S. Rozov, E. Armengaud|arXiv (Cornell University)|Jan 25, 2010
Nuclear Physics and ApplicationsPhysics and Astronomy14 citations
TL;DR

This study presents continuous, high-precision monitoring of thermal neutron flux using a 3He-filled proportional counter in the deep underground LSM laboratory. It reports a daily average flux of 3.57 × 10⁻⁶ neutrons/cm²/sec near the EDELWEISS-II experiment, with variations exceeding a factor of three between locations, and confirms no significant temporal fluctuations over days or weeks.

ABSTRACT

This paper describes precise measurements of the thermal neutron flux in the LSM underground laboratory in proximity of the EDELWEISS-II dark matter search experiment together with short measurements at various other locations. Monitoring of the flux of thermal neutrons is accomplished using a mobile detection system with low background proportional counter filled with $^3$He. On average 75 neutrons per day are detected with a background level below 1 count per day (cpd). This provides a unique possibility of a day by day study of variations of the neutron field in a deep underground site. The measured average 4$π$ neutron flux per cm$^{2}$ in the proximity of EDELWEISS-II is $Φ_{MB}=3.57\pm0.05^{stat}\pm0.27^{syst} imes 10^{-6}$ neutrons/sec. We report the first experimental observation that the point-to-point thermal neutron flux at LSM varies by more than a factor two.

Motivation & Objective

  • To measure and monitor the thermal neutron flux in the deep underground LSM laboratory with high precision.
  • To assess temporal stability of the neutron flux over daily and weekly timescales.
  • To investigate spatial variations in thermal neutron flux across different locations in the LSM facility.
  • To support background characterization for next-generation dark matter and double beta decay experiments at LSM.
  • To validate the use of 3He proportional counters for low-background, high-sensitivity neutron flux monitoring in underground environments.

Proposed method

  • A mobile 3He-filled proportional counter (CHM-57) was used to detect thermal neutrons via the n + ³He → T + p reaction.
  • The detector was equipped with Teflon and copper layers to suppress α-decay background from U/Th progeny.
  • Energy spectra were acquired using a charge-sensitive preamplifier and 12-bit ADC, with pulse height analysis focused on the 660–830 keV region.
  • Calibration was performed using a PuBe neutron source moderated by polyethylene to simulate thermal neutron response.
  • Measurements were conducted at 10 distinct locations across LSM, including near EDELWEISS-II, NEMO-3, and control rooms.
  • Statistical analysis of time-series data used χ² tests to evaluate constancy of flux over 3-hour, daily, and weekly intervals.

Experimental results

Research questions

  • RQ1What is the spatial distribution of thermal neutron flux across different locations in the LSM underground laboratory?
  • RQ2How stable is the thermal neutron flux over short timescales (3 hours, days, weeks) at a fixed location?
  • RQ3What is the absolute thermal neutron flux near the EDELWEISS-II dark matter experiment?
  • RQ4To what extent do material composition and water content influence local neutron flux variations?
  • RQ5Can a 3He-based detector achieve sufficient sensitivity and low background to enable continuous, day-by-day monitoring of thermal neutrons?

Key findings

  • The average thermal neutron flux near the EDELWEISS-II experiment is Φ_MB = 3.57 × 10⁻⁶ neutrons/cm²/sec with statistical and systematic uncertainties of ±0.05 and ±0.27 × 10⁻⁶, respectively.
  • Thermal neutron flux varies by more than a factor of three between different locations in the LSM laboratory, with the highest flux (6.6 × 10⁻⁶ cm⁻²s⁻¹) measured near the laboratory wall.
  • No significant temporal variations in neutron flux were observed over 3-hour, daily, or weekly intervals at any location, with χ²/ndf values consistent with a constant flux.
  • The detector achieved a daily count rate of 75 counts per day with a background level below 1 count per day, enabling high-precision monitoring.
  • Systematic uncertainties were dominated by detector sensitivity (6.2%) and active volume (4.6%), with total systematic error estimated at 7.7%.
  • The study reports the first experimental observation of point-to-point thermal neutron flux variations exceeding a factor of two in a deep underground site.