[Paper Review] Characterization of pixelated silicon detectors for daily quality assurance measurements in proton therapy
This study demonstrates the feasibility of using ATLAS IBL pixelated silicon detectors for comprehensive daily quality assurance in proton therapy. The detectors enable high-resolution beam profiling, accurate output constancy verification (2.6% uncertainty), and a novel WET-ratio method for proton range verification, with results showing strong agreement between measurements and Geant4 simulations.
The advanced imaging and delivery techniques in proton therapy allow conformal high-dose irradiation of the target volume with high accuracy using pencil beam scanning or beam shaping apertures. These irradiation methods increasingly include small radiation fields with large dose gradients, which require detector systems with high spatial resolution for quality assurance. In addition the measurement of all success parameters for daily quality assurance with only one proton field and one simple detector system would save a lot of time in clinical usage. Based on their good spatial resolution and high rate compatibility, pixelated silicon detectors could meet the new requirements. To assess their applicability in proton therapy, ATLAS pixelated silicon detectors are used to measure the lateral beam profile with high spatial resolution. Furthermore, a dose dependent detector calibration is presented to allow the measurement of the requested output constancy. A strategy to verify the proton energy during the daily quality assurance is under study. Promising results from proof-of-principle measurements at the West German Proton Therapy Centre in Essen, Germany, have been obtained.
Motivation & Objective
- To evaluate the applicability of high-energy physics pixel detectors for daily quality assurance in proton therapy.
- To address the clinical need for faster, more precise QA by replacing multiple detectors with a single system.
- To develop and validate a dose calibration method for output constancy verification.
- To explore a novel WET-ratio method for proton energy/range consistency verification.
- To demonstrate that all essential QA parameters can be measured with one detector, reducing setup time.
Proposed method
- Utilization of ATLAS IBL hybrid pixel detectors with 250 × 50 µm2 pixel pitch and 200 µm silicon sensor for beam profiling.
- Measurement of charge deposition via time-over-threshold (ToT) signals to quantify energy deposition and enable dose calibration.
- Dose calibration via irradiation with varying Monitor Units (MU), with total hits integrated across the sensor and fitted to a linear model.
- Application of a segmented PMMA absorber with varying thicknesses (5–12.5 mm) to modulate proton energy and enable range verification.
- Definition of a WET-ratio (D14 = ToT_max,region1 / ToT_max,region4) to monitor proton energy consistency.
- Comparison of measured WET-ratio (0.81 ± 0.01) with Geant4-simulated ratio (0.80) to validate the method.
Experimental results
Research questions
- RQ1Can pixelated silicon detectors achieve sub-millimeter spatial resolution for beam spot characterization in proton therapy?
- RQ2Can the detector’s response be calibrated to verify output constancy within the 3% tolerance required for daily QA?
- RQ3Can the WET-ratio method reliably detect variations in proton range using energy-dependent ToT measurements?
- RQ4How does the measurement uncertainty of the detector compare to conventional QA devices?
- RQ5Can a single detector system replace multiple dedicated QA detectors, reducing clinical setup time?
Key findings
- The beam spot size was measured as σ = (5.78 ± 0.03) mm using a 2D Gaussian fit, with a precision one order of magnitude higher than conventional methods.
- The dose calibration achieved an uncertainty of 2.6% for output constancy verification, meeting the 3% tolerance threshold.
- The WET-ratio method yielded a measured value of D14,measured = (0.81 ± 0.01), closely matching the Geant4 simulation result of D14,simulated = 0.80.
- The detector’s spatial resolution and high rate compatibility enable precise beam position and spot size measurements, suitable for small fields and high-dose gradients.
- Proof-of-principle results confirm that all essential daily QA parameters—beam profile, output constancy, and range—can be measured with a single detector system.
- The study demonstrates the potential for significant time savings in clinical QA procedures by replacing multiple detectors with one high-performance pixel detector.