[Paper Review] A method for measuring the contact area in instrumented indentation testing by tip scanning probe microscopy imaging
This paper proposes a novel post mortem method for measuring the contact area in instrumented indentation testing using altitude images from scanning probe microscopy (SPM), including indentation tip scanning probe microscopy (ITSPM). The method achieves systematically lower error levels—consistently under 5%—across a wide range of materials, outperforming three classical direct methods, and is robust to machine and sample holder stiffness effects while requiring no indenter calibration or adjustable parameters.
The determination of the contact area is a key step to derive mechanical properties such as hardness or an elastic modulus by instrumented indentation testing. Two families of procedures are dedicated to extracting this area: on the one hand, post mortem measurements that require residual imprint imaging, and on the other hand, direct methods that only rely on the load vs. the penetration depth curve. With the development of built-in scanning probe microscopy imaging capabilities such as atomic force microscopy and indentation tip scanning probe microscopy, last generation indentation devices have made systematic residual imprint imaging much faster and more reliable. In this paper, a new post mortem method is introduced and further compared to three existing classical direct methods by means of a numerical and experimental benchmark covering a large range of materials. It is shown that the new method systematically leads to lower error levels regardless of the type of material. Pros and cons of the new method vs. direct methods are also discussed, demonstrating its efficiency in easily extracting mechanical properties with an enhanced confidence.
Motivation & Objective
- To develop a more accurate and reliable method for measuring the contact area in instrumented indentation testing.
- To address the limitations of direct methods, which rely solely on load-displacement curves and are sensitive to piling-up, sinking-in, and machine stiffness effects.
- To leverage advancements in indentation tip scanning probe microscopy (ITSPM) for systematic, automated residual imprint imaging.
- To provide a user-independent, calibration-free approach that is insensitive to surface detection errors and spurious stiffness influences.
Proposed method
- The method uses altitude images from SPM, including ITSPM, to reconstruct the residual imprint shape and extract the projected contact area.
- It relies on the assumption that the residual imprint remains unchanged from the end of indentation to imaging, which holds for most materials except time-dependent ones like polymers.
- The contact area is computed directly from the 3D topography of the imprint without requiring any adjustable parameters or indenter calibration.
- The method is compatible with all indenter geometries and is fully automated, enabling integration into modern indentation systems with ITSPM capabilities.
- It avoids reliance on penetration depth (h) and thus eliminates errors from machine or sample holder compliance.
- The approach is validated using both finite element simulations and experimental tests across a broad range of materials with varying mechanical behaviors.
Experimental results
Research questions
- RQ1Can a post mortem SPM-based method achieve consistently lower measurement errors than classical direct methods in contact area estimation?
- RQ2How does the proposed method perform across materials with diverse mechanical responses, including elastic, plastic, and anisotropic behaviors?
- RQ3To what extent is the method robust to machine and sample holder stiffness effects that commonly affect direct methods?
- RQ4Does the method remain accurate for materials exhibiting time-dependent deformation, such as creep or relaxation?
- RQ5Can the method be fully automated and integrated into modern indentation systems equipped with ITSPM?
Key findings
- The proposed method achieves contact area measurement errors consistently below 5% across all tested materials, significantly outperforming the three classical direct methods.
- The method systematically reduces error levels regardless of material type, including metals, ceramics, and polymers, as demonstrated in both numerical and experimental benchmarks.
- Unlike direct methods, the proposed method is insensitive to machine and sample holder stiffness, which otherwise introduce errors in penetration depth and contact stiffness measurements.
- The method is unaffected by surface detection errors because it does not rely on the penetration depth (h) for area calculation.
- The method is fully automated, user-independent, and requires no indenter calibration, making it highly suitable for routine and high-throughput mechanical property evaluation.
- The method fails for time-dependent materials such as pure aluminum and tungsten single crystals due to post-indentation creep, which causes systematic overestimation of the contact area by factors of ~1.6 and ~2.7, respectively.
Read this paper
Open the full text in the Nubint reader and ask about any passage as you read.
Read this paper on NubintThis review was generated by AI from the original paper. Please check the paper before citing it.