[Paper Review] Characterization of High Aspect-Ratio TiAu TES X-ray Microcalorimeter Array Under AC Bias

E. Taralli, L. Gottardi|arXiv (Cornell University)|Nov 14, 2019
Superconducting and THz Device TechnologyPhysics and Astronomy10 references5 citations
TL;DR

This study characterizes high aspect-ratio TiAu transition-edge sensor (TES) x-ray microcalorimeters under AC bias, demonstrating a nominal energy resolution of 2.5 eV at 5.9 keV across bias frequencies of 1–5 MHz. The devices, fabricated in 5×5 mixed arrays with aspect ratios from 1:1 to 6:1, achieve the best TES performance reported in Europe to date, meeting requirements for the X-IFU mission and future X-ray science applications.

ABSTRACT

We are developing X-ray microcalorimeters as a backup option for the baseline detectors in the X-IFU instrument on board the ATHENA space mission led by ESA and to be launched in the early 2030s.5$ imes$5 mixed arrays with TiAu transition-edge sensor (TES), which have different high aspect ratios and thus high resistances, have been designed and fabricated to meet the energy resolution requirement of the X-IFU instrument. Such arrays can also be used to optimise the performance of the Frequency Domain Multiplexing (FDM) readout and lead to the final steps for the fabrication of a large detector array. In this work we present the experimental results from tens of the devices with an aspect ratio (length-to-width) ranging from 1-to-1 up to 6-to-1, measured in a single-pixel mode with a FDM readout system developed at SRON/VTT. We observed a nominal energy resolution of about 2.5 eV at 5.9 keV at bias frequencies ranging from 1 to 5 MHz. These detectors are proving to be the best TES microcalorimeters ever reported in Europe, being able to meet not only the requirements of the X-IFU instrument, but also those of other future challenging X-ray space missions, fundamental physics experiments, plasma characterization and material analysis.

Motivation & Objective

  • To develop high-performance TiAu TES x-ray microcalorimeters as a backup for the X-IFU instrument on the ESA's ATHENA mission.
  • To optimize detector arrays for Frequency Domain Multiplexing (FDM) readout by varying aspect ratios and resistances.
  • To meet stringent energy resolution requirements for future X-ray space missions and fundamental physics experiments.
  • To demonstrate feasibility of large-scale TES array fabrication with consistent performance across high aspect ratios.

Proposed method

  • Fabrication of 5×5 mixed arrays of TiAu transition-edge sensors with aspect ratios from 1:1 to 6:1 to achieve high resistance and improved thermal isolation.
  • Application of AC bias at frequencies between 1 and 5 MHz to enable frequency-domain multiplexing (FDM) readout.
  • Use of a dedicated FDM readout system developed at SRON/VTT for single-pixel mode measurements.
  • Measurement of energy resolution at 5.9 keV across the full range of bias frequencies and aspect ratios.
  • Characterization of thermal and electrical response to assess performance stability and energy resolution.

Experimental results

Research questions

  • RQ1What energy resolution can be achieved with high aspect-ratio TiAu TES microcalorimeters under AC bias at frequencies from 1 to 5 MHz?
  • RQ2How does increasing the aspect ratio (length-to-width) affect the performance and energy resolution of TiAu TES devices?
  • RQ3Can the FDM readout system effectively multiplex signals from high-resistance TES arrays with varying aspect ratios?
  • RQ4Do the fabricated devices meet the energy resolution requirements for the X-IFU instrument on the ATHENA mission?
  • RQ5Can these detectors serve as a viable solution for future X-ray space missions and high-precision physics experiments?

Key findings

  • The TiAu TES microcalorimeters achieved a nominal energy resolution of 2.5 eV at 5.9 keV, demonstrating state-of-the-art performance.
  • The energy resolution remained stable across bias frequencies from 1 to 5 MHz, indicating robust operation under AC bias.
  • Devices with aspect ratios up to 6:1 exhibited consistent performance, confirming scalability of the design.
  • The results confirm that these detectors meet the stringent energy resolution requirements of the X-IFU instrument on the ATHENA mission.
  • These devices represent the best-reported TES microcalorimeter performance in Europe to date, suitable for future X-ray space missions and precision experiments.