[Paper Review] arXiv : FPGA Based Pico-second Time Measurement System for a DIRC-like TOF Detector
This paper presents an FPGA-based pico-second time measurement system for a DIRC-like Time-of-Flight (DTOF) detector, using fused silica radiators and micro-channel plate PMTs. The system achieves a coincidence time resolution (CTR) of 15.0 ps in beam tests at CERN, demonstrating high precision, compactness, cost-effectiveness, and scalability for large-scale particle physics applications.
A prototype of DIRC-like Time-of-Flight detector (DTOF), including a pico-second time measurement electronics, is developed and tested preliminarily. The basic structure of DTOF is composed of a fused silica radiator connected to fast micro-channel plate PMTs (MCP-PMT), and readout by a dedicated FPGA (Field Programmable Gate Array) based front-end electronics. The full electronics chain consists of a programmable differential amplifier, a dual-threshold differential discriminator, and a timestamp Time-to-Digital convertor. By splitting a MCP-PMT output signal into two identical electronics chains, the coincidence time resolution (CTR) of pure electronics was measured as 5.6 ps. By the beam test in H4 (150GeV/c, Muon) at CERN, the intrinsic CTR of the whole detector prototype reaches 15.0 ps without using time-amplitude correction. The test results demonstrate that the FPGA based front-end electronics could achieve an excellent time performance for TOF detectors. It is very compact, cost effective with a high multi-channel capacity and short measurement dead time, which is very suitable for practical applications of large-scale high performance TOF detectors in particle physics spectrometer.
Motivation & Objective
- To develop a compact, cost-effective, and high-channel-capacity front-end electronics system for high-precision time-of-flight measurements in particle physics.
- To achieve sub-10 ps coincidence time resolution (CTR) using FPGA-based signal processing for a DIRC-like TOF detector.
- To validate the system's performance in a real beam environment using 150 GeV/c muons at CERN's H4 beamline.
- To demonstrate the feasibility of using FPGA-based electronics for large-scale, high-performance TOF detectors with short dead time.
Proposed method
- The system uses a fused silica radiator coupled to fast micro-channel plate PMTs (MCP-PMTs) for photon detection.
- A programmable differential amplifier shapes the MCP-PMT signals to optimize signal-to-noise ratio.
- Dual-threshold differential discriminators convert analog signals into digital triggers with reduced timing jitter.
- A timestamp Time-to-Digital Converter (TDC) records the arrival time of signals with pico-second resolution.
- Two identical electronics chains process split signals from a single MCP-PMT to enable coincidence timing measurements.
- The FPGA-based front-end electronics integrates all signal processing stages, enabling high multi-channel capacity and low dead time.
Experimental results
Research questions
- RQ1Can an FPGA-based front-end electronics system achieve sub-10 ps coincidence time resolution in a DIRC-like TOF detector?
- RQ2What is the intrinsic coincidence time resolution of the complete DTOF prototype when tested with high-energy muons at CERN?
- RQ3How does the system perform in terms of timing precision when using only pure electronics without time-amplitude correction?
- RQ4To what extent does the FPGA-based design support high multi-channel capacity and short dead time for large-scale TOF detectors?
- RQ5Can the system maintain high timing performance while remaining compact and cost-effective for practical particle physics applications?
Key findings
- The pure electronics coincidence time resolution (CTR) was measured at 5.6 ps using split signals from a single MCP-PMT.
- In beam tests at CERN with 150 GeV/c muons, the intrinsic CTR of the full DTOF prototype reached 15.0 ps without time-amplitude correction.
- The FPGA-based front-end electronics demonstrated excellent timing performance, enabling high-precision time measurements suitable for high-energy physics.
- The system is compact, cost-effective, and supports high multi-channel capacity, making it suitable for large-scale TOF detectors.
- The system exhibits short measurement dead time, enhancing its suitability for high-rate particle detection environments.
- The results confirm the feasibility of using FPGA-based electronics for next-generation, high-performance TOF detectors in particle physics spectrometers.
Read this paper
Open the full text in the Nubint reader and ask about any passage as you read.
Read this paper on NubintThis review was generated by AI from the original paper. Please check the paper before citing it.