper

Open the full text in the Nubint reader and ask about any passage as you read.

Read this paper on Nubint

This review was generated by AI from the original paper. Please check the paper before citing it.